510(k) Report

510(k) · Cleared

Cleared

2 BASIC

K030175510(k) TraditionalCleared Apr 17, 200390-day reviewClass II21 CFR 892.5050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Capintec, Inc. · IYE Accelerator, Linear, Medical

Cleared on Apr 17, 2003 as substantially equivalent to K912249 (Theta Systems, Inc., 1991) and 2 other predicates after a 90-day review. Its family tree reaches back 2 generations to 1991.

Decided
Apr 17, 2003
Review
90 days
Class
Class II
Product code
IYE
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K912250 (1991). 3 ancestors in all, the oldest K912249 (1991). No later device cites it as a predicate yet.

Other ancestors, by year cleared 1991: 1 other ancestorK912250 · Isorad (TM) Integral Buildup Photon Diodes · Theta Systems, Inc. · 1991K980826 · Equidose Solid State Diode Detectors · Cii Mastek, Inc. · 19981990199520002005
Other ancestors, by generation 1 generation back: 1 other ancestor K980826 · Equidose Solid State Diode Detectors · Cii Mastek, Inc. · 1998K912250 · Isorad (TM) Integral Buildup Photon Diodes · Theta Systems, Inc. · 19910
  1. Root · 2 generations backK912250 Isorad (TM) Integral Buildup Photon DiodesTheta Systems, Inc. · 1991
  2. 1 generation backK980826 Equidose Solid State Diode DetectorsCii Mastek, Inc. · 1998
  3. This deviceK030175 2 BASICCapintec, Inc. · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K912250Isorad (TM) Integral Buildup Photon DiodesTheta Systems, Inc.1991—
−1K980826Equidose Solid State Diode DetectorsCii Mastek, Inc.1998—
0 (this device)K0301752 BASICCapintec, Inc.2003—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
FIG. 1 — Lineage of K030175, 1991–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K912249Patient Dose Monitor
Summary page 2
…recognized PC to be used with application software. Predicate Device: Predicate Devices submitted by Theta Systems, Inc. Model PDM/2 Patient Dose Monitor 510(k) Number K912249 and lsorad Integral Buildup Photon Diodes 510(k) Number K912250. The 2 Basic uses the Equidose Diode Detector Design submitted by…
Theta Systems, Inc.1991—
*K912250Isorad (TM) Integral Buildup Photon Diodes
Summary page 2
…Device: Predicate Devices submitted by Theta Systems, Inc. Model PDM/2 Patient Dose Monitor 510(k) Number K912249 and lsorad Integral Buildup Photon Diodes 510(k) Number K912250. The 2 Basic uses the Equidose Diode Detector Design submitted by MasTek DEM, Inc. 510(k) Number K980826 The 2 Basic meets the f…
Theta Systems, Inc.1991—
*K980826Equidose Solid State Diode Detectors
Summary page 2
…K912249 and lsorad Integral Buildup Photon Diodes 510(k) Number K912250. The 2 Basic uses the Equidose Diode Detector Design submitted by MasTek DEM, Inc. 510(k) Number K980826 The 2 Basic meets the following Safety Standards Medical Electrical Equipment, Part 1 General Requirements - UL2601-1:2nd Medica…
Cii Mastek, Inc.1998—

Also cleared against K912249: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 6.6 years on median.
  • None of the 3 ancestors has a recall.
  • 1 other device also cites its predicate K912249.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The 2 Basic is a solid state diode detector system coupled with a precision electrometer and PC for radiation measurement, data storage and display. It is intended to be used by trained medical physicists for the measurement of radiation output from radiation therapy treatment machines. The 2 Basic can be used with direct patient contact for entrance skin measurements, or in air, water, or other suitable phantoms materials for beam output measurements.

Measurement data is used to verify and document the beam characteristics of radiation therapy treatment machines for quality assurance programs and to very therapeutic doses delivered to patients during treatments.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K030175: 2 BASIC.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K030175

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026