510(k) Report

510(k) · Cleared

Recalled · 2004

Modification to Accu-Chek Inform Meter

K012210510(k) SpecialCleared Aug 9, 200124-day reviewClass II21 CFR 862.1345Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · NBW System, Test, Blood Glucose, Over The Counter

Cleared on Aug 9, 2001 after a 24-day review. The FDA summary names its predicate without a number. It has been recalled 3 times.

Decided
Aug 9, 2001
Review
24 days
Class
Class II
Product code
NBW
Recalls
3
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K012210Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NBW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NBW device

  1. #1 K260545 CONTOUR® PLUS BLUE Blood Glucose Monitoring…score 14.6 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Roche Diagnostics Accu-Chek Advantage Meter
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 19, 2004Not classifiedZ-0372-05OtherA firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa.
Jun 30, 2004Not classifiedZ-1060-04OtherExposure to excessive disinfectant or cleaning solution may cause overheating, melting, or smoke.
Jun 30, 2004Not classifiedZ-1061-04OtherExposure to excessive disinfectant or cleaning solution may cause overheating, melting, or smoke.

No ancestor or descendant has a Class I recall.

How does its category compare?

All NBW devices

NBW System, Test, Blood Glucose, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NBW

2in 2025

2007: 292008: 372009: 382010: 312011: 302012: 302013: 472014: 242015: 322016: 202017: 252018: 192019: 102020: 72021: 72022: 122023: 52024: 72025: 22026 (partial year): 5
2007high 47 (2013) · low 2 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NBW

243in 2024

2007: 85 days median2008: 128 days median2009: 140 days median2010: 132 days median2011: 328 days median2012: 230 days median2013: 167 days median2014: 202 days median2015: 237 days median2016: 233 days median2017: 244 days median2018: 168 days median2019: 127 days median2020: 121 days median2021: 465 days median2022: 410 days median2023: 193 days median2024: 243 days median2026 (partial year): 190 days median
2007high 465 (2021) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NBW

1in 2025

2007: 92008: 22009: 52010: 112011: 12012: 12013: 5, 2 Class I2014: 12016: 12019: 142025: 1
2007high 14 (2019) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NBW

4,745in 2025

2007: 7,9912008: 7,7042009: 10,9632010: 11,6312011: 22,9092012: 23,4492013: 35,4082014: 38,5242015: 62,1112016: 76,6212017: 37,5382018: 16,9112019: 14,4722020: 10,7032021: 9,1982022: 8,1032023: 7,2562024: 6,2202025: 4,745
2007high 76,621 (2016) · low 4,745 (2025)2025

MAUDE reports naming product code NBW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007298597,991
20083712827,704
200938140510,963
2010311321111,631
201130328122,909
201230230123,449
2013471675 (2)35,408
201424202138,524
201532237062,111
201620233176,621
201725244037,538
201819168016,911
2019101271414,472
20207121010,703
2021746509,198
20221241008,103
2023519307,256
2024724306,220
20252—14,745
202651900—

What is it indicated for?

The Accu-Chek Inform Meter is designed to quantitatively measure the concentration of glucose in whole blood samples. The device is indicated for use by health care professionals and persons with diabetes mellitus.

The Accu-Chek Inform Meter is an addition to the Accu-Chek brand of blood glucose monitors that incorporates the fundamental scientific technology currently found in the Accu-Chek Advantage Meter and Accu-Chek Complete Meter. The Accu-Chek Inform Meter is designed for use in conjunction with either the Accu-Chek Advantage or Accu-Chek Comfort Curve test strips.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K012210: Modification to Accu-Chek Inform Meter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K012210

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Page last updated Nov 3, 2026