Recalled · Class II · 2016Tubing and Connectors with Trillium Biopassive Surface
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medtronic Vascular · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Cleared on Aug 23, 2001 as substantially equivalent to K800178 (Extracorporeal Medical Specialities, Inc., 1980) after a 16-day review. Its family tree reaches back 1 generation to 1980. 14 later devices descend from it. It has been recalled 5 times.
- Decided
- Aug 23, 2001
- Review
- 16 days
- Class
- Class II
- Product code
- DWF
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 1 generation to K800178 (1980). 14 later devices descend from it, 2 citing it directly.
- Root · 1 generation backK800178 Extracorp.'S Intersept Cus. Tubing PK.Extracorporeal Medical Specialities, Inc. · 1980 · Recalled
- This deviceK012538 Tubing and Connectors with Trillium Biopassive SurfaceMedtronic Vascular · 2001
- Descendant · cites it directly · 7 of its ownK113845 Tubing, Conmnectors, and Accessories Wit Balance BiosurfaceMedtronic, Inc. · 2012 · Recalled
- Descendant · 2 generations on · 5 of its ownK122811 Connector Components with Balance BiosurfaceMedtronic, Inc. · 2012 · Recalled
- Descendant · cites it directly · 5 of its ownK111972 Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm…Medtronic, Inc. · 2011
- Descendant · 2 generations on · 4 of its ownK151523 Uncoated Tubing and Connector ComponentsMedtronic, Inc. · 2015
- Descendant · 2 generations on · 4 of its ownK120988 DLP Single Stage Venous Cannula, DLP Right Angle Single Stage Venous…Medtronic, Inc. · 2012 · Recalled
Showing 5 of 14 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 14 · See all- K113845Tubing, Conmnectors, and Accessories Wit Balance BiosurfaceMedtronic, Inc. · 2012 · cites it directly7descendants
- K122811Connector Components with Balance BiosurfaceMedtronic, Inc. · 2012 · 2 generations on5descendants
- K111972Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm Cardioplegia Deliv.,Affinity…Medtronic, Inc. · 2011 · cites it directly5descendants
- K151523Uncoated Tubing and Connector ComponentsMedtronic, Inc. · 2015 · 2 generations on4descendants
- K120988DLP Single Stage Venous Cannula, DLP Right Angle Single Stage Venous Cannula, DLP Single…Medtronic, Inc. · 2012 · 2 generations on4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K800178 | Extracorp.'S Intersept Cus. Tubing PK. | Extracorporeal Medical Specialities, Inc. | 1980 | 10 |
| 0 (this device) | K012538 | Tubing and Connectors with Trillium Biopassive Surface | Medtronic Vascular | 2001 | — |
| +1 | K113845 | Tubing, Conmnectors, and Accessories Wit Balance Biosurface | Medtronic, Inc. | 2012 | 1 |
| +2 | K122811 | Connector Components with Balance Biosurface | Medtronic, Inc. | 2012 | 1 |
| +1 | K111972 | Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm Cardioplegia… | Medtronic, Inc. | 2011 | — |
| +2 | K151523 | Uncoated Tubing and Connector Components | Medtronic, Inc. | 2015 | — |
| +2 | K120988 | DLP Single Stage Venous Cannula, DLP Right Angle Single Stage Venous Cannula… | Medtronic, Inc. | 2012 | 2 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2012 | 4 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2017 | 2 | 0 |
| 2022 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DWF device
- #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K800178 | Extracorp.'S Intersept Cus. Tubing PK.Summary page 1…ger, Regulatory Affairs Device Name Tubing and Connectors with Trillium™ Biopassive Surface Name of Predicated or Legally Marketed Device Intersept Blood Tubing Packs (K800178) Affinity Hollow Fiber Oxygenator with Trillium™ Biopassive Surface (K973760) Description of Device The tubing consists of bioco… | 1980 |
| † | K973760 | Affinity Hollow Fiber Oxygenator with Trillium Biopassive Surface | 1998 |
What descends from it? (14)
See all 142 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 4, 2016 | Class III | Z-1023-2016 | Component design/selection | Medtronic is notifying customers who have or may receive specific lots of Medtronic Perfusion Tubing Packs that include the potentially affected cuvettes manufactured by Terumo Cardiovascular Systems (Terumo CVS). Terumo Cardiovascular… |
| Oct 28, 2015 | Class II | Z-0268-2016 | Packaging process control | This recall is being initiated due to a potential breach of the Tubing Pack Header Bags manufactured in these select lots. The seal in question is between the Tyvek and the Poly side of the header bag. |
| Aug 18, 2015 | Class II | Z-2736-2015 | Nonconforming Material/Component | Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic… |
| Jul 17, 2013 | Class II | Z-2106-2013 | Process control | Medtronic was notified that Edwards Lifesciences has initiated a product recall for the Rigid Suction Wand, model S099B. The Wands are included within certain Perfusion Tubing Packs that are manufactured and distributed by Medtronic. |
| Nov 19, 2012 | Class II | Z-0618-2013 | Package design/selection | Medtronic is initiating an Urgent Medical Device Customer Notification. We have confirmed that certain Medtronic Intersept Tubing Packs contain packaging trays that are susceptible to damage that can compromise product sterility. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DWF devicesDWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DWF
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DWF
89in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DWF
4in 2024
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DWF
1,020in 2025
MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 25 | 121 |
| 2008 | 6 | 95 | 25 | 126 |
| 2009 | 8 | 148 | 21 | 99 |
| 2010 | 3 | 62 | 26 | 197 |
| 2011 | 5 | 85 | 13 | 212 |
| 2012 | 10 | 110 | 37 | 233 |
| 2013 | 9 | 110 | 14 (1) | 238 |
| 2014 | 14 | 110 | 6 | 325 |
| 2015 | 9 | 46 | 3 (1) | 218 |
| 2016 | 3 | 32 | 3 | 144 |
| 2017 | 5 | 127 | 2 (1) | 101 |
| 2018 | 5 | 30 | 10 | 118 |
| 2019 | 3 | 120 | 2 | 179 |
| 2020 | 3 | 30 | 0 | 142 |
| 2021 | 5 | 358 | 3 | 162 |
| 2022 | 2 | — | 0 | 276 |
| 2023 | 2 | — | 0 | 470 |
| 2024 | 3 | 29 | 4 | 401 |
| 2025 | 4 | 89 | 0 | 1,020 |
| 2026 | 8 | 81 | 20 | — |
What is it indicated for?
This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass surgery.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K012538 · printed Nov 6, 2026
Cite this page
510(k) Report. “K012538: Tubing and Connectors with Trillium Biopassive Surface.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K012538
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026