510(k) · Cleared
Recalled · Class II · 2013Introcan Safety IV Catheter
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
B.Braun Medical, Inc. · DYB Introducer, Catheter
Cleared on Jul 3, 2002 as substantially equivalent to K982805 (B.Braun Medical, Inc., 1999) and 3 other predicates after a 90-day review. Its family tree reaches back 2 generations to 1980. 26 later devices descend from it. It has been recalled once.
- Decided
- Jul 3, 2002
- Review
- 90 days
- Class
- Class II
- Product code
- DYB
- Recalls
- 1
- Predicates
- 4
Its longest line of descent runs back 2 generations to K801941 (1980). 5 ancestors in all. 26 later devices descend from it, 3 citing it directly.
- Root · 2 generations backK801941 Burron Microcath Intra. Cath. PlacementtBurron Medical Products, Inc. · 1980
- 1 generation backK982805 Introcan Safety IV CatheterB.Braun Medical, Inc. · 1999 · Recalled
- This deviceK021094 Introcan Safety IV CatheterB.Braun Medical, Inc. · 2002
- Descendant · cites it directly · 15 of its ownK111236 Introcan Safety (R) 3 Closed IV Catheter 18GA X 1-3/4 (45MM) and…B.Braun Melsungen AG · 2011 · Recalled
- Descendant · 2 generations on · 9 of its ownK163358 Introcan Safety 3 Closed IV CatheterB.Braun Melsungen AG · 2017
- Descendant · 2 generations on · 8 of its ownK182870 Introcan Safety 3 Closed IV CatheterB.Braun Medical, Inc. · 2019 · Recalled
- Descendant · cites it directly · 7 of its ownK030135 Safety Introducer NeedleB.Braun Medical, Inc. · 2003
- Descendant · 3 generations on · 6 of its ownK192676 Introcan Safety 2 IV Catheter 18-24 gaugeB.Braun Medical, Inc. · 2020
Showing 5 of 26 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 26 · See all- K111236Introcan Safety (R) 3 Closed IV Catheter 18GA X 1-3/4 (45MM) and 18GA X 1-1/4 (32MM) /…B.Braun Melsungen AG · 2011 · cites it directly15descendants
- K163358Introcan Safety 3 Closed IV CatheterB.Braun Melsungen AG · 2017 · 2 generations on9descendants
- K182870Introcan Safety 3 Closed IV CatheterB.Braun Medical, Inc. · 2019 · 2 generations on8descendants
- K030135Safety Introducer NeedleB.Braun Medical, Inc. · 2003 · cites it directly7descendants
- K192676Introcan Safety 2 IV Catheter 18-24 gaugeB.Braun Medical, Inc. · 2020 · 3 generations on6descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K801941 | Burron Microcath Intra. Cath. Placementt | Burron Medical Products, Inc. | 1980 | — |
| −1 | K982805 | Introcan Safety IV Catheter | B.Braun Medical, Inc. | 1999 | 1 |
| 0 (this device) | K021094 | Introcan Safety IV Catheter | B.Braun Medical, Inc. | 2002 | — |
| +1 | K111236 | Introcan Safety (R) 3 Closed IV Catheter 18GA X 1-3/4 (45MM) and 18GA X 1-1/4… | B.Braun Melsungen AG | 2011 | 1 |
| +2 | K163358 | Introcan Safety 3 Closed IV Catheter | B.Braun Melsungen AG | 2017 | — |
| +2 | K182870 | Introcan Safety 3 Closed IV Catheter | B.Braun Medical, Inc. | 2019 | 1 |
| +1 | K030135 | Safety Introducer Needle | B.Braun Medical, Inc. | 2003 | — |
| +3 | K192676 | Introcan Safety 2 IV Catheter 18-24 gauge | B.Braun Medical, Inc. | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 2 | 0 |
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 1 | 0 |
| 2005 | 1 | 0 |
| 2007 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 4 | 0 |
| 2021 | 1 | 0 |
| 2022 | 3 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 2 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DYB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DYB device
