# K013480: Cobe Smarxt Optimin Surface Modified Hollow Fiber Membrane Oxygenator

- Pathway: 510(k) (Special)
- Applicant: Cobe Cardiovascular, Inc.
- Decided: Nov 16, 2001 after 28 days
- Product code: DTZ (Oxygenator, Cardiopulmonary Bypass)
- Device class: 2
- Regulation: 21 CFR 870.4350
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K960074 (1996). 3 ancestors in all.
No later device cites it as a predicate yet.
0 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K991452: Cobe Optimin Hollow Fiber Membrane Oxygenator, Cobe Cardiovascular, Inc. (1999) Summary page 1.

## Recalls of this device
- Jun 26, 2006 · Not classified · Z-1424-06: Cooling and rewarming time is extended. Efficiency of the heat exchangers is reduced because many of the folds that are critical to heat exchange on the blood side of the units were blocked with a met
- Jun 26, 2006 · Not classified · Z-1427-06: Cooling and rewarming time is extended. Efficiency of the heat exchangers is reduced because many of the folds that are critical to heat exchange on the blood side of the units were blocked with a met

## Indications for use
The COBE® SMARXT® Optimin™ is indicated for use in surgical procedures requiring extracorporeal gas exchange support and blood temperature control. It is intended to be used in procedures requiring a maximum blood flow rate of 5 liters/min and lasting up to six hours.