# K002082: Speedlink Transverse Connector

- Pathway: 510(k) (Traditional)
- Applicant: Spinal Concepts, Inc.
- Decided: Oct 6, 2000 after 88 days
- Product code: KWP (Appliance, Fixation, Spinal Interlaminal)
- Device class: 2
- Regulation: 21 CFR 888.3050
- Recalls of this device: 8 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Recalls of this device
- Sep 21, 2009 · Not classified · Z-0525-2010: The titanium stock used to manufacturer the product is reported to exhibit areas of alloy segregation, which could result in a weakening of the strength of the metal.
- Jan 1, 2009 · Not classified · Z-0028-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th
- Jan 1, 2009 · Not classified · Z-0030-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th
- Jan 1, 2009 · Not classified · Z-0032-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th
- Jan 1, 2009 · Not classified · Z-0033-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th
- Jan 1, 2009 · Not classified · Z-0026-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th
- Jan 1, 2009 · Not classified · Z-0027-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th
- Jan 1, 2009 · Not classified · Z-0029-2011: The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. Th