- #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K982805 | Introcan Safety IV CatheterSummary page 1…r Catheter and CLASSIFICATION: 21 CFR 5 870.1340: Catheter Introducer PREDICATE DEVICE: B. Braun Medical Inc. Introcan Safety IV Catheter (K982805) B. Braun Medical Inc. Introducer Catheter (Kit component of… | 1999 |
| * | K810460 | Burron Percutaneous Inducer SetSummary page 1…Introcan Safety IV Catheter (K982805) B. Braun Medical Inc. Introducer Catheter (Kit component of K8 10460 / K8 10461) Becton Dickinson Infusion Therapy Systems Inc., BD IntrosyteTM… | 1981 |
| * | K810461 | Burron Central Vein Catherization KitSummary page 1…afety IV Catheter (K982805) B. Braun Medical Inc. Introducer Catheter (Kit component of K8 10460 / K8 10461) Becton Dickinson Infusion Therapy Systems Inc., BD IntrosyteTM… | 1981 |
| * | K013304 | BD Introsyte Precision Introducer, BD Introsyte Autogaurd Shielded IntroducerSummary page 1…1) Becton Dickinson Infusion Therapy Systems Inc., BD IntrosyteTM AutoguardT" Shielded Introducer (K013304) DESCRIPTION: The Introcan Safety IV Catheter consists of an over-the-needle… | 2002 |
What descends from it? (26)
See all 263 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 27, 2013 | Class II | Z-0559-2014 | Labeling False and Misleading | One lot of Introcan Safety IV Catheters may not contain the anti-needlestick safety clip which is used for the prevention of needlestick injuries. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 22 years and passes through 2 companies; a new generation was cleared every 18.4 years on median.
- 1 of the 5 ancestors has a recall of any class.
- Of its descendants, K111236 (B.Braun Melsungen AG, 2011) has gone furthest, with 15 descendants of its own.
- 4 other devices also cite its predicate K982805.
How does its category compare?
All DYB devicesDYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DYB
16in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DYB
63in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DYB
3in 2025
Class I (48 since 2007) Class II/III not classified
Adverse event reports per year · DYB
4,539in 2025
MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 95 | 15 | 316 |
| 2008 | 17 | 37 | 9 | 423 |
| 2009 | 14 | 52 | 25 | 399 |
| 2010 | 18 | 123 | 18 | 445 |
| 2011 | 17 | 100 | 4 | 628 |
| 2012 | 22 | 128 | 3 | 898 |
| 2013 | 26 | 118 | 5 (1) | 743 |
| 2014 | 30 | 122 | 2 | 780 |
| 2015 | 17 | 129 | 2 (1) | 913 |
| 2016 | 19 | 95 | 45 (3) | 1,234 |
| 2017 | 24 | 70 | 22 (1) | 1,672 |
| 2018 | 19 | 156 | 1 | 1,903 |
| 2019 | 18 | 47 | 2 | 2,450 |
| 2020 | 9 | 49 | 21 (21) | 2,465 |
| 2021 | 12 | 78 | 6 | 2,835 |
| 2022 | 13 | 143 | 2 | 4,084 |
| 2023 | 15 | 55 | 10 | 3,775 |
| 2024 | 18 | 56 | 2 | 3,931 |
| 2025 | 16 | 63 | 3 | 4,539 |
| 2026 | 17 | 59 | 23 (21) | — |
What is it indicated for?
The Introcan® Safety™ IV Catheter is a passive anti-needle stick device to provide venous or arterial access for the infusion of fluids, drugs, and/or blood components, or to facilitate the placement of Vascular Access devices such as guidewires, indwelling central venous catheters, peripherally inserted central Catheters, and midline catheters into the vascular system.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K021094 · printed Nov 4, 2026
Cite this page
510(k) Report. “K021094: Introcan Safety IV Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K021094
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026