510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Multi-Snare Set

K011783510(k) TraditionalCleared Jan 24, 2002231-day reviewClass II21 CFR 870.5150Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pfm Produkte Fur Die Medizin AG · DXE Catheter, Embolectomy

Cleared on Jan 24, 2002 as substantially equivalent to K925439 (Microvena Corp., 1993) and 1 other predicate after a 231-day review. Its family tree reaches back 1 generation to 1993. It has been recalled once.

Decided
Jan 24, 2002
Review
231 days
Class
Class II
Product code
DXE
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K920823 (1993). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1993: 1 other ancestorK920823 · Welter Snare Retriever · Cook, Inc. · 199319901992199419961998200020022004
Other ancestors, by generation 1 generation back: 1 other ancestor K920823 · Welter Snare Retriever · Cook, Inc. · 19930
  1. Root · 1 generation backK920823 Welter Snare RetrieverCook, Inc. · 1993
  2. This deviceK011783 Multi-Snare SetPfm Produkte Fur Die Medizin AG · 2002
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K920823Welter Snare RetrieverCook, Inc.1993—
0 (this device)K011783Multi-Snare SetPfm Produkte Fur Die Medizin AG2002—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
FIG. 1 — Lineage of K011783, 1993–2002Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXE device

  1. #1 K240078 Fogarty Thru-Lumen Embolectomy Catheterscore 7.9 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K925439Microsnare
Summary page 1
…e Name: Multi-Snare Set Common Name: Embolectomy Catheter Classification Name: Same Equivalent Devices: Manufacturer: Microvena Corporation Name: Microsnare 510(k) #: K925439 Manufacturer: Cook Inc. Name: Welter Snare Retriever 510(k) #: K920823 Device Description: The Multi-Snare Set is designed for…
Microvena Corp.1993—
*K920823Welter Snare Retriever
Summary page 1
…Name: Same Equivalent Devices: Manufacturer: Microvena Corporation Name: Microsnare 510(k) #: K925439 Manufacturer: Cook Inc. Name: Welter Snare Retriever 510(k) #: K920823 Device Description: The Multi-Snare Set is designed for the following indications: e Retrieval and manipulation of foreign objec…
Cook, Inc.1993—

Also cleared against K925439: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 21, 2026Class IIZ-2526-2026Under Investigation by firmA snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • 1 other device also cites its predicate K925439.

How does its category compare?

All DXE devices

DXE Catheter, Embolectomy: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DXE

5in 2024

2008: 12010: 22015: 12017: 22020: 12021: 12022: 12024: 5
2007high 5 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXE

173in 2024

2024: 173 days median
2007high 173 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXE

1in 2023

2007: 32009: 52010: 12011: 12015: 132021: 4, 4 Class I2022: 32023: 12026 (partial year): 2
2007high 13 (2015) · low 0 (2008)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DXE

306in 2025

2007: 1672008: 1342009: 1242010: 1362011: 1642012: 2002013: 1502014: 1922015: 2662016: 7962017: 7702018: 6152019: 5942020: 5472021: 3332022: 4632023: 4362024: 3772025: 306
2007high 796 (2016) · low 124 (2009)2025

MAUDE reports naming product code DXE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—3167
20081—0134
20090—5124
20102—1136
20110—1164
20120—0200
20130—0150
20140—0192
20151—13266
20160—0796
20172—0770
20180—0615
20190—0594
20201—0547
20211—4 (4)333
20221—3463
20230—1436
202451730377
20250—0306
20260—2—

What is it indicated for?

The Multi-Snare Set (and its components) is intended for:

Retrieval and manipulation of foreign objects from

the vascular system and hollow viscera

Assistance in creating loops where cross-over

technique is applied

Repositioning of indwelling venous catheters

Assistance in performing venipuncture to obtain

access to central vein

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K011783: Multi-Snare Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K011783

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026