510(k) Report

De Novo · Granted

Granted

Terragene Bionova Photon Biological Indicator (BT225); Terragene Bionova Photon Auto-Reader Incubator (BPH)

DEN220042De NovoGranted Jul 12, 2024742-day reviewClass II21 CFR 880.2806

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Terragene S.A. · QVB Biological Sterilization Indicator With Indirect Growth Detection

Granted on Jul 12, 2024 as a new device type through De Novo after a 742-day review. 2 later devices descend from it.

Decided
Jul 12, 2024
Review
742 days
Class
Class II
Product code
QVB
Recalls
None

2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2026: 2 descendants cleared2020202220242026
  1. This deviceDEN220042 Terragene Bionova Photon Biological Indicator (BT225); Terragene…Terragene S.A. · 2024
  2. Descendant · cites it directly · 0 of its ownK251559 InstaTRU Instant Biological Indicator for Steam (IBI-05)True Indicating, LLC · 2026
  3. Descendant · cites it directly · 0 of its ownK252306 Verrix EVA™ STEAM Biological Indicator (BI) (Model: 1200), Verrix…Steritec Products · 2026

Most influential descendants

Showing 2 of 2 · See all
  • K251559InstaTRU Instant Biological Indicator for Steam (IBI-05)True Indicating, LLC · 2026 · cites it directly0descendants
  • K252306Verrix EVA™ STEAM Biological Indicator (BI) (Model: 1200), Verrix EVA™ STEAM Process…Steritec Products · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)DEN220042Terragene Bionova Photon Biological Indicator (BT225); Terragene Bionova…Terragene S.A.2024—
+1K251559InstaTRU Instant Biological Indicator for Steam (IBI-05)True Indicating, LLC2026—
+1K252306Verrix EVA™ STEAM Biological Indicator (BI) (Model: 1200), Verrix EVA™ STEAM…Steritec Products2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202620
FIG. 1 — Lineage of DEN220042, 2024–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA De Novo database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QVB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QVB device

  1. #1 K251559 InstaTRU Instant Biological Indicator for…score 11.5 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

Does it have a predicate?

De Novo decisions do not claim a predicate device.

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251559InstaTRU Instant Biological Indicator for Steam (IBI-05)True Indicating, LLC20260
K252306Verrix EVA™ STEAM Biological Indicator (BI) (Model: 1200), Verrix EVA™ STEAM Process…Steritec Products20260

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K251559 (True Indicating, LLC, 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All QVB devices

QVB Biological Sterilization Indicator With Indirect Growth Detection: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · QVB

2so far in 2026, the most since 2007

2024: 12026 (partial year): 2
2007high 2 (2026) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QVB

Too few QVB 510(k)s in any year since 2007 to give a median.

Recalls per year · QVB

No recall names QVB since 2007.

Adverse event reports per year · QVB

No MAUDE report names QVB since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20190—00
20200—00
20210—00
20220—00
20230—00
20241—00
20250—00
20262—0—

What is it indicated for?

Terragene® Bionova® Photon Biological Indicator (BT225) is a 7-second readout Self-Contained Biological Indicator (SCBI) inoculated with a minimum of 106 viable Geobacillus stearothermophilus bacterial spores and is intended for routine monitoring of the efficacy of steam sterilization processes. BT225 SCBI is not recommended and should not be used for qualification testing purposes. On each Terragene® Bionova® SCBI is a chemical process indicator that…

Read the full statement (132 words)

Terragene® Bionova® Photon Biological Indicator (BT225) is a 7-second readout Self-Contained Biological Indicator (SCBI) inoculated with a minimum of 106 viable Geobacillus stearothermophilus bacterial spores and is intended for routine monitoring of the efficacy of steam sterilization processes. BT225 SCBI is not recommended and should not be used for qualification testing purposes. On each Terragene® Bionova® SCBI is a chemical process indicator that changes color from pink to brown when exposed to steam.

- Gravity-displacement Steam Sterilization Cycles

132 ºC, 25 minutes

132 ºC, 15 minutes

132 ºC, 10 minutes

135 ºC, 10 minutes

- Dynamic-air-removal Steam Sterilization Cycles

132 ºC, 4 minutes

135 ºC, 3 minutes

Terragene® Bionova® Photon Auto-reader Incubator (BPH) incubates at 60 °C and reads the Terragene® Bionova® Photon SCBIs at the times prescribed in the User Manual.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “DEN220042: Terragene Bionova Photon Biological Indicator (BT225); Terragene Bionova Photon Auto-Reader Incubator (BPH).” Lineage record from FDA De Novo data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/DEN220042

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.