IM2001 Pulse Oximeter

K012706510(k) TraditionalCleared Mar 7, 2003570-day reviewClass II21 CFR 870.2700

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Imagyn Medical Technologies, Inc. · DQA Oximeter

Cleared on Mar 7, 2003 as substantially equivalent to K991823 (Nellcor Puritan Bennett, Inc., 1999) and 1 other predicate after a 570-day review. Its family tree reaches back 4 generations to 1997. 2 later devices descend from it.

Decided
Mar 7, 2003
Review
570 days
Class
Class II
Product code
DQA
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K962603 (1997). 6 ancestors in all, the oldest K955642 (1996). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 19970
  1. Root · 4 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 3 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 2 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 1 generation backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. This deviceK012706 IM2001 Pulse OximeterImagyn Medical Technologies, Inc. · 2003
  6. Descendant · cites it directly · 1 of its ownK032831 Pro 2 Monitor, Pro 2 Sensor, PRO2 Holder, PRO2 Neonate Holder…Conmed Corporation · 2004
  7. Descendant · 2 generations on · 0 of its ownK061112 Conmed PRO2 Pulse Reflectance Oximeter, Model Pro-MConmed Corporation · 2006

Most influential descendants

Showing 2 of 2 · See all
  • K032831Pro 2 Monitor, Pro 2 Sensor, PRO2 Holder, PRO2 Neonate Holder, Models, Pro- 200…Conmed Corporation · 2004 · cites it directly1descendant
  • K061112Conmed PRO2 Pulse Reflectance Oximeter, Model Pro-MConmed Corporation · 2006 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−3K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−2K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−1K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
0 (this device)K012706IM2001 Pulse OximeterImagyn Medical Technologies, Inc.2003—
+1K032831Pro 2 Monitor, Pro 2 Sensor, PRO2 Holder, PRO2 Neonate Holder, Models, Pro-…Conmed Corporation2004—
+2K061112Conmed PRO2 Pulse Reflectance Oximeter, Model Pro-MConmed Corporation2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200610
FIG. 1 — Lineage of K012706, 1997–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K991823N-395 Pulse Oximeter, Model N-395
Summary page 1
…ation No. 870.2700) and they are reviewed by the Cardiovascular Panel. Predicate Device: N-395 Pulse Oximeter (Nellcor Puritan Bennett, Inc) cleared under K991823 and K993637 when operates with its RS-10 sensor (Nellcor Puritan Bennett, Inc) cleared under K904039.…
Nellcor Puritan Bennett, Inc.1999—
*K993637N-395 Pulse Oximeter, with Extended Device Claims
Summary page 1
…0.2700) and they are reviewed by the Cardiovascular Panel. Predicate Device: N-395 Pulse Oximeter (Nellcor Puritan Bennett, Inc) cleared under K991823 and K993637 when operates with its RS-10 sensor (Nellcor Puritan Bennett, Inc) cleared under K904039.…
Nellcor Puritan Bennett, Inc.19991

Also cleared against K991823: 20 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032831Pro 2 Monitor, Pro 2 Sensor, PRO2 Holder, PRO2 Neonate Holder, Models, Pro- 200…Conmed Corporation20041

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 4 companies; a new generation was cleared every 11 months on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K032831 (Conmed Corporation, 2004) has gone furthest, with 1 descendant of its own.
  • 20 other devices also cite its predicate K991823.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

The IM2001 Pulse Oximeter is indicated for the continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate. For use with pediatric and adult population, in hospitals, hospital-type facilities and intra-hospital transport environment. For prescription use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K012706: IM2001 Pulse Oximeter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K012706

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026