Motility Visualization System

K031169510(k) TraditionalCleared Dec 12, 2003242-day reviewClass II21 CFR 876.1725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sierra Scientific Instruments, Inc. · FFX System, Gastrointestinal Motility (Electrical)

Cleared on Dec 12, 2003 as substantially equivalent to K011472 (Medtronic Functional Diagnostics A/S, 2001) after a 242-day review. Its family tree reaches back 1 generation to 2001. 15 later devices descend from it. It has been recalled once.

Decided
Dec 12, 2003
Review
242 days
Class
Class II
Product code
FFX
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K011472 (2001). 15 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2015: 2 descendants cleared2009: 1 descendant cleared2012: 1 descendant cleared2007: 1 descendant cleared2010: 1 descendant cleared2019: 1 descendant cleared2016: 1 descendant cleared2017: 1 descendant cleared2018: 1 descendant cleared2024: 2 descendants cleared2025: 2 descendants cleared2021: 1 descendant clearedK011472 · Polygram 98 Anorectal Function Testing Application, Polygram 98 Esophageal Manometry Application, Polygraf ID · Medtronic Functional Diagnostics A/S · 2001200020052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 7 descendants3 generations forward: 4 descendants K011472 · Polygram 98 Anorectal Function Testing Application, Polygram 98 Esophageal Manometry Application, Polygraf ID · Medtronic Functional Diagnostics A/S · 20010
  1. Root · 1 generation backK011472 Polygram 98 Anorectal Function Testing Application, Polygram 98…Medtronic Functional Diagnostics A/S · 2001
  2. This deviceK031169 Motility Visualization SystemSierra Scientific Instruments, Inc. · 2003
  3. Descendant · cites it directly · 7 of its ownK120088 Mcompass Anorectal Manometry SystemMedspira, LLC · 2012 · Recalled
  4. Descendant · 2 generations on · 4 of its ownK161785 THD AnopressThd Spa · 2017
  5. Descendant · 2 generations on · 3 of its ownK180135 THD Anopress with THD SensyProbeThd Spa · 2018
  6. Descendant · cites it directly · 2 of its ownK062222 Unitip High Resolution CatheterUnisensor AG · 2007
  7. Descendant · 2 generations on · 1 of its ownK143031 mcompass Biofeedback Anorectal Manometry SystemMedspira, LLC · 2015 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2007: 12009: 12010: 12012: 12015: 22016: 12017: 12018: 12019: 12021: 12024: 22025: 2
2007high 2 (2015) · low 0 (2008)2026

Most influential descendants

Showing 5 of 15 · See all
  • K120088Mcompass Anorectal Manometry SystemMedspira, LLC · 2012 · cites it directly7descendants
  • K161785THD AnopressThd Spa · 2017 · 2 generations on4descendants
  • K180135THD Anopress with THD SensyProbeThd Spa · 2018 · 2 generations on3descendants
  • K062222Unitip High Resolution CatheterUnisensor AG · 2007 · cites it directly2descendants
  • K143031mcompass Biofeedback Anorectal Manometry SystemMedspira, LLC · 2015 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K011472Polygram 98 Anorectal Function Testing Application, Polygram 98 Esophageal…Medtronic Functional Diagnostics A/S2001—
0 (this device)K031169Motility Visualization SystemSierra Scientific Instruments, Inc.2003—
+1K120088Mcompass Anorectal Manometry SystemMedspira, LLC20121
+2K161785THD AnopressThd Spa2017—
+2K180135THD Anopress with THD SensyProbeThd Spa2018—
+1K062222Unitip High Resolution CatheterUnisensor AG2007—
+2K143031mcompass Biofeedback Anorectal Manometry SystemMedspira, LLC20151

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200910
201010
201210
201520
201610
201710
201810
201910
202110
202420
202520
FIG. 1 — Lineage of K031169, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FFX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FFX device

  1. #1 K262621 Rapid Barostat Bag (RBB) Pump and Catheter…score 12.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011472Polygram 98 Anorectal Function Testing Application, Polygram 98 Esophageal Manometry…
Summary page 1
…name: Gastrointestinal Motility (Electrical) System Classification: Class II Product Code: FFX Substantial equivalence claimed to: i. K011472 - Medtronic, Polygraf ID and Polygram 98 software Description: The Motility Visualization System (MVS) obtains a high resolutio…
Medtronic Functional Diagnostics A/S2001—

Also cleared against K011472: 1 other device →

What descends from it? (15)

See all 15

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120088Mcompass Anorectal Manometry SystemMedspira, LLC20127
K062222Unitip High Resolution CatheterUnisensor AG20072
K142355Vaginal Tactile ImagerAdvanced Tactile Imaging, Inc.20151
K091070Manoscan Motility with Impedance Visualization SystemSierra Scientific Instruments, Inc.20091

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 12, 2011Not classifiedZ-0135-2012Nonconforming Material/ComponentThe recall was initiated after Given Imaging received a test report during a Canadian licensing process where the AR ManoShield exhibited pinholes and did not pass the test, the Field Safety Corrective Action (FSCA) Committee was…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K120088 (Medspira, LLC, 2012) has gone furthest, with 7 descendants of its own.
  • 1 other device also cites its predicate K011472.

How does its category compare?

All FFX devices

FFX System, Gastrointestinal Motility (Electrical): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FFX

1in 2025

2007: 22008: 12009: 22010: 22012: 42013: 12016: 32017: 22018: 12019: 42020: 12023: 32024: 12025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FFX

123in 2023

2012: 120 days median2016: 145 days median2019: 240 days median2023: 123 days median
2007high 240 (2019) · low 120 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FFX

2so far in 2026, the most since 2007

2009: 12011: 12026 (partial year): 2
2007high 2 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FFX

32in 2025

2009: 22010: 62011: 62012: 42013: 42014: 42015: 102016: 152017: 102018: 172019: 272020: 202021: 262022: 312023: 512024: 592025: 32
2007high 59 (2024) · low 0 (2007)2025

MAUDE reports naming product code FFX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20081—00
20092—12
20102—06
20110—16
2012412004
20131—04
20140—04
20150—010
20163145015
20172—010
20181—017
20194240027
20201—020
20210—026
20220—031
20233123051
20241—059
20251—032
20261—2—

What is it indicated for?

The Motility Visualization System obtains a high resolution mapping of pressures within tubular organs of the gastrointestinal tract. These organs include the esophagus from the pharyngeal region to the stomach, the proximal gut (stomach/duodenum), and the anus/rectum. It is used in a medical clinical setting to sense the pressures and store the corresponding data. The system also provides analysis information. The real time data as well as the analysis information can then be viewed by a physician for diagnostic and analysis purposes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031169: Motility Visualization System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031169

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.