# DEN150008: INVOcell

- Pathway: De Novo (Direct)
- Applicant: Invo Bioscience
- Decided: Nov 2, 2015 after 252 days
- Product code: OYO (Culture, Intravaginal, Assisted Reproduction)
- Device class: 2
- Regulation: 21 CFR 884.6165
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
1 later device descends from it; 1 cites it directly.
0 of 2 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. De Novo decisions do not claim a predicate device.

## Most influential descendants
- K222932: INVOcell Intravaginal Culture System (2023), 0 descendants of its own

## Indications for use
The INVOcell Intravaginal Culture System consists of the following components: The INVOcell Culture Device is indicated for use in preparing, holding, and transferring human gametes or embryos during In Vitro Fertilization/Intra Vaginal Culture (IVF/IVC) and Intra-cytoplasmic Sperm Injection Fertilization/Intravaginal Culture (ICSI/IVC) procedures. The INVOcell Culture Device is indicated for use with the INVOcell Retention Device and the INVOcell Holding Block. The INVOcell Culture Device is not indicated for incubation periods exceeding 72h. The INVOcell Retention Device is indicated for use with the INVOcell Culture Device to aid in retention of the INVOcell Culture Device in the vaginal cavity during the incubation period. The INVOcell Retention Device is not indicated for use exceeding 72 hours. The INVOcell Holding Block is indicated for use with the INVOcell Culture Device to aid in temperature maintenance of the INVOcell Culture Device during loading and collection procedures and to aid in positioning and observation of the INVOcell Culture Device during human gamete/embryo loading and collection procedures.

Source page: https://510kreport.com/device/DEN150008
Methodology: https://510kreport.com/methodology