510(k) Report

510(k) · Cleared

Recalled · 2007

Breathtek - Ubit Ubt for H. Pylori

K014225510(k) SpecialCleared Jan 17, 200222-day reviewClass I21 CFR 866.3110Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Meretek Diagnostics, Inc. · MSQ Test, Urea (Breath Or Blood)

Cleared on Jan 17, 2002 as substantially equivalent to K013371 (Otsuka Pharmaceutical Co., Ltd., 2001) after a 22-day review. Its family tree reaches back 2 generations to 1997. It has been recalled once.

Decided
Jan 17, 2002
Review
22 days
Class
Class I
Product code
MSQ
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K974322 (1997). No later device cites it as a predicate yet.

  1. Root · 2 generations backK974322 Abca-NT Gas Isotope Ratio Mass Spectrometer SystemEuropa Scientific, Ltd. · 1997
  2. 1 generation backK013371 Ubit-IR300 Infrared Spectrometry SystemOtsuka Pharmaceutical Co., Ltd. · 2001 · Recalled
  3. This deviceK014225 Breathtek - Ubit Ubt for H. PyloriMeretek Diagnostics, Inc. · 2002
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K974322Abca-NT Gas Isotope Ratio Mass Spectrometer SystemEuropa Scientific, Ltd.1997—
−1K013371Ubit-IR300 Infrared Spectrometry SystemOtsuka Pharmaceutical Co., Ltd.20011
0 (this device)K014225Breathtek - Ubit Ubt for H. PyloriMeretek Diagnostics, Inc.2002—
FIG. 1 — Lineage of K014225, 1997–2002Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSQ device

  1. #1 K220494 BreathID Hp Lab System, BreathID Smart Systemscore 8.4 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013371Ubit-IR300 Infrared Spectrometry System
Our OCR of the scanned summary read “K013371” beside “510(k)”; FDA’s K013371 is the only number that fits it and carries the summary’s name for the device.
Summary page 4
…IRMS spectrometer and the bag samples using an UBiT IR spectrophotometer. The BreathTek™ - UBIiT® UBT kit will be advertised for use with the UBIT IR 300 System (510(k) #K013371). The BreathTek™ kit will be advertised for use with the ABCA-NT Gas Isotope Ratio Mass Spectrometer System and the ABCA Gas Isot…
Otsuka Pharmaceutical Co., Ltd.20011

Also cleared against K013371: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 2, 2007Not classifiedZ-0003-2008Labeling False and MisleadingDiagnostic test kit for Helicobacter pylori was distributed for adult and pediatric use, but was not approved for pediatric use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 3 companies; a new generation was cleared every 4.0 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K013371.

How does its category compare?

All MSQ devices

MSQ Test, Urea (Breath Or Blood): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSQ

3in 2022

2010: 22013: 12016: 12018: 22020: 12022: 3
2007high 3 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSQ

121in 2022

2022: 121 days median
2007high 121 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSQ

1in 2011

2007: 12011: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MSQ

1in 2025

2007: 22009: 12011: 22012: 42013: 12014: 22015: 42016: 22017: 22022: 12024: 12025: 1
2007high 4 (2012) · low 0 (2008)2025

MAUDE reports naming product code MSQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—12
20080—00
20090—01
20102—00
20110—12
20120—04
20131—01
20140—02
20150—04
20161—02
20170—02
20182—00
20190—00
20201—00
20210—00
2022312101
20230—00
20240—01
20250—01
20260—0—

What is it indicated for?

The BreathTek™ - UBiT® UBT for H. pylori collection kit is intended for use in the qualitative detection of urease associated with Helicobacter pylori in the human stomach and as an aid in the initial diagnosis and post-treatment monitoring of Helicobacter pylori infection in adult patients. The test may be used for monitoring treatment if used at least four weeks following completion of therapy.

For these purposes, the test utilizes the UBIT IR spectrophotometer for the measurement of the ratio of 13CO; to 12COzcontent in breath gas.

For administration by health care professionals. To be administered under a physician’s supervision.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K014225: Breathtek - Ubit Ubt for H. Pylori.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K014225

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026