# K022895: Elcam Stopcocks and Manifolds

- Pathway: 510(k) (Traditional)
- Applicant: Medical Device Consultants, Inc.
- Decided: Oct 18, 2002 after 45 days
- Product code: FMG (Stopcock, I.V. Set)
- Device class: 2
- Regulation: 21 CFR 880.5440
- Recalls of this device: 14 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K862691 (1986). 1 ancestor in all.
29 later devices descend from it; 5 cite it directly.
0 of 31 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K862691: Stopcock, Ram Development (1986) Summary page 1.

## Most influential descendants
- K032393: Induction and Sampling Manifold (or Stopcock) (2003), 18 descendants of its own
- K033793: Cleaner II Rotational Thrombectomy System (2003), 12 descendants of its own
- K040252: Cleaner Rotational Thrombectomy System Kit (2004), 11 descendants of its own
- K060904: Cleaner Rotational Thrombectomy System (2006), 10 descendants of its own
- K091029: Cleaner Rotational Thrombectomy System (9 MM X 65 CM) and (9 MM X 120 CM) (2010), 9 descendants of its own

## Recalls of this device
- Oct 29, 2025 · Class II · Z-0642-2026: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
- Oct 29, 2025 · Class II · Z-0644-2026: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
- Oct 29, 2025 · Class II · Z-0641-2026: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
- Oct 29, 2025 · Class II · Z-0643-2026: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
- Oct 29, 2025 · Class II · Z-0717-2026: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
- Aug 8, 2024 · Class II · Z-3144-2024: Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.
- Nov 20, 2019 · Class II · Z-0936-2020: Potential for leakage at the joint between the blood filters and tubing
- Nov 20, 2019 · Class II · Z-0938-2020: Potential for leakage at the joint between the blood filters and tubing
- Dec 15, 2005 · Not classified · Z-0654-06: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Dec 15, 2005 · Not classified · Z-0655-06: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Dec 15, 2005 · Not classified · Z-0656-06: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Dec 15, 2005 · Not classified · Z-0646-06: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Dec 15, 2005 · Not classified · Z-0653-06: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Apr 21, 2005 · Not classified · Z-0943-05: There is an unsecure connection between the male Luer and the stopcock on the I.V. extension set.

## Indications for use
Elcam Stopcocks and Manifolds are indicated for fluid flow directional control and for providing access port(s) for administration of solutions. Typical uses include pressure monitoring, intravenous fluid administration and transfusion.