510(k) Report

510(k) · Cleared

Recalled · 2004

Glucopro Evo Infusion Pump

K013309510(k) TraditionalCleared Jun 24, 2002263-day reviewClass II21 CFR 880.5725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nipro Diabetes Systems, Inc. · LZG Pump, Infusion, Insulin

Cleared on Jun 24, 2002 as substantially equivalent to K973044 (Disetronic Medical Systems, 1997) and 1 other predicate after a 263-day review. Its family tree reaches back 2 generations to 1990. 1 of the 17 devices in its family, counting itself, has a Class I recall. 12 later devices descend from it. It has been recalled once.

Decided
Jun 24, 2002
Review
263 days
Class
Class II
Product code
LZG
Recalls
1
Predicates
2

Its longest line of descent runs back 2 generations to K901588 (1990). 4 ancestors in all. 12 later devices descend from it, 1 citing it directly. 1 of the 17 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1991: 1 other ancestor2005: 1 descendant cleared2007: 1 descendant cleared2011: 1 descendant cleared, 1 with a Class I recall12014: 2 descendants cleared2016: 3 descendants cleared2015: 1 descendant cleared2024: 1 descendant cleared2026: 2 descendants clearedK901588 · Minimed (R) Model 506 External Insulin Pump · Medtronic Minimed · 1990K960001 · Minimed Infusion Pump, Model MMT-507 · Medtronic Minimed · 19961990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant, 1 with a Class I recall14 generations forward: 2 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants7 generations forward: 1 descendant8 generations forward: 1 descendant K960001 · Minimed Infusion Pump, Model MMT-507 · Medtronic Minimed · 1996K901588 · Minimed (R) Model 506 External Insulin Pump · Medtronic Minimed · 19900+5
  1. Root · 2 generations backK901588 Minimed (R) Model 506 External Insulin PumpMedtronic Minimed · 1990
  2. 1 generation backK960001 Minimed Infusion Pump, Model MMT-507Medtronic Minimed · 1996
  3. This deviceK013309 Glucopro Evo Infusion PumpNipro Diabetes Systems, Inc. · 2002
  4. Descendant · cites it directly · 11 of its ownK050312 Amigo Insulin PumpNipro Diabetes Systems, Inc. · 2005
  5. Descendant · 2 generations on · 10 of its ownK071613 Amigo Insulin PumpNipro Diabetes Systems, Inc. · 2007
  6. Descendant · 3 generations on · 9 of its ownK111210 T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc. · 2011 · Recalled (Class I)
  7. Descendant · 4 generations on · 8 of its ownK133593 T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014
  8. Descendant · 4 generations on · 7 of its ownK141758 T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2005: 12007: 12011: 12014: 22015: 12016: 32024: 12026: 2
2005high 3 (2016) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K050312Amigo Insulin PumpNipro Diabetes Systems, Inc. · 2005 · cites it directly11descendants
  • K071613Amigo Insulin PumpNipro Diabetes Systems, Inc. · 2007 · 2 generations on10descendants
  • K111210T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc. · 2011 · 3 generations on9descendants
  • K133593T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014 · 4 generations on8descendants
  • K141758T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014 · 4 generations on7descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K901588Minimed (R) Model 506 External Insulin PumpMedtronic Minimed1990—
−1K960001Minimed Infusion Pump, Model MMT-507Medtronic Minimed1996—
0 (this device)K013309Glucopro Evo Infusion PumpNipro Diabetes Systems, Inc.2002—
+1K050312Amigo Insulin PumpNipro Diabetes Systems, Inc.2005—
+2K071613Amigo Insulin PumpNipro Diabetes Systems, Inc.2007—
+3K111210T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc.2011Class I
+4K133593T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc.2014—
+4K141758T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200710
201111
201420
201510
201630
202410
202620
FIG. 1 — Lineage of K013309, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZG device

  1. #1 K253534 Pivot Insulin Delivery Systemscore 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973044Disetronic H-Tron Plus V 100 Insulin Infusion Pump
Summary page 1
…&on infusion pump (21 CFR §SS0.5725),C h 3807.92 (a)(3) Legally Marketed Substantially t Gquivalent Devices: Disetronic EI-Tran Plus V 100 infusion pump (K973044) and Minim& 'vfodel507 (K960001) 13escriptiori of Device: The GhicIPro Inksion Pump i s intended far use with tile disposa…
Disetronic Medical Systems19975
*K960001Minimed Infusion Pump, Model MMT-507
Summary page 1
…725),C h 3807.92 (a)(3) Legally Marketed Substantially t Gquivalent Devices: Disetronic EI-Tran Plus V 100 infusion pump (K973044) and Minim& 'vfodel507 (K960001) 13escriptiori of Device: The GhicIPro Inksion Pump i s intended far use with tile disposable compatible infUsicPn sets an…
Medtronic Minimed1996—

Also cleared against K973044: 4 other devices →

What descends from it? (12)

See all 12

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050312Amigo Insulin PumpNipro Diabetes Systems, Inc.200511

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 7, 2004Not classifiedZ-1335-04OtherA FDA inspection of the firm revealed deficiencies in the company's quality system. Problems exhibited may include motor failure or unintended operation.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111210T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 6.1 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K050312 (Nipro Diabetes Systems, Inc., 2005) has gone furthest, with 11 descendants of its own.
  • 4 other devices also cite its predicate K973044.

How does its category compare?

All LZG devices

LZG Pump, Infusion, Insulin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LZG

2in 2024

2007: 32008: 62009: 22010: 22011: 42012: 32013: 22014: 42015: 12016: 42017: 12018: 12019: 22021: 12022: 12023: 12024: 22026 (partial year): 2
2007high 6 (2008) · low 0 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZG

116in 2016

2007: 112 days median2008: 129 days median2011: 189 days median2012: 73 days median2014: 123 days median2016: 116 days median
2007high 189 (2011) · low 73 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZG

1in 2022

2007: 12008: 132009: 22010: 42011: 22012: 42013: 3, 1 Class I2014: 2, 1 Class I2015: 2, 2 Class I2018: 3, 2 Class I2019: 52022: 1, 1 Class I2026 (partial year): 2, 2 Class I
2007high 13 (2008) · low 0 (2016)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · LZG

12,004in 2025

2007: 5,1102008: 5,7112009: 9,4712010: 10,1752011: 12,8582012: 18,3672013: 30,0592014: 49,1882015: 69,1362016: 56,5812017: 44,4722018: 32,6522019: 29,0272020: 32,5882021: 31,5222022: 28,2252023: 21,9722024: 18,2282025: 12,004
2007high 69,136 (2015) · low 5,110 (2007)2025

MAUDE reports naming product code LZG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007311215,110
20086129135,711
20092—29,471
20102—410,175
20114189212,858
2012373418,367
20132—3 (1)30,059
201441232 (1)49,188
20151—2 (2)69,136
20164116056,581
20171—044,472
20181—3 (2)32,652
20192—529,027
20200—032,588
20211—031,522
20221—1 (1)28,225
20231—021,972
20242—018,228
20250—012,004
20262—2 (2)—

What is it indicated for?

The GlucoPro Infusion Pump is intended for the subcutaneous infusion of insulin.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K013309: Glucopro Evo Infusion Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K013309

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026