510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Restoration Modular System

K022549510(k) TraditionalCleared Sep 23, 200253-day reviewClass II21 CFR 888.3353Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Howmedica Osteonics Corp. · LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Cleared on Sep 23, 2002 as substantially equivalent to K013106 (Howmedica Osteonics Corp., 2002) after a 53-day review. Its family tree reaches back 4 generations to 1994. 54 later devices descend from it. It has been recalled 3 times.

Decided
Sep 23, 2002
Review
53 days
Class
Class II
Product code
LZO
Recalls
3
Predicates
1

Its longest line of descent runs back 4 generations to K921308 (1994). 14 ancestors in all. 54 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1997: 1 other ancestor1996: 3 other ancestors1995: 4 other ancestors1994: 1 other ancestor2006: 1 descendant cleared2021: 3 descendants cleared2020: 4 descendants cleared2025: 9 descendants cleared2003: 1 descendant cleared2009: 1 descendant cleared2012: 4 descendants cleared2015: 5 descendants cleared2010: 2 descendants cleared2013: 1 descendant cleared2016: 4 descendants cleared2022: 3 descendants cleared2014: 2 descendants cleared2019: 3 descendants cleared2023: 2 descendants cleared2017: 4 descendants cleared2024: 2 descendants cleared2026: 3 descendants clearedK921308 · HG Multilock Hip Prosthesis W/ TI-Nidium · Zimmer, Inc. · 1994 · recalledK972893 · Partnership Revision Femoral Stems · Howmedica Corp. · 1997K983404 · Type 3 Femoral Components · Howmedica Corp. · 1998 · recalledK013106 · 2 Piece Modular Hip Stem · Howmedica Osteonics Corp. · 2002 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 8 other ancestors 1 generation forward: 6 descendants2 generations forward: 10 descendants3 generations forward: 17 descendants4 generations forward: 18 descendants5 generations forward: 3 descendants K013106 · 2 Piece Modular Hip Stem · Howmedica Osteonics Corp. · 2002 · recalledK983404 · Type 3 Femoral Components · Howmedica Corp. · 1998 · recalledK972893 · Partnership Revision Femoral Stems · Howmedica Corp. · 1997K921308 · HG Multilock Hip Prosthesis W/ TI-Nidium · Zimmer, Inc. · 1994 · recalled0+5
  1. Root · 4 generations backK921308 HG Multilock Hip Prosthesis W/ TI-NidiumZimmer, Inc. · 1994 · Recalled
  2. 3 generations backK972893 Partnership Revision Femoral StemsHowmedica Corp. · 1997
  3. 2 generations backK983404 Type 3 Femoral ComponentsHowmedica Corp. · 1998 · Recalled
  4. 1 generation backK013106 2 Piece Modular Hip StemHowmedica Osteonics Corp. · 2002 · Recalled
  5. This deviceK022549 Restoration Modular SystemHowmedica Osteonics Corp. · 2002
  6. Descendant · cites it directly · 37 of its ownK090757 Biomet Modular Femoral Revision SystemBiomet, Inc. · 2009 · Recalled
  7. Descendant · 2 generations on · 27 of its ownK092331 Djo Surgical Revision Femoral Hip System, Model 428-14/24-140/200…Encore Medical L.P. · 2010 · Recalled
  8. Descendant · cites it directly · 24 of its ownK032380 Wagner Cone ProsthesisCenterpulse Orthopedics, Inc. · 2003
  9. Descendant · 2 generations on · 13 of its ownK112158 Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups…Lima Corporate S.P.A. · 2012
  10. Descendant · 2 generations on · 9 of its ownK113556 Wagner Cone Prosthesis SystemZimmer GmbH · 2012 · Recalled

Showing 5 of 54 descendants. See all descendants

Descendants cleared per year

2003: 12006: 12009: 12010: 22012: 42013: 12014: 22015: 52016: 42017: 42019: 32020: 42021: 32022: 32023: 22024: 22025: 92026: 3
2003high 9 (2025) · low 0 (2004)2026

