510(k) Report

510(k) · Cleared

Recalled · Class I · 2014

Espirt

K001208510(k) SpecialCleared May 12, 200028-day reviewClass II21 CFR 868.5895Recalls 8 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on May 12, 2000 after a 28-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 of the 2 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 8 times, including a Class I recall.

Decided
May 12, 2000
Review
28 days
Class
Class II
Product code
CBK
Recalls
8 (1 Class I)
Predicates
0

1 later device descends from it, 1 citing it directly. 1 of the 2 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared200020052010201520202025
  1. This deviceK001208 EspirtRespironics, Inc. · 2000
  2. Descendant · cites it directly · 0 of its ownK041412 Esprit Ventilator with Nico-Esprit Interface Option, Model V1000Respironics California, Inc. · 2004 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K041412Esprit Ventilator with Nico-Esprit Interface Option, Model V1000Respironics California, Inc. · 2004 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K001208EspirtRespironics, Inc.2000—
+1K041412Esprit Ventilator with Nico-Esprit Interface Option, Model V1000Respironics California, Inc.20046

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
FIG. 1 — Lineage of K001208, 2000–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041412Esprit Ventilator with Nico-Esprit Interface Option, Model V1000Respironics California, Inc.20040

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2014Class IIZ-0485-2015Device DesignIf the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if…
Sep 17, 2014Class IZ-0008-2015Unknown/Undetermined by firmA failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not…
Apr 22, 2009Not classifiedZ-1813-2009Device DesignThe recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is erroneously diverted…
Jan 25, 2008Not classifiedZ-2002-2008Device DesignThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize…
Jan 25, 2008Not classifiedZ-2003-2008Device DesignThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize…
Jan 25, 2008Not classifiedZ-2004-2008Device DesignThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize…
Oct 29, 2007Class IIZ-2084-2013Device DesignThe power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
Feb 20, 2002Not classifiedZ-1230-04OtherFlow sensors would exhibit drift.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K041412 (Respironics California, Inc., 2004) has gone furthest, with 0 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—
Cite this page

510(k) Report. “K001208: Espirt.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K001208

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026