510(k) Report

510(k) · Cleared

Recalled · 2010

Oscor Pace 203/Pace 203H Dual-Chamber Temporary Pacemaker, ST. Jude Medical, Model 3085 Dual-Chamber External Pulse Gene

K020896510(k) AbbreviatedCleared Jun 14, 200287-day reviewClass II21 CFR 870.3600Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Osypka Medical, Inc. · DTE Pulse-Generator, Pacemaker, External

Cleared on Jun 14, 2002 as substantially equivalent to K910237 (Medtronic Vascular, 1991) and 3 other predicates after an 87-day review. Its family tree reaches back 1 generation to 1989. 2 later devices descend from it. It has been recalled 2 times.

Decided
Jun 14, 2002
Review
87 days
Class
Class II
Product code
DTE
Recalls
2
Predicates
4

Its longest line of descent runs back 1 generation to K893633 (1989). 4 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestor2002: 1 descendant cleared2003: 1 descendant clearedK893633 · Model 5330 External a-V Sequential Demand Pulse · Medtronic Vascular · 19891990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 2 descendants K893633 · Model 5330 External a-V Sequential Demand Pulse · Medtronic Vascular · 19890
  1. Root · 1 generation backK893633 Model 5330 External a-V Sequential Demand PulseMedtronic Vascular · 1989
  2. This deviceK020896 Oscor Pace 203/Pace 203H Dual-Chamber Temporary Pacemaker, ST. Jude…Osypka Medical, Inc. · 2002
  3. Descendant · cites it directly · 1 of its ownK022939 Oscor Pace 101/101H and Accessories, ST.Jude Medical Model 3077 and…Osypka Medical, Inc. · 2002
  4. Descendant · cites it directly · 0 of its ownK033130 Pace 101/Pace 101H, Model 3077 Ssi Temporary Pulse GeneratorOsypka Medical, Inc. · 2003

Most influential descendants

Showing 2 of 2 · See all
  • K022939Oscor Pace 101/101H and Accessories, ST.Jude Medical Model 3077 and…Osypka Medical, Inc. · 2002 · cites it directly1descendant
  • K033130Pace 101/Pace 101H, Model 3077 Ssi Temporary Pulse GeneratorOsypka Medical, Inc. · 2003 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K893633Model 5330 External a-V Sequential Demand PulseMedtronic Vascular1989—
0 (this device)K020896Oscor Pace 203/Pace 203H Dual-Chamber Temporary Pacemaker, ST. Jude Medical…Osypka Medical, Inc.2002—
+1K022939Oscor Pace 101/101H and Accessories, ST.Jude Medical Model 3077 and…Osypka Medical, Inc.2002—
+1K033130Pace 101/Pace 101H, Model 3077 Ssi Temporary Pulse GeneratorOsypka Medical, Inc.2003—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199310
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200310
FIG. 1 — Lineage of K020896, 1989–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTE device

  1. #1 K232721 Lifetech Cardio Temporary Pacemakerscore 7.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K910237Modified Model 5330 AV Demand Pulse Generator
Summary page 1
…Classification 870.3600 Pulse-Generator, Pacemaker, External Names: 870.2900 Cables, Transducer and Electrode Predicate K910237 MEDTRONIC" MODIFIED MODEL 5330 Devices: A-V Demand Pulse Generator…
Medtronic Vascular1991—
*K893633Model 5330 External a-V Sequential Demand Pulse
Summary page 1
…rode Predicate K910237 MEDTRONIC" MODIFIED MODEL 5330 Devices: A-V Demand Pulse Generator K893633 MEDTRONICTMODEL 5330 External A-V Sequential Demand Pulse Generator…
Medtronic Vascular1989—
*K970497PACE 101H
Summary page 1
…K893633 MEDTRONICTMODEL 5330 External A-V Sequential Demand Pulse Generator K970497 OSCOR" D-1 / D-13 / 0-5 / D-9 / D-10 Extension Cables PACE 101H Arm Strap…
Sulzer Oscor, Inc.1997—
*K923621External Demand Pacemaker
Summary page 1
…K970497 OSCOR" D-1 / D-13 / 0-5 / D-9 / D-10 Extension Cables PACE 101H Arm Strap K923621 OSCOR" DX-2 / 0-5 / D-I 0 Extension Cables PACE 100H Arm Strap Devic…
Oscor Medical Corp.1993—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K022939Oscor Pace 101/101H and Accessories, ST.Jude Medical Model 3077 and…Osypka Medical, Inc.20021
K033130Pace 101/Pace 101H, Model 3077 Ssi Temporary Pulse GeneratorOsypka Medical, Inc.20030

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 30, 2010Not classifiedZ-2443-2010Device DesignThe recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a…
Jun 30, 2010Not classifiedZ-2444-2010Device DesignThe recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 4 ancestors has a recall.
  • Of its descendants, K022939 (Osypka Medical, Inc., 2002) has gone furthest, with 1 descendant of its own.

How does its category compare?

All DTE devices

DTE Pulse-Generator, Pacemaker, External: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTE

1in 2025

2013: 22015: 12016: 22018: 22019: 22020: 12024: 12025: 1
2007high 2 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTE

Too few DTE 510(k)s in any year since 2007 to give a median.

Recalls per year · DTE

2in 2010

2010: 2
2007high 2 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DTE

636in 2025

2007: 172008: 222009: 392010: 3352011: 1,8652012: 1,7862013: 982014: 672015: 3422016: 4962017: 6542018: 9562019: 1,0012020: 1,0922021: 8992022: 8332023: 6502024: 6202025: 636
2007high 1,865 (2011) · low 17 (2007)2025

MAUDE reports naming product code DTE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—017
20080—022
20090—039
20100—2335
20110—01,865
20120—01,786
20132—098
20140—067
20151—0342
20162—0496
20170—0654
20182—0956
20192—01,001
20201—01,092
20210—0899
20220—0833
20230—0650
20241—0620
20251—0636
20260—0—

What is it indicated for?

The PACE 203 / PACE 203H / MODEL 3085 external pulse generator / temporary pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203 / PACE 203H / MODEL 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes. Specific indications include, but are not limited to, the following: Sick Sinus Syndrome; Bradycardia with…

Read the full statement (187 words)

The PACE 203 / PACE 203H / MODEL 3085 external pulse generator / temporary pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203 / PACE 203H / MODEL 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.

Specific indications include, but are not limited to, the following:

Sick Sinus Syndrome;

Bradycardia with congestive heart failure;

Complete heart block;

Acute myocardial infarction complicated by heart block;

Sinus bradycardia;

Cardiac arrest with ventricular asystole;

Atrial and/or ventricular ectopic arrhythmia;

Postoperatively after cardiac surgery;

Temporary application during implantation or exchange of permanent pacemaker.

Indication for atrial overdrive stimulation (rapid atrial pacing):

Supra-ventricular tachycardia.

The PACE 203 / PACE 203H / MODEL 3085 can be used to measure atrial or ventricular voltage stimulation thresholds and determine the approximate P wave potentials sensed via the atrial pacing lead electrodes and the approximate R wave potentials sensed via the ventricular pacing lead electrodes.

The XI.TME and XI.RAC Extension Cables are used with external pulse generators / temporary cardiac pacemakers and pacing system analyzers.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K020896: Oscor Pace 203/Pace 203H Dual-Chamber Temporary Pacemaker, ST. Jude Medical, Model 3085 Dual-Chamber External Pulse Gene.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020896

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026