510(k) · Cleared
Recalled · 2011Cardio Vations Steerable Coronary Sinus Catheter
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
ETHICON, Inc. · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Cleared on Dec 11, 2002 after a 20-day review. The FDA summary names its predicate without a number. It has been recalled 2 times.
- Decided
- Dec 11, 2002
- Review
- 20 days
- Class
- Class II
- Product code
- DWF
- Recalls
- 2
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DWF device
- #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Endocoronary Sinus Catheter Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 15, 2011 | Not classified | Z-2777-2011 | Employee error | Possibility of separation of the introducer sheath from the hub on Coronary Sinus Catheters. |
| Dec 3, 2010 | Not classified | Z-0992-2011 | Device Design | Defective product may result in the silicone valve becoming dislodged from the introducer which can result in injury. |
No ancestor or descendant has a Class I recall.
How does its category compare?
All DWF devicesDWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DWF
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DWF
89in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DWF
4in 2024
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DWF
1,020in 2025
MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 25 | 121 |
| 2008 | 6 | 95 | 25 | 126 |
| 2009 | 8 | 148 | 21 | 99 |
| 2010 | 3 | 62 | 26 | 197 |
| 2011 | 5 | 85 | 13 | 212 |
| 2012 | 10 | 110 | 37 | 233 |
| 2013 | 9 | 110 | 14 (1) | 238 |
| 2014 | 14 | 110 | 6 | 325 |
| 2015 | 9 | 46 | 3 (1) | 218 |
| 2016 | 3 | 32 | 3 | 144 |
| 2017 | 5 | 127 | 2 (1) | 101 |
| 2018 | 5 | 30 | 10 | 118 |
| 2019 | 3 | 120 | 2 | 179 |
| 2020 | 3 | 30 | 0 | 142 |
| 2021 | 5 | 358 | 3 | 162 |
| 2022 | 2 | — | 0 | 276 |
| 2023 | 2 | — | 0 | 470 |
| 2024 | 3 | 29 | 4 | 401 |
| 2025 | 4 | 89 | 0 | 1,020 |
| 2026 | 8 | 81 | 20 | — |
What is it indicated for?
The Steerable Coronary Sinus Catheter is indicated for occlusion of the coronary sinus, delivery of cardioplegic solution, and monitoring of coronary sinus pressure during cardiopulmonary bypass.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K023880 · printed Nov 5, 2026
Cite this page
510(k) Report. “K023880: Cardio Vations Steerable Coronary Sinus Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K023880
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026