Most influential descendants

Showing 5 of 54 · See all
  • K090757Biomet Modular Femoral Revision SystemBiomet, Inc. · 2009 · cites it directly37descendants
  • K092331Djo Surgical Revision Femoral Hip System, Model 428-14/24-140/200; 428-00-050/110;Encore Medical L.P. · 2010 · 2 generations on27descendants
  • K032380Wagner Cone ProsthesisCenterpulse Orthopedics, Inc. · 2003 · cites it directly24descendants
  • K112158Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups, Lock Bipolar HeadsLima Corporate S.P.A. · 2012 · 2 generations on13descendants
  • K113556Wagner Cone Prosthesis SystemZimmer GmbH · 2012 · 2 generations on9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K921308HG Multilock Hip Prosthesis W/ TI-NidiumZimmer, Inc.19941
−3K972893Partnership Revision Femoral StemsHowmedica Corp.1997—
−2K983404Type 3 Femoral ComponentsHowmedica Corp.19981
−1K0131062 Piece Modular Hip StemHowmedica Osteonics Corp.200237
0 (this device)K022549Restoration Modular SystemHowmedica Osteonics Corp.2002—
+1K090757Biomet Modular Femoral Revision SystemBiomet, Inc.200918
+2K092331Djo Surgical Revision Femoral Hip System, Model 428-14/24-140/200…Encore Medical L.P.20101
+1K032380Wagner Cone ProsthesisCenterpulse Orthopedics, Inc.2003—
+2K112158Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups, Lock…Lima Corporate S.P.A.2012—
+2K113556Wagner Cone Prosthesis SystemZimmer GmbH20121

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199540
199630
199710
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200610
200910
201020
201240
201310
201420
201550
201640
201740
201930
202040
202130
202230
202320
202420
202590
202630
FIG. 1 — Lineage of K022549, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZO device

  1. #1 K192827 Quadra-Pscore 19.3 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K0131062 Piece Modular Hip StemHowmedica Osteonics Corp.200237

Also cleared against K013106: 5 other devices →

What descends from it? (54)

See all 54

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090757Biomet Modular Femoral Revision SystemBiomet, Inc.200937
K032380Wagner Cone ProsthesisCenterpulse Orthopedics, Inc.200324
K213129Restoration Modular 115mm Conical Distal StemHowmedica Osteonics Corp., Dba Stryker Orthopaedics20218
K062383Unisyn HA Plasma Modular HipHayes Medical, Inc.20061
K242315Resolve Modular Revision Hip StemUnited Orthopedic Corporation20250
K193233Restoration® Modular Hip SystemHowmedica Osteonics Corp. (Aka Stryker Orthopaedics)20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 2, 2012Class IIZ-2487-2012Device DesignStryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.
Apr 2, 2009Not classifiedZ-1642-2009Process controlPackaging burst test values for the Restoration Modular Cone Body family were found to be incorrectly transferred from the finalized validation to the work specification. Testing showed that burst test results for recalled products fell…
Nov 21, 2008Not classifiedZ-0884-2009Labeling mix-upsStryker became aware of a lot for lot mix up.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 3.3 years on median.
  • 8 of the 14 ancestors have a recall of any class.
  • Of its descendants, K090757 (Biomet, Inc., 2009) has gone furthest, with 37 descendants of its own.
  • 5 other devices also cite its predicate K013106.

How does its category compare?

All LZO devices

LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LZO

14in 2025

2007: 92008: 112009: 72010: 152011: 132012: 142013: 262014: 132015: 122016: 102017: 162018: 182019: 282020: 182021: 122022: 112023: 102024: 112025: 142026 (partial year): 9
2007high 28 (2019) · low 7 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZO

119in 2025

2007: 87 days median2008: 63 days median2009: 83 days median2010: 78 days median2011: 87 days median2012: 98 days median2013: 100 days median2014: 85 days median2015: 115 days median2016: 177 days median2017: 257 days median2018: 163 days median2019: 109 days median2020: 112 days median2021: 170 days median2022: 279 days median2023: 122 days median2024: 111 days median2025: 119 days median2026 (partial year): 102 days median
2007high 279 (2022) · low 63 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZO

1in 2025

2007: 22008: 132009: 422011: 102012: 142014: 12015: 12016: 92017: 52018: 22019: 42020: 32022: 42023: 22024: 12025: 1
2007high 42 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZO

3,244in 2025

2007: 1382008: 1252009: 1892010: 3942011: 5442012: 9632013: 1,3702014: 1,4772015: 1,3482016: 1,7282017: 2,1652018: 3,1332019: 3,0662020: 1,9632021: 2,1992022: 1,9852023: 2,0662024: 2,5862025: 3,244
2007high 3,244 (2025) · low 125 (2008)2025

MAUDE reports naming product code LZO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20079872138
2008116313125
200978342189
201015780394
2011138710544
2012149814963
20132610001,370
2014138511,477
20151211511,348
20161017791,728
20171625752,165
20181816323,133
20192810943,066
20201811231,963
20211217002,199
20221127941,985
20231012222,066
20241111112,586
20251411913,244
202691020—

What is it indicated for?

The RestorationTMModular System is intended to be used for primary or revision total hip arthroplasty, as well as in the presence of severe proximal bone loss. The stems are intended to be used with Howmedica Osteonics femoral heads, unipolar and bipolar components, and acetabular Components. These femoral stems are designed to be press fit into the proximal femur.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K022549: Restoration Modular System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K022549

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026