510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Access Thyroglobulin Reagents on the Access Immunoassay Systems

K031269510(k) SpecialCleared May 2, 200311-day reviewClass II21 CFR 866.6010Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · MSW System, Test, Thyroglobulin

Cleared on May 2, 2003 as substantially equivalent to K002905 (Beckman Coulter, Inc., 2000) after an 11-day review. Its family tree reaches back 1 generation to 2000. It has been recalled 6 times.

Decided
May 2, 2003
Review
11 days
Class
Class II
Product code
MSW
Recalls
6
Predicates
1

Its longest line of descent runs back 1 generation to K002905 (2000). No later device cites it as a predicate yet.

  1. Root · 1 generation backK002905 Access Thyroglobulin Reagents on the Access AnalyzerBeckman Coulter, Inc. · 2000 · Recalled
  2. This deviceK031269 Access Thyroglobulin Reagents on the Access Immunoassay SystemsBeckman Coulter, Inc. · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K002905Access Thyroglobulin Reagents on the Access AnalyzerBeckman Coulter, Inc.20004
0 (this device)K031269Access Thyroglobulin Reagents on the Access Immunoassay SystemsBeckman Coulter, Inc.2003—
FIG. 1 — Lineage of K031269, 2000–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSW device

  1. #1 K220972 Access Thyroglobulinscore 12.0 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002905Access Thyroglobulin Reagents on the Access Analyzer
Summary page 1
…Class I Class II Legally Marketed Access@Thyroglobulin Reagents on the AccessB (Unmodified) Device Immunoassay Systems (K002905, cleared 10/19/00) Device Description The AccessB Thyroglobulin reagents consist of reagent packs, calibrators, substrate and w…
Beckman Coulter, Inc.20004
†K023764Unicel Dxi 800 Access Immunoassay System, Model 973100Beckman Coulter, Inc.200331

Also cleared against K002905: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 10, 2025Class IIZ-2322-2025Under Investigation by firmAccess Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy…
Jul 6, 2023Class IIZ-0100-2024OtherIt has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.
Apr 30, 2019Class IIZ-2240-2020OtherThe recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687…
May 9, 2017Class IIZ-2636-2017Device DesignThe Access immunoassays are susceptible to biotin interference.
Jul 28, 2015Class IIZ-2733-2015Under Investigation by firmBeckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.
Oct 18, 2010Not classifiedZ-2215-2011OtherThe recall was initiated because In addition to Access Tg Sample Diluent, an additional diluent, Access Wash Buffer II, was added to Thyroglobulin instructions for use (IFU) for evaluation of samples having Tg levels above the reportable…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 2 other devices also cite its predicate K002905.

How does its category compare?

All MSW devices

MSW System, Test, Thyroglobulin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSW

1in 2025

2023: 22024: 22025: 1
2007high 2 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSW

Too few MSW 510(k)s in any year since 2007 to give a median.

Recalls per year · MSW

1in 2023

2010: 12022: 22023: 1
2007high 2 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MSW

3in 2025

2012: 22013: 22015: 12016: 12017: 32018: 52019: 22022: 572023: 42024: 32025: 3
2007high 57 (2022) · low 0 (2007)2025

MAUDE reports naming product code MSW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—10
20110—00
20120—02
20130—02
20140—00
20150—01
20160—01
20170—03
20180—05
20190—02
20200—00
20210—00
20220—257
20232—14
20242—03
20251—03
20260—0—

What is it indicated for?

The Access Thyroglobulin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin levels in human serum and plasma, using the Access Immunoassay Systems. This device is intended to aid in the monitoring for the presence of local and metastatic thyroid tissue in patients who have had thyroid gland ablation (using thyroid surgery with or without radioactivity) and who lack serum thyroglobulin antibodies.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031269: Access Thyroglobulin Reagents on the Access Immunoassay Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031269

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026