# K011296: Infinion Enhancement Package

- Pathway: 510(k) (Traditional)
- Applicant: Philips Medical Systems (Cleveland), Inc.
- Decided: Jul 19, 2001 after 80 days
- Product code: LNH (System, Nuclear Magnetic Resonance Imaging)
- Device class: 2
- Regulation: 21 CFR 892.1000
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Recalls of this device
- Sep 19, 2008 · Not classified · Z-2432-2010: Water may collect and freeze in the vent pipe. This can rupture the magnet vessel during a cryogen transfer, and release helium into the MRI suite.
- Sep 19, 2008 · Not classified · Z-2433-2010: Water may collect and freeze in the vent pipe. This can rupture the magnet vessel during a cryogen transfer, and release helium into the MRI suite.
- Apr 17, 2003 · Not classified · Z-0272-05: There is a potential for the patient to be exposed to cryogenic fluids (liquid air) dripping from the vent during a quench due to the location of the vent.