Acs:180 and Advia Centaur Ipth Immunoassay

K020217510(k) TraditionalCleared Mar 15, 200252-day reviewClass I21 CFR 862.1660Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bayer Diagnostics Corp. · JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

Cleared on Mar 15, 2002 as substantially equivalent to K954418 (Nichols Institute Diagnostics, 1995) after a 52-day review. Its family tree reaches back 1 generation to 1995. 12 later devices descend from it. It has been recalled 4 times.

Decided
Mar 15, 2002
Review
52 days
Class
Class I
Product code
JJX
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K954418 (1995). 12 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2013: 3 descendants cleared2014: 4 descendants cleared2015: 2 descendants cleared2018: 2 descendants cleared2022: 1 descendant clearedK954418 · Intact Parathyroid Hormone (PTH) Immunoassay (Modification) · Nichols Institute Diagnostics · 19951995200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 4 descendants3 generations forward: 4 descendants4 generations forward: 1 descendant K954418 · Intact Parathyroid Hormone (PTH) Immunoassay (Modification) · Nichols Institute Diagnostics · 19950
  1. Root · 1 generation backK954418 Intact Parathyroid Hormone (PTH) Immunoassay (Modification)Nichols Institute Diagnostics · 1995
  2. This deviceK020217 Acs:180 and Advia Centaur Ipth ImmunoassayBayer Diagnostics Corp. · 2002
  3. Descendant · cites it directly · 5 of its ownK133124 Immulite 2000; Total T3 Calibration Verification Material, Total T4…Siemens Healthcare Diagnostics, Inc. · 2013
  4. Descendant · cites it directly · 4 of its ownK132515 Liaison N-Tact PTH Gen II, Control Set, Calibration VerifiersDiaSorin, Inc. · 2013
  5. Descendant · 2 generations on · 3 of its ownK141463 Liason XL1,25 Dihydroxyvitamin D; Control Set, Calibration VerifiersDiaSorin, Inc. · 2014
  6. Descendant · 2 generations on · 3 of its ownK140818 Immulite: 2000 Free T4 Calibration Verification Material, 2000 Igf-I…Siemens Healthcare Diagnostics, Inc. · 2014
  7. Descendant · 3 generations on · 1 of its ownK182698 LIAISON Calprotectin, LIAISON Calprotectin Control Set, LIAISON…DiaSorin, Inc. · 2018

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2013: 32014: 42015: 22018: 22022: 1
2013high 4 (2014) · low 0 (2016)2026

Most influential descendants

Showing 5 of 12 · See all
  • K133124Immulite 2000; Total T3 Calibration Verification Material, Total T4 Calibration…Siemens Healthcare Diagnostics, Inc. · 2013 · cites it directly5descendants
  • K132515Liaison N-Tact PTH Gen II, Control Set, Calibration VerifiersDiaSorin, Inc. · 2013 · cites it directly4descendants
  • K141463Liason XL1,25 Dihydroxyvitamin D; Control Set, Calibration VerifiersDiaSorin, Inc. · 2014 · 2 generations on3descendants
  • K140818Immulite: 2000 Free T4 Calibration Verification Material, 2000 Igf-I Calibration…Siemens Healthcare Diagnostics, Inc. · 2014 · 2 generations on3descendants
  • K182698LIAISON Calprotectin, LIAISON Calprotectin Control Set, LIAISON Calprotectin Calibration…DiaSorin, Inc. · 2018 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K954418Intact Parathyroid Hormone (PTH) Immunoassay (Modification)Nichols Institute Diagnostics1995—
0 (this device)K020217Acs:180 and Advia Centaur Ipth ImmunoassayBayer Diagnostics Corp.2002—
+1K133124Immulite 2000; Total T3 Calibration Verification Material, Total T4…Siemens Healthcare Diagnostics, Inc.2013—
+1K132515Liaison N-Tact PTH Gen II, Control Set, Calibration VerifiersDiaSorin, Inc.2013—
+2K141463Liason XL1,25 Dihydroxyvitamin D; Control Set, Calibration VerifiersDiaSorin, Inc.2014—
+2K140818Immulite: 2000 Free T4 Calibration Verification Material, 2000 Igf-I…Siemens Healthcare Diagnostics, Inc.2014—
+3K182698LIAISON Calprotectin, LIAISON Calprotectin Control Set, LIAISON Calprotectin…DiaSorin, Inc.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201330
201440
201520
201820
202210
FIG. 1 — Lineage of K020217, 1995–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K954418Intact Parathyroid Hormone (PTH) Immunoassay (Modification)
Summary page 1
…Nichols Institute Diagnostics Intact Parathyroid Hormone (PTH) kit / Nichols IRMA Intact PTH Nichols Institute Diagnostics 33608 Ortega Highway San Juan Capistrano, Ca K954418 4. Device Description The ACS:180 Intact PTH assay is a two-site “sandwich” immunoassay using direct, chemilumin…
Nichols Institute Diagnostics1995—

