510(k) Report

510(k) · Cleared

Recalled · 2003

Modification to: Datex-Ohmeda Cardiocap 5 (Rev B) and Accessories

K012837510(k) SpecialCleared Sep 14, 200122-day reviewClass II21 CFR 870.1025Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda · MLD Monitor, St Segment With Alarm

Cleared on Sep 14, 2001 as substantially equivalent to K992323 (Datex-Ohmeda, Inc., 1999) after a 22-day review. Its family tree reaches back 1 generation to 1999. It has been recalled 2 times.

Decided
Sep 14, 2001
Review
22 days
Class
Class II
Product code
MLD
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K992323 (1999). No later device cites it as a predicate yet.

  1. Root · 1 generation backK992323 Datex-Ohmeda Cardiocap 5, F-MX, F-MXG and AccessoriesDatex-Ohmeda, Inc. · 1999 · Recalled
  2. This deviceK012837 Modification to: Datex-Ohmeda Cardiocap 5 (Rev B) and AccessoriesDatex-Ohmeda · 2001
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K992323Datex-Ohmeda Cardiocap 5, F-MX, F-MXG and AccessoriesDatex-Ohmeda, Inc.19991
0 (this device)K012837Modification to: Datex-Ohmeda Cardiocap 5 (Rev B) and AccessoriesDatex-Ohmeda2001—
FIG. 1 — Lineage of K012837, 1999–2001Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K992323Datex-Ohmeda Cardiocap 5, F-MX, F-MXG and Accessories
Summary page 2
…x-Ohmeda Cardiocap 5 Rev B and accessories is substantially equivalent ) in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda Cardiocap 5 Rev A (K992323). DEVICE DESCRIPTION as required by 807.92(a)(4) The Cardiocap 5 monitor is a multi-parameter monitor that is factory configure…
Datex-Ohmeda, Inc.19991
†K974747CS/3 Arrhythmia WorkstationDatex-Engstrom, Inc.1998—

Also cleared against K992323: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 17, 2003Not classifiedZ-0143-04OtherDatex-Ohmeda Cardiocap/5 Patient Monitors have bracket pins which do not adequately secure them to their optional mounting stand so the monitors may fall.
Sep 16, 2002Not classifiedZ-0611-03OtherFalsely high blood oxygen saturation readings in some circumstances which could result in no alarm in a critical life-threatening situation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 1 other device also cites its predicate K992323.

How does its category compare?

All MLD devices

MLD Monitor, St Segment With Alarm: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MLD

1in 2010

2008: 12010: 1
2007high 1 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLD

Too few MLD 510(k)s in any year since 2007 to give a median.

Recalls per year · MLD

No recall names MLD since 2007.

Adverse event reports per year · MLD

1in 2019

2007: 32008: 52009: 32010: 32011: 62012: 72014: 12015: 12017: 12018: 12019: 1
2007high 7 (2012) · low 0 (2013)2025

MAUDE reports naming product code MLD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—03
20081—05
20090—03
20101—03
20110—06
20120—07
20130—00
20140—01
20150—01
20160—00
20170—01
20180—01
20190—01
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Datex-Ohmeda Cardiocap 5 (REV. B) and accessories are indicated for indoor monitoring of hemodynamic (ECG, Impedance respiration, NIBP, Temperature, SPO2 and invasive pressure) , respiratory (CO2, 02, N20, respiration rate, anesthetic agent and agent identification), ventilatory (airway pressure, volume and flow) and relaxation status (NMT) of all hospital patients. With the N-XOSAT option, monitoring of arterial oxygen saturation includes monitoring…

Read the full statement (102 words)

The Datex-Ohmeda Cardiocap 5 (REV. B) and accessories are indicated for indoor monitoring of hemodynamic (ECG, Impedance respiration, NIBP, Temperature, SPO2 and invasive pressure) , respiratory (CO2, 02, N20, respiration rate, anesthetic agent and agent identification), ventilatory (airway pressure, volume and flow) and relaxation status (NMT) of all hospital patients.

With the N-XOSAT option, monitoring of arterial oxygen saturation includes monitoring hospital patients under clinical motion conditions.

Impedance Respiration measurement is indicated for patients ages 3 years and older. Cardiocap/5 is indicated for patients weighing 5 kg (11 Ib.) or more.

The monitor is indicated for use by qualified medical personnel only.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K012837: Modification to: Datex-Ohmeda Cardiocap 5 (Rev B) and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K012837

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026