510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

Fresenius Capd Stay.Safe Disposable Administration Sets with Stay.Safe Connector

K022412510(k) TraditionalCleared Mar 5, 2003224-day reviewClass II21 CFR 876.5630Recalls 5 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care North America · KDJ Set, Administration, For Peritoneal Dialysis, Disposable

Cleared on Mar 5, 2003 after a 224-day review. The FDA summary names its predicate without a number. 1 of the 3 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 5 times, including 5 Class I recalls.

Decided
Mar 5, 2003
Review
224 days
Class
Class II
Product code
KDJ
Recalls
5 (5 Class I)
Predicates
0

2 later devices descend from it, 1 citing it directly. 1 of the 3 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2018: 1 descendant cleared200020052010201520202025
  1. This deviceK022412 Fresenius Capd Stay.Safe Disposable Administration Sets with…Fresenius Medical Care North America · 2003
  2. Descendant · cites it directly · 1 of its ownK041792 Fresenius Staysafe Patient Connectors, Model 050-87211Fresenius Medical Care North America · 2004
  3. Descendant · 2 generations on · 0 of its ownK173651 Multiple Tubing Segment Set with stay safe PIN Connectors, stay safe…Fresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K041792Fresenius Staysafe Patient Connectors, Model 050-87211Fresenius Medical Care North America · 2004 · cites it directly1descendant
  • K173651Multiple Tubing Segment Set with stay safe PIN Connectors, stay safe Drain SetFresenius Medical Care Renal Therapies Group, LLC · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K022412Fresenius Capd Stay.Safe Disposable Administration Sets with Stay.Safe ConnectorFresenius Medical Care North America2003—
+1K041792Fresenius Staysafe Patient Connectors, Model 050-87211Fresenius Medical Care North America2004—
+2K173651Multiple Tubing Segment Set with stay safe PIN Connectors, stay safe Drain SetFresenius Medical Care Renal Therapies Group, LLC20181

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
201810
FIG. 1 — Lineage of K022412, 2003–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDJ device

  1. #1 K250404 stay•safe® catheter extension set with Safe-…score 9.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—CAPD Safe-LockTMTransfer Sets with Pre-filled Connector
Named without a submission number
—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041792Fresenius Staysafe Patient Connectors, Model 050-87211Fresenius Medical Care North America20041

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 23, 2024Class IZ-1323-2024Device DesignThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
Jan 23, 2024Class IZ-1324-2024Device DesignThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
Jan 23, 2024Class IZ-1326-2024Device DesignThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
Jan 23, 2024Class IZ-1327-2024Device DesignThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
Jan 23, 2024Class IZ-1325-2024Device DesignThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K041792 (Fresenius Medical Care North America, 2004) has gone furthest, with 1 descendant of its own.

How does its category compare?

All KDJ devices

KDJ Set, Administration, For Peritoneal Dialysis, Disposable: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KDJ

1in 2025

2008: 12013: 12015: 22016: 12018: 32019: 12022: 12024: 12025: 1
2007high 3 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDJ

90in 2018

2018: 90 days median
2007high 90 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDJ

21in 2024

2009: 22013: 12015: 12020: 12022: 122023: 12024: 21, 10 Class I
2007high 21 (2024) · low 0 (2007)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDJ

2,460in 2025

2007: 2192008: 4112009: 2332010: 1,8792011: 4,3152012: 4,2942013: 5,6232014: 7,6802015: 6,8702016: 7,0452017: 4,6362018: 3,3792019: 2,4562020: 2,8702021: 3,1672022: 2,9272023: 2,9472024: 3,1132025: 2,460
2007high 7,680 (2014) · low 219 (2007)2025

MAUDE reports naming product code KDJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—0219
20081—0411
20090—2233
20100—01,879
20110—04,315
20120—04,294
20131—15,623
20140—07,680
20152—16,870
20161—07,045
20170—04,636
201839003,379
20191—02,456
20200—12,870
20210—03,167
20221—122,927
20230—12,947
20241—21 (10)3,113
20251—02,460
20260—0—

What is it indicated for?

The Fresenius CAPD stayesafe@Disposable Administration Sets with stayesafe@Connector is indicated for use in end stage acute and chronic renal disease.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K022412: Fresenius Capd Stay.Safe Disposable Administration Sets with Stay.Safe Connector.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K022412

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026