510(k) Report

510(k) · Cleared

Recalled · 2010

Inion CPS Bsso Screw

K020266510(k) SpecialCleared Feb 22, 200228-day reviewClass II21 CFR 888.3040Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Inion, Ltd. · HWC Screw, Fixation, Bone

Cleared on Feb 22, 2002 as substantially equivalent to K010352 (Inion, Ltd., 2001) after a 28-day review. Its family tree reaches back 1 generation to 2001. 1 later device descends from it. It has been recalled 3 times.

Decided
Feb 22, 2002
Review
28 days
Class
Class II
Product code
HWC
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K010352 (2001). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2025: 1 descendant clearedK010352 · CPS 1.5/2.0/2.5 Bioabsorbable Fixation System · Inion, Ltd. · 2001 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant K010352 · CPS 1.5/2.0/2.5 Bioabsorbable Fixation System · Inion, Ltd. · 2001 · recalled0
  1. Root · 1 generation backK010352 CPS 1.5/2.0/2.5 Bioabsorbable Fixation SystemInion, Ltd. · 2001 · Recalled
  2. This deviceK020266 Inion CPS Bsso ScrewInion, Ltd. · 2002
  3. Descendant · cites it directly · 0 of its ownK251441 Inion CPS 1.5 Baby Bioabsorbable Fixation System; Inion CPS…Inion OY · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K251441Inion CPS 1.5 Baby Bioabsorbable Fixation System; Inion CPS 1.5/2.0/2.5 Bioabsorbable…Inion OY · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K010352CPS 1.5/2.0/2.5 Bioabsorbable Fixation SystemInion, Ltd.20013
0 (this device)K020266Inion CPS Bsso ScrewInion, Ltd.2002—
+1K251441Inion CPS 1.5 Baby Bioabsorbable Fixation System; Inion CPS 1.5/2.0/2.5…Inion OY2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K020266, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010352CPS 1.5/2.0/2.5 Bioabsorbable Fixation System
Summary page 1
…CFR 888.3040 screws are classified as Class II. Screws have been assigned Product Code HWC. PREDICATE DEVICE (1) Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System (K010352) Special 510(k) 23.1.2002 Final i D/L K 0 0alofr a 8 INTENDED USE The Inion CPS" BSSO Screw is intended for u…
Inion, Ltd.20013

Also cleared against K010352: 8 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251441Inion CPS 1.5 Baby Bioabsorbable Fixation System; Inion CPS 1.5/2.0/2.5 Bioabsorbable…Inion OY20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 14, 2010Not classifiedZ-0630-2011OtherInadequate Instructions for Use. The stated strength retention is ambiguous and the maximum immersion time of the plates/meshed is not defined.
Apr 14, 2010Not classifiedZ-0631-2011OtherInadequate Instructions for Use. The stated strength retention is ambiguous and the maximum immersion time of the plates/meshed is not defined.
Mar 5, 2010Not classifiedZ-1550-2010Nonconforming Material/ComponentDuring a sampling inspection, 2 lots of screws did not meet shear and torsion testing specifications.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K251441 (Inion OY, 2025) has gone furthest, with 0 descendants of its own.
  • 8 other devices also cite its predicate K010352.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

Indications:

A. General indications: The Inion CPS™ BSSO Screw is intended for use in trauma and reconstructive procedures in the craniofacial skeleton, midface, maxilla and mandible (in conjunction with appropriate maxillomandibular fixation) as a part of the Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System.

B. Specific indications:

Fractures of the cranium, midface, maxilla and mandible.

Infant craniofacial surgery (i.e. craniosynostosis, congenital malformations).

LeFort (I, II, III) osteotomies.

Pediatric reconstructive procedures.

Orthognathic or reconstructive procedures of the cranium, midface, maxilla or mandible.

Craniotomy flap fixation.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020266: Inion CPS Bsso Screw.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020266

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026