Recalled · 2010Heartstream FR2 Aed, Model M3860A,M3861A
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Agilent Technologies, Inc. · MKJ Automated External Defibrillators (Non-Wearable)
Cleared on Dec 20, 2000 after a 30-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 of the 11 devices in its family, counting itself, has a Class I recall. 10 later devices descend from it. It has been recalled 12 times.
- Decided
- Dec 20, 2000
- Review
- 30 days
- Class
- Class III
- Product code
- MKJ
- Recalls
- 12
- Predicates
- 0
10 later devices descend from it, 4 citing it directly. 1 of the 11 devices in this family, counting this one, has a Class I recall.
- This deviceK003565 Heartstream FR2 Aed, Model M3860A,M3861AAgilent Technologies, Inc. · 2000
- Descendant · cites it directly · 3 of its ownK023854 Samaritan Aed, Model Sam-01Heartsine Technologies, Inc. · 2003 · Recalled
- Descendant · cites it directly · 2 of its ownK111693 Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems · 2011 · Recalled
- Descendant · cites it directly · 2 of its ownK040904 Philips Heartstart Home Defibrillator, Model M5068APhilips Medical Systems · 2004 · Recalled
- Descendant · 2 generations on · 2 of its ownK041067 Heartsine Samaritan PadHeartsine Technologies, Inc. · 2004 · Recalled (Class I)
- Descendant · 3 generations on · 1 of its ownK123881 Heartsine Samaritan Pad 350PHeartsine Technologies, Inc. · 2013 · Recalled
Showing 5 of 10 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K041067 Heartsine Samaritan Pad2 generations on · 2004
Most influential descendants
Showing 5 of 10 · See all- K023854Samaritan Aed, Model Sam-01Heartsine Technologies, Inc. · 2003 · cites it directly3descendants
- K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems · 2011 · cites it directly2descendants
- K040904Philips Heartstart Home Defibrillator, Model M5068APhilips Medical Systems · 2004 · cites it directly2descendants
- K041067Heartsine Samaritan PadHeartsine Technologies, Inc. · 2004 · 2 generations on2descendants
- K123881Heartsine Samaritan Pad 350PHeartsine Technologies, Inc. · 2013 · 3 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K003565 | Heartstream FR2 Aed, Model M3860A,M3861A | Agilent Technologies, Inc. | 2000 | — |
| +1 | K023854 | Samaritan Aed, Model Sam-01 | Heartsine Technologies, Inc. | 2003 | 1 |
| +1 | K111693 | Philips Heartstart FR3 Automated External Defibrillator | Philips Medical Systems | 2011 | 1 |
| +1 | K040904 | Philips Heartstart Home Defibrillator, Model M5068A | Philips Medical Systems | 2004 | 7 |
| +2 | K041067 | Heartsine Samaritan Pad | Heartsine Technologies, Inc. | 2004 | Class I |
| +3 | K123881 | Heartsine Samaritan Pad 350P | Heartsine Technologies, Inc. | 2013 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
| 2004 | 2 | 1 |
| 2009 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 2 | 0 |
| 2015 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (10)
4 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 22, 2010 | Not classified | Z-0487-2011 | Nonconforming Material/Component | A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture. |
| Sep 22, 2010 | Not classified | Z-0483-2011 | Nonconforming Material/Component | A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture. |
| Sep 22, 2010 | Not classified | Z-0485-2011 | Nonconforming Material/Component | A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture. |
| Sep 22, 2010 | Not classified | Z-0486-2011 | Nonconforming Material/Component | A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture. |
| Sep 22, 2010 | Not classified | Z-0488-2011 | Nonconforming Material/Component | A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture. |
| Sep 22, 2010 | Not classified | Z-0484-2011 | Nonconforming Material/Component | A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture. |
| May 28, 2010 | Not classified | Z-0470-2011 | Nonconforming Material/Component | Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable. |
| May 28, 2010 | Not classified | Z-0472-2011 | Nonconforming Material/Component | Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable. |
| May 28, 2010 | Not classified | Z-0473-2011 | Nonconforming Material/Component | Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable. |
| May 28, 2010 | Not classified | Z-0471-2011 | Nonconforming Material/Component | Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable. |
| Feb 28, 2007 | Not classified | Z-0641-2007 | Nonconforming Material/Component | Failure of tactile switches on one of two printed circuit board assemblies (PCBA) used in the FR2+ device. Potential delay or prevention of shock treatment due to PCBA board contamination, causing intermittent switch operation. |
| Dec 19, 2006 | Not classified | Z-0308-2007 | Process design | Potential for shock treatment to be delayed or prevented due to board contamination that causes intermittent switch operation |
The 5 latest of 12 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K041067Heartsine Samaritan PadHeartsine Technologies, Inc. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- Of its descendants, K023854 (Heartsine Technologies, Inc., 2003) has gone furthest, with 3 descendants of its own.
How does its category compare?
All MKJ devicesMKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MKJ
48in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MKJ
156in 2015
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MKJ
1in 2025
Class I (23 since 2007) Class II/III not classified
Adverse event reports per year · MKJ
13,315in 2025
MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 69 | 12 | 3,958 |
| 2008 | 13 | 141 | 24 | 4,752 |
| 2009 | 8 | 165 | 37 | 6,782 |
| 2010 | 4 | 250 | 35 | 7,399 |
| 2011 | 11 | 36 | 7 | 8,170 |
| 2012 | 10 | 197 | 5 (1) | 9,732 |
| 2013 | 7 | 191 | 4 | 14,955 |
| 2014 | 11 | 196 | 24 (15) | 15,399 |
| 2015 | 6 | 156 | 1 | 16,129 |
| 2016 | 0 | — | 2 (1) | 17,062 |
| 2017 | 4 | — | 6 (4) | 16,684 |
| 2018 | 20 | — | 3 | 16,591 |
| 2019 | 25 | — | 6 (2) | 16,861 |
| 2020 | 52 | — | 2 | 16,197 |
| 2021 | 41 | — | 4 | 16,300 |
| 2022 | 19 | — | 3 | 14,867 |
| 2023 | 24 | — | 11 | 13,766 |
| 2024 | 27 | — | 5 | 15,109 |
| 2025 | 48 | — | 1 | 13,315 |
| 2026 | 31 | — | 3 | — |
https://510kreport.com/device/K003565 · printed Nov 4, 2026
Cite this page
510(k) Report. “K003565: Heartstream FR2 Aed, Model M3860A,M3861A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K003565
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026