510(k) Report

510(k) · Cleared

Recalled · 2007

Vitros Immunodiagnostic Products Troponin I Reagent Pack and Calibrators

K031031510(k) TraditionalCleared Sep 29, 2003181-day reviewClass II21 CFR 862.1215Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · MMI Immunoassay Method, Troponin Subunit

Cleared on Sep 29, 2003 as substantially equivalent to K973650 (Dade Chemistry Systems, Inc., 1997) after a 181-day review. Its family tree reaches back 1 generation to 1997. It has been recalled 2 times.

Decided
Sep 29, 2003
Review
181 days
Class
Class II
Product code
MMI
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K973650 (1997). No later device cites it as a predicate yet.

K973650 · Troponin (Trop) Method · Dade Chemistry Systems, Inc. · 199719961998200020022004
  1. Root · 1 generation backK973650 Troponin (Trop) MethodDade Chemistry Systems, Inc. · 1997
  2. This deviceK031031 Vitros Immunodiagnostic Products Troponin I Reagent Pack and…Ortho-Clinical Diagnostics, Inc. · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K973650Troponin (Trop) MethodDade Chemistry Systems, Inc.1997—
0 (this device)K031031Vitros Immunodiagnostic Products Troponin I Reagent Pack and CalibratorsOrtho-Clinical Diagnostics, Inc.2003—
FIG. 1 — Lineage of K031031, 1997–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MMI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MMI device

  1. #1 K253051 i-STAT hs-TnI cartridge with the i-STAT…score 13.0 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973650Troponin (Trop) Method
Summary page 2
…oducts Troponin I Calibrators are substantially equivalent to the DADE Dimension RxL Cardiac Troponin-I (TROP) Method, which was cleared by the FDA (K973650) for IVD use. Additional data is being provided on hemoglobin interference and precision. A summary of this data is…
Dade Chemistry Systems, Inc.1997—

Also cleared against K973650: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 27, 2007Not classifiedZ-0820-2007Nonconforming Material/ComponentQuality control and patient results can be lower than expected when the true Troponin I value is less than or equal to 0.2 ng/mL which may result in the device failing to detect myocardial injury.
Jun 7, 2004Not classifiedZ-1053-04OtherRandom occurrence of false positive test results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 3 other devices also cite its predicate K973650.

How does its category compare?

All MMI devices

MMI Immunoassay Method, Troponin Subunit: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MMI

4in 2025

2007: 32008: 12009: 12010: 22011: 12013: 22017: 12018: 42019: 32021: 12022: 12023: 32024: 22025: 42026 (partial year): 1
2007high 4 (2018) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MMI

266in 2025

2007: 190 days median2018: 344 days median2019: 88 days median2023: 270 days median2025: 266 days median
2007high 344 (2018) · low 88 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MMI

1in 2024

2007: 62008: 22009: 32010: 52011: 12012: 62014: 12015: 12019: 12020: 12021: 12024: 1
2007high 6 (2007) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MMI

805in 2025

2007: 3522008: 8372009: 1,0102010: 1,3352011: 2,4062012: 1,8412013: 1,3942014: 1,1502015: 8312016: 4872017: 6182018: 4872019: 5212020: 3872021: 3562022: 5072023: 7482024: 9212025: 805
2007high 2,406 (2011) · low 352 (2007)2025

MAUDE reports naming product code MMI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731906352
20081—2837
20091—31,010
20102—51,335
20111—12,406
20120—61,841
20132—01,394
20140—11,150
20150—1831
20160—0487
20171—0618
201843440487
20193881521
20200—1387
20211—1356
20221—0507
202332700748
20242—1921
202542660805
20261—0—

What is it indicated for?

For the in vitro quantitative measurement of Troponin I (cTnI) in human heparin plasma to aid in the diagnosis of myocardial infarction.

For use in the calibration of the Vitros Immunodiagnostic System for the quantitative measurement of cardiac Troponin I (cTnI) in human hepan

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031031: Vitros Immunodiagnostic Products Troponin I Reagent Pack and Calibrators.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031031

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026