510(k) · Cleared
Recalled · Class II · 2020Improved Try-in Gel
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Sybron Dental Specialties, Inc. · EBF Material, Tooth Shade, Resin
Cleared on Aug 28, 2000 after a 67-day review. FDA has not posted a summary or statement for it, so its predicates are not known. It has been recalled 2 times.
- Decided
- Aug 28, 2000
- Review
- 67 days
- Class
- Class II
- Product code
- EBF
- Recalls
- 2
- Predicates
- 0
FDA has not posted a summary or statement for it, so its predicates are not known.
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in EBF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new EBF device
- #1 K202846 Tera Harzscore 18.5 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 25, 2020 | Class II | Z-1824-2020 | Process control | An issue within production process used to package trial kits let to the inclusion of the incorrect bonding agent for the kit. Use of the incorrect bonding agent may lead to increased risk of bond failures. |
| Sep 13, 2013 | Class II | Z-2259-2013 | Mixed-up of materials/components | Kerr Corporation is voluntarily recalling one lot of NX3 Try-In Gel, because some of the NX3 Try-In Gel syringes in the affected lot contain a different product material. The material in the affected syringe does not match the shade of… |
No ancestor or descendant has a Class I recall.
How does its category compare?
All EBF devicesEBF Material, Tooth Shade, Resin: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · EBF
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · EBF
43in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · EBF
1in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · EBF
1in 2025
MAUDE reports naming product code EBF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 19 | 49 | 1 | 6 |
| 2008 | 14 | 79 | 42 | 10 |
| 2009 | 20 | 84 | 1 | 8 |
| 2010 | 19 | 75 | 6 | 31 |
| 2011 | 19 | 81 | 7 | 31 |
| 2012 | 11 | 86 | 0 | 62 |
| 2013 | 13 | 153 | 0 | 246 |
| 2014 | 10 | 112 | 0 | 195 |
| 2015 | 17 | 123 | 0 | 119 |
| 2016 | 20 | 205 | 0 | 143 |
| 2017 | 15 | 108 | 4 | 42 |
| 2018 | 15 | 115 | 0 | 8 |
| 2019 | 17 | 159 | 2 | 11 |
| 2020 | 15 | 190 | 1 | 2 |
| 2021 | 13 | 178 | 0 | 2 |
| 2022 | 17 | 63 | 0 | 2 |
| 2023 | 9 | 87 | 0 | 8 |
| 2024 | 18 | 120 | 0 | 5 |
| 2025 | 15 | 43 | 0 | 1 |
| 2026 | 16 | 59 | 2 | — |
https://510kreport.com/device/K001899 · printed Nov 5, 2026
Cite this page
510(k) Report. “K001899: Improved Try-in Gel.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K001899
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Page last updated Nov 3, 2026