# K031038: Advia Centaur B-Type Natriuretic Peptide (BNP) Assay

- Pathway: 510(k) (Traditional)
- Applicant: Bayer Healthcare, LLC
- Decided: Jun 23, 2003 after 83 days
- Product code: NBC (Test, Natriuretic Peptide)
- Device class: 2
- Regulation: 21 CFR 862.1117
- Recalls of this device: 6 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K010266 (2001). 1 ancestor in all.
4 later devices descend from it; 3 cite it directly.
0 of 6 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K010266: Triage B-Type Natriuretic Peptide (BNP) Test, Catalog #98000, Biosite Incorporated (2001) Summary page 1.

## Most influential descendants
- K192380: ST Aia-Pack BNP (2020), 1 descendant of its own
- K040425: Acs:180 and Advia Centaur B-Type Natriuretic Peptide (BNP) Assays (2004), 1 descendant of its own
- K211199: ST AIA-PACK BNP Assay (2021), 0 descendants of its own
- K043228: Acs:180 and Advia Centaur B-Type Natriuretic Peptide (BNP) Assays (2005), 0 descendants of its own

## Recalls of this device
- Dec 18, 2017 · Class II · Z-2053-2018: In certain scenarios the ADVIA Centaur CP, XP and XPT immunoassay analyzers may exhibit reduced dilution recovery or become inoperable when used with Multi-Diluent 15 and certain lots of BNP and TSH3R
- Oct 6, 2014 · Class III · Z-0229-2015: Siemens Healthcare Diagnostics has confirmed that approximately 2000 ReadyPacks from the ADVIA Centaur Systems BNP reagent contains an ADVIA Centaur Systems FT4 human-readable side label.
- Jun 23, 2014 · Class II · Z-2149-2014: Complaints of failed calibration.
- Jun 23, 2014 · Class II · Z-2150-2014: Complaints of failed calibration.
- Nov 14, 2012 · Class II · Z-0506-2013: Operating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur syst
- Jul 25, 2012 · Class II · Z-2313-2012: Decrease in onboard dilution recovery when using Multi-Diluent 1 that has been stored on board the ADVIA Centaur and ADVIA Centaur XP systems for for BNP and TSH3 Ultra assays

## Indications for use
For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur@System. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis of heart failure. This assay is not intended for use on any other system.