# K023369: Access Total T4 Assay

- Pathway: 510(k) (Traditional)
- Applicant: Beckman Coulter, Inc.
- Decided: Dec 9, 2002 after 62 days
- Product code: KLI (Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine)
- Device class: 2
- Regulation: 21 CFR 862.1700
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K926475 (1993). 1 ancestor in all.
No later device cites it as a predicate yet.
0 of 2 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K926475: Access (R) Thyroxine Assay, Bio-Rad Laboratories, Inc. (1993) Summary page 1.

## Recalls of this device
- Mar 16, 2026 · Class II · Z-1922-2026: Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient resul
- Feb 8, 2024 · Class II · Z-1303-2024: Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
- Jul 27, 2011 · Class II · Z-0613-2013: A recall communication was initiated because Beckman has identified that Access Total T4 calibrator lots 019818, 021654, and 024072 do not meet their 12 month expiration date claim.
- Mar 10, 2011 · Not classified · Z-1949-2011: The recall was initiated because the concentration of total T4 in samples from pregnant women is erroneously low (20%) when measured using the Access Total T4 assay. Erroneously low results which are 
- Aug 26, 2010 · Not classified · Z-3027-2011: The recall was initiated because Beckman Coulter has confirmed that the stability claims for the Access Total T4 calibrators may not be met. This issue is associated with calibrator lots 917074 and 00

## Indications for use
The Access Total T4 assay provides in vitro quantitative measurement of total thyroxine (T4) levels in human serum or plasma. The Access Total T4 assay is indicated for use with patients where an evaluation of their thyroid status is desired.