510(k) Report

510(k) · Cleared

Cleared

Puritan Bennett, Goodknight 420 Evolution, Model M-113903-US

K031470510(k) SpecialCleared Jun 5, 200327-day reviewClass II21 CFR 868.5905

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tyco Healthcare Nancy · BZD Ventilator, Non-Continuous (Respirator)

Cleared on Jun 5, 2003 as substantially equivalent to K020886 (Tyco Healthcare Nancy, 2002) and 1 other predicate after a 27-day review. Its family tree reaches back 5 generations to 1990.

Decided
Jun 5, 2003
Review
27 days
Class
Class II
Product code
BZD
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K903764 (1990). 9 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK903764 Companion 318 Nasal Cpap SystemPuritan Bennett Corp. · 1990
  2. 4 generations backK960518 The Puritan-Bennett 318 Plus Nasal Cpap SystemPuritan Bennett Corp. · 1996
  3. 3 generations backK980365 Goodknight 418 Cpap SystemNellcor Puritan Bennett, Inc. · 1998
  4. 2 generations backK991150 Nellcor Puritan Bennett, Goodknight 418GNellcor Puritan Bennett France Dvpt. · 1999
  5. 1 generation backK993584 Mallinckrodt, Goodknight 418A, Goodknight 418P, Silverlining 2Nellcor Puritan Bennett France Dvpt. · 2000
  6. This deviceK031470 Puritan Bennett, Goodknight 420 Evolution, Model M-113903-USTyco Healthcare Nancy · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K903764Companion 318 Nasal Cpap SystemPuritan Bennett Corp.1990—
−4K960518The Puritan-Bennett 318 Plus Nasal Cpap SystemPuritan Bennett Corp.1996—
−3K980365Goodknight 418 Cpap SystemNellcor Puritan Bennett, Inc.1998—
−2K991150Nellcor Puritan Bennett, Goodknight 418GNellcor Puritan Bennett France Dvpt.1999—
−1K993584Mallinckrodt, Goodknight 418A, Goodknight 418P, Silverlining 2Nellcor Puritan Bennett France Dvpt.2000—
0 (this device)K031470Puritan Bennett, Goodknight 420 Evolution, Model M-113903-USTyco Healthcare Nancy2003—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
199610
200210
FIG. 1 — Lineage of K031470, 1990–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BZD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZD device

  1. #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020886Purtian Bennett, Good Knight 420G Model M-113900-US; Purtan Bennett, Goodknight 420S…Tyco Healthcare Nancy2002—
*K993584Mallinckrodt, Goodknight 418A, Goodknight 418P, Silverlining 2
Summary page 1
…), per 21 CFR 868.5905 3. Predicate Device Equivalence Puritan 13ennett GoodKnight 420s CPAP system, KO20886 Puritan 13ennett GoodKnight 41 8P CPAP system, K993584 4. Device Description The GoodKnight 420 Evolution is designed to deliver Continuous Positive Airway Pressure betwe…
Nellcor Puritan Bennett France Dvpt.2000—

Also cleared against K020886: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.2 years on median.
  • None of the 9 ancestors has a recall.
  • 2 other devices also cite its predicate K020886.

How does its category compare?

All BZD devices

BZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BZD

7in 2025

2007: 212008: 172009: 232010: 122011: 232012: 212013: 232014: 142015: 112016: 122017: 52018: 52019: 62020: 62021: 62022: 42023: 52024: 102025: 72026 (partial year): 10
2007high 23 (2009) · low 4 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZD

189in 2025

2007: 83 days median2008: 108 days median2009: 90 days median2010: 122 days median2011: 125 days median2012: 133 days median2013: 127 days median2014: 172 days median2015: 224 days median2016: 217 days median2017: 221 days median2018: 280 days median2019: 231 days median2020: 137 days median2021: 192 days median2022: 253 days median2023: 204 days median2024: 132 days median2025: 189 days median2026 (partial year): 214 days median
2007high 280 (2018) · low 83 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZD

8in 2023

2007: 292008: 22009: 12011: 12019: 4, 4 Class I2023: 8, 7 Class I2026 (partial year): 2, 2 Class I
2007high 29 (2007) · low 0 (2010)2026

Class I (13 since 2007) Class II/III not classified

Adverse event reports per year · BZD

60,279in 2025

2007: 652008: 672009: 1472010: 1432011: 1892012: 1392013: 822014: 862015: 1142016: 1182017: 1482018: 1012019: 1882020: 1262021: 6,0662022: 90,7682023: 27,9872024: 68,2622025: 60,279
2007high 90,768 (2022) · low 65 (2007)2025

MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721832965
200817108267
200923901147
2010121220143
2011231251189
2012211330139
201323127082
201414172086
2015112240114
2016122170118
201752210148
201852800101
201962314 (4)188
202061370126
2021619206,066
20224253090,768
202352048 (7)27,987
202410132068,262
20257189060,279
2026102142 (2)—

What is it indicated for?

The Puritan Bennett GoodKnight 420 Evolution is intended for use in treating

obstructive sleep apnea (OSA) in spontaneously breathing patients weighing over 30 kg

within a homecare and hospital environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031470: Puritan Bennett, Goodknight 420 Evolution, Model M-113903-US.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031470

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026