Over the Wire Embolectomy Catheter, Model 2302

K022145510(k) TraditionalCleared Jan 14, 2003196-day reviewClass II21 CFR 870.5150Recalls 11 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

LeMaitre Vascular, Inc. · DXE Catheter, Embolectomy

Cleared on Jan 14, 2003 after a 196-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 1 of the 53 devices in its family, counting itself, has a Class I recall. 52 later devices descend from it. It has been recalled 11 times, including a Class I recall.

Decided
Jan 14, 2003
Review
196 days
Class
Class II
Product code
DXE
Recalls
11 (1 Class I)
Predicates
0

52 later devices descend from it, 1 citing it directly. 1 of the 53 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2015: 2 descendants cleared2008: 1 descendant cleared2009: 1 descendant cleared2012: 1 descendant cleared2013: 1 descendant cleared2019: 5 descendants cleared2018: 8 descendants cleared2016: 1 descendant cleared2020: 8 descendants cleared2017: 2 descendants cleared2023: 3 descendants cleared2022: 4 descendants cleared2021: 4 descendants cleared2024: 3 descendants cleared2025: 4 descendants cleared2026: 3 descendants cleared200020052010201520202025
  1. This deviceK022145 Over the Wire Embolectomy Catheter, Model 2302LeMaitre Vascular, Inc. · 2003
  2. Descendant · cites it directly · 51 of its ownK040010 Genesis Medical Interventional F.a.S.T. (Facilitated…Genesis Medical Interventional, Inc. · 2004
  3. Descendant · 2 generations on · 35 of its ownK143563 Infusion Aspiration Catheter SystemInari Medical, Inc. · 2015
  4. Descendant · 2 generations on · 7 of its ownK092623 F.a.S.T. System Sed and CXDGenesis Medical Interventional, Inc. · 2009
  5. Descendant · 3 generations on · 6 of its ownK163549 ClotTriever Thrombectomy SystemInari Medical, Inc. · 2017
  6. Descendant · 3 generations on · 6 of its ownK162970 FlowTriever Retrieval/Aspiration System, Aspiration Guide Catheter…Inari Medical · 2016

Showing 5 of 52 descendants. See all descendants

Descendants cleared per year

2004: 12008: 12009: 12012: 12013: 12015: 22016: 12017: 22018: 82019: 52020: 82021: 42022: 42023: 32024: 32025: 42026: 3
2004high 8 (2018) · low 0 (2005)2026

Most influential descendants

Showing 5 of 52 · See all
  • K040010Genesis Medical Interventional F.a.S.T. (Facilitated Aspiration/Suction Thrombectomy)…Genesis Medical Interventional, Inc. · 2004 · cites it directly51descendants
  • K143563Infusion Aspiration Catheter SystemInari Medical, Inc. · 2015 · 2 generations on35descendants
  • K092623F.a.S.T. System Sed and CXDGenesis Medical Interventional, Inc. · 2009 · 2 generations on7descendants
  • K163549ClotTriever Thrombectomy SystemInari Medical, Inc. · 2017 · 3 generations on6descendants
  • K162970FlowTriever Retrieval/Aspiration System, Aspiration Guide Catheter, Retraction AspiratorInari Medical · 2016 · 3 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K022145Over the Wire Embolectomy Catheter, Model 2302LeMaitre Vascular, Inc.2003—
+1K040010Genesis Medical Interventional F.a.S.T. (Facilitated Aspiration/Suction…Genesis Medical Interventional, Inc.2004—
+2K143563Infusion Aspiration Catheter SystemInari Medical, Inc.2015—
+2K092623F.a.S.T. System Sed and CXDGenesis Medical Interventional, Inc.2009—
+3K163549ClotTriever Thrombectomy SystemInari Medical, Inc.2017—
+3K162970FlowTriever Retrieval/Aspiration System, Aspiration Guide Catheter, Retraction…Inari Medical2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200810
200910
201210
201310
201520
201610
201720
201880
201950
202080
202140
202240
202330
202430
202540
202630
FIG. 1 — Lineage of K022145, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXE device

  1. #1 K240078 Fogarty Thru-Lumen Embolectomy Catheterscore 7.9 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (52)

See all 52

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K040010Genesis Medical Interventional F.a.S.T. (Facilitated Aspiration/Suction Thrombectomy)…Genesis Medical Interventional, Inc.200451

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 28, 2026Class IIZ-0117-2027Under Investigation by firmDevices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
Apr 11, 2025Class IIZ-1750-2025Packaging process controlDuring internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
Mar 23, 2020Class IZ-1567-2020Process controlThere has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.
Oct 21, 2015Class IIZ-0378-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0379-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0383-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0380-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0381-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0382-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0384-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Oct 21, 2015Class IIZ-0385-2016Packaging process controlManufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K040010 (Genesis Medical Interventional, Inc., 2004) has gone furthest, with 51 descendants of its own.

How does its category compare?

All DXE devices

DXE Catheter, Embolectomy: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DXE

5in 2024

2008: 12010: 22015: 12017: 22020: 12021: 12022: 12024: 5
2007high 5 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXE

173in 2024

2024: 173 days median
2007high 173 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXE

1in 2023

2007: 32009: 52010: 12011: 12015: 132021: 4, 4 Class I2022: 32023: 12026 (partial year): 2
2007high 13 (2015) · low 0 (2008)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DXE

306in 2025

2007: 1672008: 1342009: 1242010: 1362011: 1642012: 2002013: 1502014: 1922015: 2662016: 7962017: 7702018: 6152019: 5942020: 5472021: 3332022: 4632023: 4362024: 3772025: 306
2007high 796 (2016) · low 124 (2009)2025

MAUDE reports naming product code DXE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—3167
20081—0134
20090—5124
20102—1136
20110—1164
20120—0200
20130—0150
20140—0192
20151—13266
20160—0796
20172—0770
20180—0615
20190—0594
20201—0547
20211—4 (4)333
20221—3463
20230—1436
202451730377
20250—0306
20260—2—
Cite this page

510(k) Report. “K022145: Over the Wire Embolectomy Catheter, Model 2302.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K022145

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026