Gyrus Open Forceps

K024286510(k) TraditionalCleared Nov 17, 2003329-day reviewClass II21 CFR 878.4400Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gyrus Medical, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Nov 17, 2003 as substantially equivalent to K000496 (Everest Medical Corp., 2000) and 2 other predicates after a 329-day review. Its family tree reaches back 2 generations to 1994. 14 later devices descend from it. It has been recalled 6 times.

Decided
Nov 17, 2003
Review
329 days
Class
Class II
Product code
GEI
Recalls
6
Predicates
3

Its longest line of descent runs back 2 generations to K932293 (1994). 9 ancestors in all. 14 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2000: 1 other ancestor1996: 1 other ancestor1995: 4 other ancestors2006: 1 descendant cleared2008: 3 descendants cleared2011: 1 descendant cleared2016: 2 descendants cleared2015: 1 descendant cleared2014: 1 descendant cleared2019: 1 descendant cleared2023: 2 descendants cleared2012: 1 descendant cleared2017: 1 descendant clearedK932293 · Cabot Medical Bipolar Cutting Forceps · Cabot Medical Corp. · 1994 · recalledK981916 · Ligasure Vessel Sealing System · Valleylab, Inc. · 19981990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 5 other ancestors 1 generation forward: 4 descendants2 generations forward: 2 descendants3 generations forward: 6 descendants4 generations forward: 2 descendants K981916 · Ligasure Vessel Sealing System · Valleylab, Inc. · 1998K932293 · Cabot Medical Bipolar Cutting Forceps · Cabot Medical Corp. · 1994 · recalled0
  1. Root · 2 generations backK932293 Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp. · 1994 · Recalled
  2. 1 generation backK981916 Ligasure Vessel Sealing SystemValleylab, Inc. · 1998
  3. This deviceK024286 Gyrus Open ForcepsGyrus Medical, Inc. · 2003
  4. Descendant · cites it directly · 7 of its ownK081503 Aed Monopolar Lap AccessoriesNational Advanced Endoscopy Devices, Inc. · 2008
  5. Descendant · 2 generations on · 6 of its ownK103508 Unmiax Laparoscopic InstrumentUnimax Medical Systems, Inc. · 2011
  6. Descendant · cites it directly · 4 of its ownK061975 Plasmaseal Open Forceps, Models 2103PK and 91701PKAcmi Corporation · 2006
  7. Descendant · cites it directly · 3 of its ownK081766 PKS Omni InstrumentGyrus-Acmi L.P. · 2008 · Recalled
  8. Descendant · 2 generations on · 2 of its ownK161825 PK AIMGyrus Acmi, Inc. · 2016

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2006: 12008: 32011: 12012: 12014: 12015: 12016: 22017: 12019: 12023: 2
2006high 3 (2008) · low 0 (2007)2026

Most influential descendants

Showing 5 of 14 · See all
  • K081503Aed Monopolar Lap AccessoriesNational Advanced Endoscopy Devices, Inc. · 2008 · cites it directly7descendants
  • K103508Unmiax Laparoscopic InstrumentUnimax Medical Systems, Inc. · 2011 · 2 generations on6descendants
  • K061975Plasmaseal Open Forceps, Models 2103PK and 91701PKAcmi Corporation · 2006 · cites it directly4descendants
  • K081766PKS Omni InstrumentGyrus-Acmi L.P. · 2008 · cites it directly3descendants
  • K161825PK AIMGyrus Acmi, Inc. · 2016 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K932293Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp.19943
−1K981916Ligasure Vessel Sealing SystemValleylab, Inc.1998—
0 (this device)K024286Gyrus Open ForcepsGyrus Medical, Inc.2003—
+1K081503Aed Monopolar Lap AccessoriesNational Advanced Endoscopy Devices, Inc.2008—
+2K103508Unmiax Laparoscopic InstrumentUnimax Medical Systems, Inc.2011—
+1K061975Plasmaseal Open Forceps, Models 2103PK and 91701PKAcmi Corporation2006—
+1K081766PKS Omni InstrumentGyrus-Acmi L.P.20081
+2K161825PK AIMGyrus Acmi, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199540
199610
200010
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200830
201110
201210
201410
201510
201620
201710
201910
202320
FIG. 1 — Lineage of K024286, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K000496Evershears Bipolar Scissors
Summary page 1
…Proprietary Name: Gyrus OPEN FORCEPS The device is a Class I1 medical device. The Gyrus OPEN FORCEPS is a modification to the predicate device cleared under K000496. The Open Forceps is similar in construction and in component materials when compared to the predicate device. The indications fo…
Everest Medical Corp.2000—
*K981916Ligasure Vessel Sealing System
Summary page 1
…s is similar in construction and in component materials when compared to the predicate device. The indications for use are similar to the predicate devices cleared under K98 1916 and K010010. The device consists of grasping jaws connected to a handle. The handle has finger loops integrated into it to allow…
Valleylab, Inc.1998—
*K010010Valleylab Ligasure Precise Instrument Vessel Sealing System-Model # LS1200 & Sligaure…
Summary page 1
…in construction and in component materials when compared to the predicate device. The indications for use are similar to the predicate devices cleared under K98 1916 and K010010. The device consists of grasping jaws connected to a handle. The handle has finger loops integrated into it to allow the physicia…
Valleylab, Inc.2001—

Also cleared against K000496: 2 other devices →

What descends from it? (14)

See all 14

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081503Aed Monopolar Lap AccessoriesNational Advanced Endoscopy Devices, Inc.20087
K061975Plasmaseal Open Forceps, Models 2103PK and 91701PKAcmi Corporation20064
K081766PKS Omni InstrumentGyrus-Acmi L.P.20083
K071759LTC Vad System, Models #Vad.400 (Electrosurgical Generator); PCS-900 and PCC-900 (Hand…Live Tissue Connect, Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 18, 2009Not classifiedZ-0632-2010Environmental controlGyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field.…
Aug 18, 2009Not classifiedZ-0633-2010Environmental controlGyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field.…
Aug 18, 2009Not classifiedZ-0634-2010Environmental controlGyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field.…
Aug 18, 2009Not classifiedZ-0635-2010Environmental controlGyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field.…
Aug 18, 2009Not classifiedZ-0637-2010Environmental controlGyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field.…
Aug 18, 2009Not classifiedZ-0636-2010Environmental controlGyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 5.2 years on median.
  • 1 of the 9 ancestors has a recall of any class.
  • Of its descendants, K081503 (National Advanced Endoscopy Devices, Inc., 2008) has gone furthest, with 7 descendants of its own.
  • 2 other devices also cite its predicate K000496.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

Electrosurgical coagulation, mechanical grasping and dissection of tissue, and sealing of

vessels up to 7mm, during the performance of open general surgical procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K024286: Gyrus Open Forceps.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K024286

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026