510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Vaxcel Picc with Pasv Valve Technology

K021704510(k) SpecialCleared Jun 6, 200214-day reviewClass II21 CFR 880.5970Recalls 31

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Catheter Innovations, Inc. · LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Cleared on Jun 6, 2002 after a 14-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). 2 of the 35 devices in its family, counting itself, have a Class I recall. 34 later devices descend from it. It has been recalled 31 times.

Decided
Jun 6, 2002
Review
14 days
Class
Class II
Product code
LJS
Recalls
31
Predicates
1

34 later devices descend from it, 4 citing it directly. 2 of the 35 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 1 descendant cleared2009: 2 descendants cleared2012: 3 descendants cleared2016: 2 descendants cleared2021: 1 descendant cleared, 1 with a Class I recall12014: 4 descendants cleared, 1 with a Class I recall12013: 5 descendants cleared2011: 1 descendant cleared2015: 5 descendants cleared2018: 2 descendants cleared2020: 2 descendants cleared2017: 1 descendant cleared2025: 4 descendants cleared2022: 1 descendant cleared200020052010201520202025
  1. This deviceK021704 Vaxcel Picc with Pasv Valve TechnologyCatheter Innovations, Inc. · 2002
  2. Descendant · cites it directly · 27 of its ownK091261 Nmi PiccNavilyst Medical, Inc. · 2009 · Recalled
  3. Descendant · cites it directly · 26 of its ownK093366 Nmi Picc IINavilyst Medical, Inc. · 2009
  4. Descendant · 2 generations on · 23 of its ownK121089 Nmi Picc IiiNavilyst Medical, Inc. · 2012
  5. Descendant · 2 generations on · 14 of its ownK111906 Xcela Hybrid Picc with Pasv Valve TechnologyNavilyst Medical, Inc. · 2011
  6. Descendant · 3 generations on · 12 of its ownK122882 Nmi Hpicc IiiNavilyst Medical, Inc. · 2012

Showing 5 of 34 descendants. See all descendants

Descendants cleared per year

2007: 12009: 22011: 12012: 32013: 52014: 42015: 52016: 22017: 12018: 22020: 22021: 12022: 12025: 4
2007high 5 (2013) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 34 · See all
  • K091261Nmi PiccNavilyst Medical, Inc. · 2009 · cites it directly27descendants
  • K093366Nmi Picc IINavilyst Medical, Inc. · 2009 · cites it directly26descendants
  • K121089Nmi Picc IiiNavilyst Medical, Inc. · 2012 · 2 generations on23descendants
  • K111906Xcela Hybrid Picc with Pasv Valve TechnologyNavilyst Medical, Inc. · 2011 · 2 generations on14descendants
  • K122882Nmi Hpicc IiiNavilyst Medical, Inc. · 2012 · 3 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K021704Vaxcel Picc with Pasv Valve TechnologyCatheter Innovations, Inc.2002—
+1K091261Nmi PiccNavilyst Medical, Inc.20095
+1K093366Nmi Picc IINavilyst Medical, Inc.2009—
+2K121089Nmi Picc IiiNavilyst Medical, Inc.2012—
+2K111906Xcela Hybrid Picc with Pasv Valve TechnologyNavilyst Medical, Inc.2011—
+3K122882Nmi Hpicc IiiNavilyst Medical, Inc.2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200920
201110
201230
201350
201441
201550
201620
201710
201820
202020
202111
202210
202540
FIG. 1 — Lineage of K021704, 2002–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LJS device

  1. #1 K241910 Neonav ECG Tip Location Systemscore 18.0 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK036421PLACEMENT-plus
Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree
Summary page 1
…f Safety and Effectiveness G~+neraf Provisions Trade Name: Vaxcel” PlCC Catheter with PASVBTechnology. Name of Predicate PLACEMENT-plusm [KO036421 Device Classification Class ll Performance Performance Standards have not been established by FDA un…

What descends from it? (34)

See all 34

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091261Nmi PiccNavilyst Medical, Inc.200927
K093366Nmi Picc IINavilyst Medical, Inc.200926
K072230Powerpicc SoloC.R. Bard, Inc.20075
K121990Picc Convenience KitNavilyst Medical, Inc.20120

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 14, 2016Class IIZ-1985-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1982-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1969-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1971-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1972-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1977-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1975-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1987-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1988-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1973-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1980-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1984-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1966-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1974-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1968-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1976-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1979-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1986-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1970-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1967-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1983-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1978-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Mar 14, 2016Class IIZ-1981-2016Device DesignPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Nov 22, 2004Not classifiedZ-0403-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0407-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0408-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0409-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0404-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0405-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0410-05OtherCatheter separation immediately distal to the nose of the suture wing.
Nov 22, 2004Not classifiedZ-0406-05OtherCatheter separation immediately distal to the nose of the suture wing.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K210264BD PowerPiCC CatheterBard Access Systems, Inc. · 2021 · 2 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K131260Nmi Dialysis CatheterNavilyst Medical, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K091261 (Navilyst Medical, Inc., 2009) has gone furthest, with 27 descendants of its own.

How does its category compare?

All LJS devices

LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LJS

8in 2025

2007: 142008: 42009: 162010: 92011: 62012: 142013: 142014: 82015: 72016: 92017: 72018: 102020: 22021: 32022: 32023: 42024: 12025: 82026 (partial year): 2
2007high 16 (2009) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LJS

171in 2025

2007: 92 days median2008: 97 days median2009: 70 days median2010: 114 days median2011: 55 days median2012: 77 days median2013: 31 days median2014: 47 days median2015: 71 days median2016: 128 days median2017: 101 days median2018: 148 days median2021: 284 days median2022: 177 days median2023: 47 days median2025: 171 days median
2007high 284 (2021) · low 31 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LJS

5in 2023

2008: 12009: 12010: 122011: 42012: 402014: 152015: 22016: 182017: 92018: 32021: 12023: 5, 2 Class I2026 (partial year): 8, 1 Class I
2007high 40 (2012) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · LJS

4,372in 2025

2007: 3922008: 3692009: 3792010: 4952011: 6962012: 5642013: 6802014: 8422015: 1,3062016: 1,1232017: 2,1002018: 2,3242019: 2,0522020: 2,5392021: 3,0842022: 3,2562023: 2,7992024: 4,2432025: 4,372
2007high 4,372 (2025) · low 369 (2008)2025

MAUDE reports naming product code LJS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714920392
20084971369
200916701379
2010911412495
20116554696
2012147740564
201314310680
201484715842
201577121,306
20169128181,123
2017710192,100
20181014832,324
20190—02,052
20202—02,539
2021328413,084
2022317703,256
20234475 (2)2,799
20241—04,243
2025817104,372
20262—8 (1)—

What is it indicated for?

The Vaxcel PICC with PASV Valve Technology is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy and blood products. It is also indicated for blood specimen withdrawal.

This product is effective for central venous access in adults, children and infants who require intravenous (IV) therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K021704: Vaxcel Picc with Pasv Valve Technology.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K021704

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.