510(k) · Cleared
Recalled · Class I · 2023V*Care Manual Resuscitator
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ventlab Corp. · BTM Ventilator, Emergency, Manual (Resuscitator)
Cleared on Sep 18, 2001 after a 26-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 1 of the 2 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 21 times, including 12 Class I recalls.
- Decided
- Sep 18, 2001
- Review
- 26 days
- Class
- Class II
- Product code
- BTM
- Recalls
- 21 (12 Class I)
- Predicates
- 0
1 later device descends from it, 1 citing it directly. 1 of the 2 devices in this family, counting this one, has a Class I recall.
- This deviceK012842 V*Care Manual ResuscitatorVentlab Corp. · 2001
- Descendant · cites it directly · 0 of its ownK053466 Hsiner ResuscitatorHsiner Co., Ltd. · 2006
Most influential descendants
Showing 1 of 1 · See all- K053466Hsiner ResuscitatorHsiner Co., Ltd. · 2006 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K012842 | V*Care Manual Resuscitator | Ventlab Corp. | 2001 | — |
| +1 | K053466 | Hsiner Resuscitator | Hsiner Co., Ltd. | 2006 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new BTM device
- #1 K243861 butterflyBVMscore 9.5 · 2 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K053466 | Hsiner Resuscitator | 2006 | 0 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 11, 2023 | Class II | Z-0365-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0369-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0361-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0366-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0364-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0362-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0363-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0367-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| Oct 11, 2023 | Class II | Z-0368-2024 | Device Design | A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. |
| May 16, 2014 | Class I | Z-1948-2014 | Process change control | The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim. |
| Jul 11, 2012 | Class I | Z-0173-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0175-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0177-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0171-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0174-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0176-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0180-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0170-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0178-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0179-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
| Jul 11, 2012 | Class I | Z-0181-2013 | Component design/selection | Possible volume leakage through the inlet valve during compressions of the resuscitator bag. |
The 5 latest of 21 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K053466 (Hsiner Co., Ltd., 2006) has gone furthest, with 0 descendants of its own.
How does its category compare?
All BTM devicesBTM Ventilator, Emergency, Manual (Resuscitator): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · BTM
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · BTM
83in 2008
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · BTM
9in 2024
Class I (24 since 2007) Class II/III not classified
Adverse event reports per year · BTM
36in 2025
MAUDE reports naming product code BTM, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 65 |
| 2008 | 3 | 83 | 0 | 31 |
| 2009 | 0 | — | 23 | 21 |
| 2010 | 2 | — | 3 | 23 |
| 2011 | 2 | — | 0 | 48 |
| 2012 | 1 | — | 13 (13) | 69 |
| 2013 | 0 | — | 3 (1) | 44 |
| 2014 | 0 | — | 0 | 108 |
| 2015 | 1 | — | 0 | 36 |
| 2016 | 1 | — | 0 | 92 |
| 2017 | 0 | — | 0 | 56 |
| 2018 | 2 | — | 0 | 56 |
| 2019 | 0 | — | 0 | 46 |
| 2020 | 0 | — | 0 | 26 |
| 2021 | 1 | — | 0 | 42 |
| 2022 | 1 | — | 0 | 35 |
| 2023 | 1 | — | 9 | 46 |
| 2024 | 0 | — | 9 (9) | 33 |
| 2025 | 2 | — | 0 | 36 |
| 2026 | 1 | — | 1 (1) | — |
What is it indicated for?
The V*Care Manual Resuscitator with Built-in Pressure Monitor's intended use is to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K012842 · printed Nov 5, 2026
Cite this page
510(k) Report. “K012842: V*Care Manual Resuscitator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K012842
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026