510(k) Report

510(k) · Cleared

Recalled · Class I · 2012

Neptune Waste Management Ststem

K012991510(k) TraditionalCleared Nov 29, 200184-day reviewClass II21 CFR 878.5070Recalls 6 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Instruments · FYD Apparatus, Exhaust, Surgical

Cleared on Nov 29, 2001 after an 84-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 1 of the 8 devices in its family, counting itself, has a Class I recall. 7 later devices descend from it. It has been recalled 6 times, including 4 Class I recalls.

Decided
Nov 29, 2001
Review
84 days
Class
Class II
Product code
FYD
Recalls
6 (4 Class I)
Predicates
0

7 later devices descend from it, 4 citing it directly. 1 of the 8 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2009: 1 descendant cleared2008: 2 descendants cleared2013: 1 descendant cleared2022: 1 descendant cleared2017: 1 descendant cleared2016: 1 descendant cleared200020052010201520202025
  1. This deviceK012991 Neptune Waste Management StstemStryker Instruments · 2001
  2. Descendant · cites it directly · 4 of its ownK081047 Transposal Ultra SystemsDornoch Medical Systems, Inc. · 2008
  3. Descendant · cites it directly · 3 of its ownK132671 Stryker Neptune 2 Waste Managment SystemStryker Corporation · 2013 · Recalled
  4. Descendant · 2 generations on · 1 of its ownK153407 Stryker Neptune 3 Waste Management SystemStryker Corporation · 2016 · Recalled
  5. Descendant · cites it directly · 1 of its ownK080845 Orwell Fluid Collection and Disposal SystemCardinalhealth · 2008
  6. Descendant · 2 generations on · 0 of its ownK222552 120V Neptune S Rover (0711-001-000);V2 4-Port Manifold…Stryker Instruments · 2022

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2008: 22009: 12013: 12016: 12017: 12022: 1
2008high 2 (2008) · low 0 (2010)2026

Most influential descendants

Showing 5 of 7 · See all
  • K081047Transposal Ultra SystemsDornoch Medical Systems, Inc. · 2008 · cites it directly4descendants
  • K132671Stryker Neptune 2 Waste Managment SystemStryker Corporation · 2013 · cites it directly3descendants
  • K153407Stryker Neptune 3 Waste Management SystemStryker Corporation · 2016 · 2 generations on1descendant
  • K080845Orwell Fluid Collection and Disposal SystemCardinalhealth · 2008 · cites it directly1descendant
  • K222552120V Neptune S Rover (0711-001-000);V2 4-Port Manifold (0750-400-000);V2 Specimen…Stryker Instruments · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K012991Neptune Waste Management StstemStryker Instruments2001—
+1K081047Transposal Ultra SystemsDornoch Medical Systems, Inc.2008—
+1K132671Stryker Neptune 2 Waste Managment SystemStryker Corporation20135
+2K153407Stryker Neptune 3 Waste Management SystemStryker Corporation20161
+1K080845Orwell Fluid Collection and Disposal SystemCardinalhealth2008—
+2K222552120V Neptune S Rover (0711-001-000);V2 4-Port Manifold (0750-400-000);V2…Stryker Instruments2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200820
200910
201310
201610
201710
202210
FIG. 1 — Lineage of K012991, 2001–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FYD device

  1. #1 K240127 PlumeSafe X5 Smoke Management Systemscore 8.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (7)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081047Transposal Ultra SystemsDornoch Medical Systems, Inc.20084
K132671Stryker Neptune 2 Waste Managment SystemStryker Corporation20133
K080845Orwell Fluid Collection and Disposal SystemCardinalhealth20081
K090759Biodrain Streamway Fluid Management SystemBiodrain Medical, Inc.20090

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 8, 2012Class IZ-2063-2012Labeling False and MisleadingStryker has received two reports of serious injury as a result of tissue damage associated with the use of the Neptune 2, including an event in which one customer connected the Neptune 2 System to a passive chest drainage tube post…
Jun 8, 2012Class IZ-2062-2012Labeling False and MisleadingStryker has received two reports of serious injury as a result of tissue damage associated with the use of the Neptune 2, including an event in which one customer connected the Neptune 2 System to a passive chest drainage tube post…
Jun 8, 2012Class IZ-2061-2012Labeling False and MisleadingStryker has received two reports of serious injury as a result of tissue damage associated with the use of the Neptune 2, including an event in which one customer connected the Neptune 2 System to a passive chest drainage tube post…
Jun 8, 2012Class IZ-2064-2012Labeling False and MisleadingStryker has received two reports of serious injury as a result of tissue damage associated with the use of the Neptune 2, including an event in which one customer connected the Neptune 2 System to a passive chest drainage tube post…
Oct 12, 2004Not classifiedZ-0174-05OtherThe grounding pin may break inside of the electrical plug, resulting in a shock hazard.
Dec 3, 2002Not classifiedZ-0524-03OtherCap may be cracked and fail, exposing O.R. staff to patient blood, and cause an unexpected loud noise in the O.R.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K081047 (Dornoch Medical Systems, Inc., 2008) has gone furthest, with 4 descendants of its own.

How does its category compare?

All FYD devices

FYD Apparatus, Exhaust, Surgical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FYD

2in 2024

2009: 12014: 12015: 12017: 12018: 12019: 32020: 12021: 12022: 12024: 2
2007high 3 (2019) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FYD

71in 2019

2019: 71 days median
2007high 71 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FYD

3so far in 2026, the most since 2007

2010: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FYD

15in 2025

2007: 32008: 42009: 52010: 392011: 1502012: 1202013: 1422014: 712015: 212016: 22017: 32018: 52019: 112020: 112021: 202022: 462023: 192024: 112025: 15
2007high 150 (2011) · low 2 (2016)2025

MAUDE reports naming product code FYD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—03
20080—04
20091—05
20100—139
20110—0150
20120—0120
20130—0142
20141—071
20151—021
20160—02
20171—03
20181—05
2019371011
20201—011
20211—020
20221—046
20230—019
20242—011
20250—015
20260—3—

What is it indicated for?

The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K012991: Neptune Waste Management Ststem.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K012991

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026