510(k) Report

510(k) · Cleared

Recalled · Class I · 2020

Pro-Femur R

K003016510(k) AbbreviatedCleared Dec 13, 200077-day reviewClass II21 CFR 888.3360Recalls 9 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Wrightmedicaltechnologyinc · LWJ Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented

Cleared on Dec 13, 2000 after a 77-day review. The FDA summary names no predicate by number. 1 of the 6 devices in its family, counting itself, has a Class I recall. 5 later devices descend from it. It has been recalled 9 times, including 2 Class I recalls.

Decided
Dec 13, 2000
Review
77 days
Class
Class II
Product code
LWJ
Recalls
9 (2 Class I)
Predicates
0

5 later devices descend from it, 3 citing it directly. 1 of the 6 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2018: 1 descendant cleared2004: 1 descendant cleared2026: 1 descendant cleared2013: 1 descendant cleared2024: 1 descendant cleared200020052010201520202025
  1. This deviceK003016 Pro-Femur RWrightmedicaltechnologyinc · 2000
  2. Descendant · cites it directly · 1 of its ownK173898 MicroPort Orthopedics Total Hip Systems MR LabelingMicroport Orthopedics Inc. (Mpo) · 2018
  3. Descendant · cites it directly · 1 of its ownK041850 Medallion Modular Hip SystemBiomet Manufacturing, Inc. · 2004
  4. Descendant · cites it directly · 0 of its ownK260173 M-Vizion Femoral Revision System ExtensionMedacta International S.A. · 2026
  5. Descendant · 2 generations on · 0 of its ownK240452 Ceramic Femoral HeadMicroport Orthopedics, Inc. · 2024
  6. Descendant · 2 generations on · 0 of its ownK130149 U2 Hip System, Expanded Indications for UseUnited Orthopedic Corporation · 2013

Most influential descendants

Showing 5 of 5 · See all
  • K173898MicroPort Orthopedics Total Hip Systems MR LabelingMicroport Orthopedics Inc. (Mpo) · 2018 · cites it directly1descendant
  • K041850Medallion Modular Hip SystemBiomet Manufacturing, Inc. · 2004 · cites it directly1descendant
  • K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A. · 2026 · cites it directly0descendants
  • K240452Ceramic Femoral HeadMicroport Orthopedics, Inc. · 2024 · 2 generations on0descendants
  • K130149U2 Hip System, Expanded Indications for UseUnited Orthopedic Corporation · 2013 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K003016Pro-Femur RWrightmedicaltechnologyinc2000—
+1K173898MicroPort Orthopedics Total Hip Systems MR LabelingMicroport Orthopedics Inc. (Mpo)2018—
+1K041850Medallion Modular Hip SystemBiomet Manufacturing, Inc.2004—
+1K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A.2026—
+2K240452Ceramic Femoral HeadMicroport Orthopedics, Inc.2024—
+2K130149U2 Hip System, Expanded Indications for UseUnited Orthopedic Corporation2013—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
201310
201810
202410
202610
FIG. 1 — Lineage of K003016, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173898MicroPort Orthopedics Total Hip Systems MR LabelingMicroport Orthopedics Inc. (Mpo)20181
K041850Medallion Modular Hip SystemBiomet Manufacturing, Inc.20041
K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A.20260

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 31, 2020Class IIZ-2948-2020Labeling Change ControlMicroPort Orthopedics Inc. is voluntarily recalling any existing inventory of PROFEMUR¿ Titanium and Cobalt Chrome modular necks with previous package insert versions to replace the PROFEMUR¿ Hip System Package Insert with the most recent…
Jul 31, 2020Class IZ-2941-2020Device DesignMicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck…
Jul 31, 2020Class IZ-2942-2020Device DesignMicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck…
Jun 30, 2016Class IIZ-2588-2016Error in labelingProfemur(R) neck, Item Number PHA01222 (Short AR VV2 Ti Neck) was labeled as Item Number PHA01212 (Short AR VV1 Ti Neck).
Sep 5, 2012Class IIZ-0183-2013Packaging process controlThere may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.
Sep 5, 2012Class IIZ-0184-2013Packaging process controlThere may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.
Sep 5, 2012Class IIZ-0185-2013Packaging process controlThere may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.
Mar 27, 2007Not classifiedZ-0868-2007OtherThe titanium plasma coating was found to have missing fragments.
Mar 27, 2007Not classifiedZ-0867-2007OtherThe titanium plasma coating was found to have missing fragments.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K173898 (Microport Orthopedics Inc. (Mpo), 2018) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LWJ devices

LWJ Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LWJ

1in 2012

2007: 22008: 12009: 12011: 12012: 1
2007high 2 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWJ

Too few LWJ 510(k)s in any year since 2007 to give a median.

Recalls per year · LWJ

2in 2023

2007: 102008: 22009: 22013: 12023: 2
2007high 10 (2007) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LWJ

51in 2025

2007: 682008: 722009: 842010: 1322011: 2552012: 3902013: 5222014: 4522015: 3102016: 2932017: 3112018: 2272019: 2102020: 1892021: 1552022: 912023: 722024: 542025: 51
2007high 522 (2013) · low 51 (2025)2025

MAUDE reports naming product code LWJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—1068
20081—272
20091—284
20100—0132
20111—0255
20121—0390
20130—1522
20140—0452
20150—0310
20160—0293
20170—0311
20180—0227
20190—0210
20200—0189
20210—0155
20220—091
20230—272
20240—054
20250—051
20260—0—

What is it indicated for?

Indicated for use in total hip arthroplasty for reduction or relief of pain and for improved hip function in skeletally mature patients with the following conditions:

1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;

2) inflammatory degenerative joint disease such as rheumatoid arthritis;

3) correction of fimctional deformity;

4) revision procedures where other treatments or devices have failed; and,

5) treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K003016: Pro-Femur R.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K003016

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026