Equalizer Balloon Catheter

K021721510(k) SpecialCleared Jun 20, 200227-day reviewClass II21 CFR 870.1250Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · DQY Catheter, Percutaneous

Cleared on Jun 20, 2002 after a 27-day review. The FDA summary names its predicate without a number. 1 of the 34 devices in its family, counting itself, has a Class I recall. 33 later devices descend from it. It has been recalled 4 times.

Decided
Jun 20, 2002
Review
27 days
Class
Class II
Product code
DQY
Recalls
4
Predicates
0

33 later devices descend from it, 6 citing it directly. 1 of the 34 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2014: 1 descendant cleared2008: 2 descendants cleared2009: 1 descendant cleared2004: 1 descendant cleared2006: 1 descendant cleared2011: 2 descendants cleared2025: 4 descendants cleared2020: 4 descendants cleared2012: 3 descendants cleared2017: 3 descendants cleared2016: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2018: 1 descendant cleared2019: 2 descendants cleared2021: 4 descendants cleared2022: 1 descendant cleared2023: 1 descendant cleared200020052010201520202025
  1. This deviceK021721 Equalizer Balloon CatheterBoston Scientific Corp · 2002
  2. Descendant · cites it directly · 17 of its ownK140273 Equalizer Occlusion Balloon CatheterBoston Scientific Corp · 2014
  3. Descendant · 2 generations on · 8 of its ownK170411 ER-REBOA CatheterPrytime Medical Devices, Inc. · 2017
  4. Descendant · cites it directly · 7 of its ownK033670 Gore Tri-Lobe Balloon CatheterW. L. Gore & Associates, Inc. · 2004
  5. Descendant · 3 generations on · 4 of its ownK172790 ER-REBOA CatheterPrytime Medical Devices, Inc. · 2017
  6. Descendant · 2 generations on · 4 of its ownK153530 Bridge Occlusion BalloonSpectranetics, Inc. · 2016 · Recalled (Class I)

Showing 5 of 33 descendants. See all descendants

Descendants cleared per year

2004: 12006: 12008: 22009: 12011: 22012: 32013: 12014: 12016: 12017: 32018: 12019: 22020: 42021: 42022: 12023: 12025: 4
2004high 4 (2020) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 33 · See all
  • K140273Equalizer Occlusion Balloon CatheterBoston Scientific Corp · 2014 · cites it directly17descendants
  • K170411ER-REBOA CatheterPrytime Medical Devices, Inc. · 2017 · 2 generations on8descendants
  • K033670Gore Tri-Lobe Balloon CatheterW. L. Gore & Associates, Inc. · 2004 · cites it directly7descendants
  • K172790ER-REBOA CatheterPrytime Medical Devices, Inc. · 2017 · 3 generations on4descendants
  • K153530Bridge Occlusion BalloonSpectranetics, Inc. · 2016 · 2 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K021721Equalizer Balloon CatheterBoston Scientific Corp2002—
+1K140273Equalizer Occlusion Balloon CatheterBoston Scientific Corp2014—
+2K170411ER-REBOA CatheterPrytime Medical Devices, Inc.2017—
+1K033670Gore Tri-Lobe Balloon CatheterW. L. Gore & Associates, Inc.2004—
+3K172790ER-REBOA CatheterPrytime Medical Devices, Inc.2017—
+2K153530Bridge Occlusion BalloonSpectranetics, Inc.2016Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200610
200820
200910
201120
201230
201310
201410
201611
201730
201810
201920
202040
202140
202210
202310
202540
FIG. 1 — Lineage of K021721, 2002–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Large Occlusion Balloon Catheter
Named without a submission number
—
*—Blue Max Balloon Catheter
Named without a submission number
—

What descends from it? (33)

See all 33

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140273Equalizer Occlusion Balloon CatheterBoston Scientific Corp201417
K033670Gore Tri-Lobe Balloon CatheterW. L. Gore & Associates, Inc.20047
K091624Stent Graft Balloon CatheterVia Biomedical, Inc.20093
K081799Gore Tri-Lobe Balloon CatheterW. L. Gore & Associates, Inc.20082
K051137Techdevice Occlusion Balloon CatheterTechdevice Corporation20061
K070126Occlusion Balloon CatheterSentreheart, Inc.20080

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 29, 2009Not classifiedZ-0194-2010OtherBoston Scientific is initiating a Recall Removal of numerous batches/lots of Equalizer Occlusion Balloon Catheter because the sterile barrier in the packaging of the affected product may be compromised. The outer pouch seal may be…
Sep 29, 2009Not classifiedZ-0195-2010OtherBoston Scientific is initiating a Recall Removal of numerous batches/lots of Equalizer Occlusion Balloon Catheter because the sterile barrier in the packaging of the affected product may be compromised. The outer pouch seal may be…
Sep 29, 2009Not classifiedZ-0196-2010OtherBoston Scientific is initiating a Recall Removal of numerous batches/lots of Equalizer Occlusion Balloon Catheter because the sterile barrier in the packaging of the affected product may be compromised. The outer pouch seal may be…
Sep 29, 2009Not classifiedZ-0197-2010OtherBoston Scientific is initiating a Recall Removal of numerous batches/lots of Equalizer Occlusion Balloon Catheter because the sterile barrier in the packaging of the affected product may be compromised. The outer pouch seal may be…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K153530Bridge Occlusion BalloonSpectranetics, Inc. · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K140273 (Boston Scientific Corp, 2014) has gone furthest, with 17 descendants of its own.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The Equalizer Balloon is indicated for use for temporary vessel occlusion in applications including arteriography, preoperative occlusion, emergency control of hemorrhage, chemotherapeutic drug infusion and renal opacification procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K021721: Equalizer Balloon Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K021721

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026