# K011314: Olympus GF UC140P-AL5 Ultrasonic Endoscope with Aloka SSD-5500 Ultrasound System

- Pathway: 510(k) (Traditional)
- Applicant: Aloka Co., Ltd.
- Decided: May 15, 2001 after 15 days
- Product code: ITX (Transducer, Ultrasonic, Diagnostic)
- Device class: 2
- Regulation: 21 CFR 892.1570
- Recalls of this device: 4 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
73 later devices descend from it; 4 cite it directly.
1 of 74 devices in its family (counting itself) has a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Most influential descendants
- K031347: Ultrasonic Gastrovideoscope, Type UC160P-AT8 and UCT160-AT8 (2003), 44 descendants of its own
- K042140: Olympus BF Type UC160-OL8 (2004), 43 descendants of its own
- K050503: Single Use Aspiration Needle NA-201SX-4022 (2005), 28 descendants of its own
- K051541: GF-UE160-AL5 Ultrasound Endoscope (2005), 27 descendants of its own
- K063683: Olympus Xeu-M60A Endoscopic Ultrasound Center; XGF-UM180 Ultrasonic Gastrovideoscope (2007), 15 descendants of its own

## Recalls of this device
- Sep 25, 2023 · Class II · Z-0333-2024: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the poten
- Mar 18, 2022 · Class II · Z-1034-2022: GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients
- Oct 16, 2020 · Class II · Z-0498-2021: A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- Oct 16, 2020 · Class II · Z-0499-2021: A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.