Oracle Oral Mask

K003894510(k) TraditionalCleared Jun 25, 2001189-day reviewClass II21 CFR 868.5905Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fisher & Paykel Healthcare, Ltd. · BZD Ventilator, Non-Continuous (Respirator)

Cleared on Jun 25, 2001 after a 189-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 41 later devices descend from it. It has been recalled 6 times.

Decided
Jun 25, 2001
Review
189 days
Class
Class II
Product code
BZD
Recalls
6
Predicates
0

41 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2004: 2 descendants cleared2006: 4 descendants cleared2005: 1 descendant cleared2013: 3 descendants cleared2011: 3 descendants cleared2008: 2 descendants cleared2009: 2 descendants cleared2017: 2 descendants cleared2012: 5 descendants cleared2019: 1 descendant cleared2018: 1 descendant cleared2014: 1 descendant cleared2020: 3 descendants cleared2022: 1 descendant cleared2016: 1 descendant cleared2021: 1 descendant cleared2023: 2 descendants cleared2024: 2 descendants cleared2025: 2 descendants cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK003894 Oracle Oral MaskFisher & Paykel Healthcare, Ltd. · 2001
  2. Descendant · cites it directly · 40 of its ownK023559 Fisher & Paykel Healthecare Oracle Oral MaskFisher & Paykel Healthcare, Ltd. · 2003
  3. Descendant · 2 generations on · 39 of its ownK033087 Oracle Oral Mask, Model HC451AFisher & Paykel Healthcare, Ltd. · 2004 · Recalled
  4. Descendant · 3 generations on · 38 of its ownK040506 Flexifil Series HC431 Full Face Mask, Model HC431AFisher & Paykel Healthcare, Ltd. · 2004 · Recalled
  5. Descendant · 3 generations on · 25 of its ownK061236 Flexifit HC432 Full Face MaskFisher & Paykel Healthcare, Ltd. · 2006
  6. Descendant · 4 generations on · 24 of its ownK063036 Opus HC482 Direct Nasal MaskFisher & Paykel Healthcare, Ltd. · 2006

Showing 5 of 41 descendants. See all descendants

Descendants cleared per year

2003: 12004: 22005: 12006: 42008: 22009: 22011: 32012: 52013: 32014: 12016: 12017: 22018: 12019: 12020: 32021: 12022: 12023: 22024: 22025: 22026: 1
2003high 5 (2012) · low 0 (2007)2026

Most influential descendants

Showing 5 of 41 · See all
  • K023559Fisher & Paykel Healthecare Oracle Oral MaskFisher & Paykel Healthcare, Ltd. · 2003 · cites it directly40descendants
  • K033087Oracle Oral Mask, Model HC451AFisher & Paykel Healthcare, Ltd. · 2004 · 2 generations on39descendants
  • K040506Flexifil Series HC431 Full Face Mask, Model HC431AFisher & Paykel Healthcare, Ltd. · 2004 · 3 generations on38descendants
  • K061236Flexifit HC432 Full Face MaskFisher & Paykel Healthcare, Ltd. · 2006 · 3 generations on25descendants
  • K063036Opus HC482 Direct Nasal MaskFisher & Paykel Healthcare, Ltd. · 2006 · 4 generations on24descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K003894Oracle Oral MaskFisher & Paykel Healthcare, Ltd.2001—
+1K023559Fisher & Paykel Healthecare Oracle Oral MaskFisher & Paykel Healthcare, Ltd.2003—
+2K033087Oracle Oral Mask, Model HC451AFisher & Paykel Healthcare, Ltd.20046
+3K040506Flexifil Series HC431 Full Face Mask, Model HC431AFisher & Paykel Healthcare, Ltd.20046
+3K061236Flexifit HC432 Full Face MaskFisher & Paykel Healthcare, Ltd.2006—
+4K063036Opus HC482 Direct Nasal MaskFisher & Paykel Healthcare, Ltd.2006—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200420
200510
200640
200820
200920
201130
201250
201330
201410
201610
201720
201810
201910
202030
202110
202210
202320
202420
202520
202610
FIG. 1 — Lineage of K003894, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BZD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZD device

  1. #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (41)

See all 41

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K023559Fisher & Paykel Healthecare Oracle Oral MaskFisher & Paykel Healthcare, Ltd.200340

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 24, 2006Not classifiedZ-0974-2007OtherPlastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and…
Nov 24, 2006Not classifiedZ-0970-2007OtherPlastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and…
Nov 24, 2006Not classifiedZ-0971-2007OtherPlastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and…
Nov 24, 2006Not classifiedZ-0973-2007OtherPlastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and…
Nov 24, 2006Not classifiedZ-0969-2007OtherPlastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and…
Nov 24, 2006Not classifiedZ-0972-2007OtherPlastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K023559 (Fisher & Paykel Healthcare, Ltd., 2003) has gone furthest, with 40 descendants of its own.

How does its category compare?

All BZD devices

BZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BZD

7in 2025

2007: 212008: 172009: 232010: 122011: 232012: 212013: 232014: 142015: 112016: 122017: 52018: 52019: 62020: 62021: 62022: 42023: 52024: 102025: 72026 (partial year): 10
2007high 23 (2009) · low 4 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZD

189in 2025

2007: 83 days median2008: 108 days median2009: 90 days median2010: 122 days median2011: 125 days median2012: 133 days median2013: 127 days median2014: 172 days median2015: 224 days median2016: 217 days median2017: 221 days median2018: 280 days median2019: 231 days median2020: 137 days median2021: 192 days median2022: 253 days median2023: 204 days median2024: 132 days median2025: 189 days median2026 (partial year): 214 days median
2007high 280 (2018) · low 83 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZD

8in 2023

2007: 292008: 22009: 12011: 12019: 4, 4 Class I2023: 8, 7 Class I2026 (partial year): 2, 2 Class I
2007high 29 (2007) · low 0 (2010)2026

Class I (13 since 2007) Class II/III not classified

Adverse event reports per year · BZD

60,279in 2025

2007: 652008: 672009: 1472010: 1432011: 1892012: 1392013: 822014: 862015: 1142016: 1182017: 1482018: 1012019: 1882020: 1262021: 6,0662022: 90,7682023: 27,9872024: 68,2622025: 60,279
2007high 90,768 (2022) · low 65 (2007)2025

MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721832965
200817108267
200923901147
2010121220143
2011231251189
2012211330139
201323127082
201414172086
2015112240114
2016122170118
201752210148
201852800101
201962314 (4)188
202061370126
2021619206,066
20224253090,768
202352048 (7)27,987
202410132068,262
20257189060,279
2026102142 (2)—
Cite this page

510(k) Report. “K003894: Oracle Oral Mask.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K003894

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026