510(k) Report

510(k) · Cleared

Recalled · 2009

Jump Start, Model 970300

K021168510(k) TraditionalCleared Jun 26, 200276-day reviewClass III21 CFR 870.5310Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medical Research Laboratories, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Jun 26, 2002 after a 76-day review. The FDA summary names no predicate by number. It has been recalled 7 times.

Decided
Jun 26, 2002
Review
76 days
Class
Class III
Product code
MKJ
Recalls
7
Predicates
0

The FDA summary names no predicate by number.

FIG. 1 — Lineage of K021168Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MKJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2009Not classifiedZ-1656-2009Labeling designPossible for users to misunderstand signals. The original product labeling told users that a flashing low battery indicator signals that the battery is low and requires changing soon. If a user experiences a flashing low battery status…
Feb 26, 2009Not classifiedZ-1389-2009Component design/selectionIssue #1: AED 10 displayed a "Do Not Use" symbol as a result of a blown F1 fuse. Issue #2: AED 10 models that incorporate a spring connector with the speaker experienced intermittent electrical connection to the circuit board causing a…
Dec 15, 2008Not classifiedZ-1005-2009Device DesignPotential for device to shut down prematurely under certain circumstances.
Dec 15, 2008Not classifiedZ-1006-2009Component design/selectionReliability issues - potential to shock a non shockable rhythm or not shocking a shockable rhythm;
Dec 15, 2008Not classifiedZ-1007-2009Process controlPotential for device to deliver less than the programmed energy.
Oct 26, 2007Not classifiedZ-0129-2008Nonconforming Material/ComponentFailure to Deliver Shock; a defective capacitor may cause the delay or non-delivery of the defibrillating shock which may result in failure to resuscitate the patient
Apr 26, 2004Not classifiedZ-0918-04OtherA faulty component has been identified which could result in an intermittent shock button (button #3) failure on the AED10 defibrillator.

No ancestor or descendant has a Class I recall.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The JUMP START is intended to be used to tréat patients in cardiopulmonary arrest. It is intended for use in either in-hospital or out-of-hospital arrests. It is intended for use by personnel who are authorized by a physician/medical director, and who have the following training and skilis: 4+ American Heart Association Heartsaver course, American Red Cross CPR/AED course, or equivalent + Training in the use of the MRL JUMP START It should only be used…

Read the full statement (130 words)

The JUMP START is intended to be used to tréat patients in cardiopulmonary arrest. It is intended for use in either in-hospital or out-of-hospital arrests.

It is intended for use by personnel who are authorized by a physician/medical director, and who have the following training and skilis: 4+ American Heart Association Heartsaver course, American Red Cross CPR/AED course, or equivalent + Training in the use of the MRL JUMP START

It should only be used on patients who are unconscious, pulseless, and not breathing spontaneously. The biphasic waveform employed by the JUMP START has not been clinically tested on pediatric patients The device has not been evaluated for cardioversion of atrial fibrillation or direct (internal) cardiac defibrillation. It should not be used on pediatric patients less than 8 years old.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K021168: Jump Start, Model 970300.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K021168

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.