Also cleared against K954418: 1 other device →

What descends from it? (12)

See all 12

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133124Immulite 2000; Total T3 Calibration Verification Material, Total T4 Calibration…Siemens Healthcare Diagnostics, Inc.20135
K132515Liaison N-Tact PTH Gen II, Control Set, Calibration VerifiersDiaSorin, Inc.20134
K132391Immulite 2000 Anti-TG AB Calibration Verification Material, Immulite 2000 Anti-Tpo AB…Siemens Helathcare Diagnostics, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 16, 2013Class IIIZ-0173-2014PackagingPTH Calibrator packaged with only high calibrator.
May 14, 2013Class IIZ-1514-2013No Marketing ApplicationiPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability
May 14, 2013Class IIZ-1515-2013No Marketing ApplicationiPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability
Oct 3, 2008Not classifiedZ-1186-2012Process controlPoor sensitivity and imprecision on ADVIA Centaur systems; Serum samples low bias; EDTA plasma samples low bias

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K133124 (Siemens Healthcare Diagnostics, Inc., 2013) has gone furthest, with 5 descendants of its own.
  • 1 other device also cites its predicate K954418.

How does its category compare?

All JJX devices

JJX Single (Specified) Analyte Controls (Assayed And Unassayed): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JJX

2in 2017

2007: 62008: 82009: 172010: 242011: 192012: 152013: 192014: 222015: 112016: 62017: 2
2007high 24 (2010) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JJX

31in 2016

2007: 53 days median2008: 52 days median2009: 80 days median2010: 80 days median2011: 62 days median2012: 40 days median2013: 37 days median2014: 30 days median2015: 65 days median2016: 31 days median
2007high 80 (2009) · low 30 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JJX

1in 2021

2007: 32008: 32010: 32011: 22014: 12015: 12016: 22018: 22021: 1
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJX

22in 2025

2007: 22009: 12010: 62011: 622012: 322013: 512014: 482015: 672016: 372017: 192018: 352019: 822020: 772021: 432022: 242023: 222024: 302025: 22
2007high 82 (2019) · low 0 (2008)2025

MAUDE reports naming product code JJX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200765332
200885230
2009178001
2010248036
20111962262
20121540032
20131937051
20142230148
20151165167
2016631237
20172—019
20180—235
20190—082
20200—077
20210—143
20220—024
20230—022
20240—030
20250—022
20260—0—

What is it indicated for?

The ACS:180 and ADVIA Centaur Intact PTH Immunoassay is intended for the quantitative

determination of intact parathyroid hormone in human EDTA plasma on the automated ACS:180 and

ADVIA Centaur analyzers marketed by Bayer Corporation. Intact PTH levels can be used to aide in the

differential diagnosis of hyperparathyroidism, hypoparathyroidism, or hypercalcemia of malignancy.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020217: Acs:180 and Advia Centaur Ipth Immunoassay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020217

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026