510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Endovive Initial Placement Peg Kit, Direct Pej Kit, and Peg Safety Kit

K031538510(k) SpecialCleared Jun 27, 200342-day reviewClass II21 CFR 876.5980Recalls 21

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · KNT Tubes, Gastrointestinal (And Accessories)

Cleared on Jun 27, 2003 as substantially equivalent to K030855 (Boston Scientific Corp, 2003) after a 42-day review. Its family tree reaches back 2 generations to 2002. 3 later devices descend from it. It has been recalled 21 times.

Decided
Jun 27, 2003
Review
42 days
Class
Class II
Product code
KNT
Recalls
21
Predicates
1

Its longest line of descent runs back 2 generations to K014297 (2002). 3 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2015: 1 descendant cleared2006: 1 descendant cleared2018: 1 descendant clearedK014297 · Endovive Initial Placement Gastrostomy Kit; Endovive Low Profile Button Replacement Gastrostomy Tube; Securi-T Initial P · Boston Scientific Corp · 2002K030855 · Modification to Endovive Initial Placement Peg Kit, Direct Pej Kit, Peg Safety Kit · Boston Scientific Corp · 2003 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K030855 · Modification to Endovive Initial Placement Peg Kit, Direct Pej Kit, Peg Safety Kit · Boston Scientific Corp · 2003 · recalledK014297 · Endovive Initial Placement Gastrostomy Kit; Endovive Low Profile Button Replacement Gastrostomy Tube; Securi-T Initial P · Boston Scientific Corp · 20020
  1. Root · 2 generations backK014297 Endovive Initial Placement Gastrostomy Kit; Endovive Low Profile…Boston Scientific Corp · 2002
  2. 1 generation backK030855 Modification to Endovive Initial Placement Peg Kit, Direct Pej Kit…Boston Scientific Corp · 2003 · Recalled
  3. This deviceK031538 Endovive Initial Placement Peg Kit, Direct Pej Kit, and Peg Safety KitBoston Scientific Corp · 2003
  4. Descendant · cites it directly · 1 of its ownK061021 Entake Standard and Safety Peg SystemConmedcorp · 2006 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK182248 Aspire Venting TubeAspire Bariatrics, Inc. · 2018
  6. Descendant · cites it directly · 0 of its ownK150679 EndoVive Initial Placement Standard PEG Kit with ENFit Connector…Boston Scientific Corp · 2015 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K061021Entake Standard and Safety Peg SystemConmedcorp · 2006 · cites it directly1descendant
  • K182248Aspire Venting TubeAspire Bariatrics, Inc. · 2018 · 2 generations on0descendants
  • K150679EndoVive Initial Placement Standard PEG Kit with ENFit Connector, EndoVive Initial…Boston Scientific Corp · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K014297Endovive Initial Placement Gastrostomy Kit; Endovive Low Profile Button…Boston Scientific Corp2002—
−1K030855Modification to Endovive Initial Placement Peg Kit, Direct Pej Kit, Peg Safety…Boston Scientific Corp20032
0 (this device)K031538Endovive Initial Placement Peg Kit, Direct Pej Kit, and Peg Safety KitBoston Scientific Corp2003—
+1K061021Entake Standard and Safety Peg SystemConmedcorp200628
+2K182248Aspire Venting TubeAspire Bariatrics, Inc.2018—
+1K150679EndoVive Initial Placement Standard PEG Kit with ENFit Connector, EndoVive…Boston Scientific Corp20153

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201510
201810
FIG. 1 — Lineage of K031538, 2002–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KNT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KNT device

  1. #1 K261890 52Fr ViSiGi LUX™; 52Fr ViSiGi 3D®score 12.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030855Modification to Endovive Initial Placement Peg Kit, Direct Pej Kit, Peg Safety Kit
Summary page 1
…Classified as a Class I1 Device PREDICATE DEVICES: EndoViveTMInitial Placement PEG Kit (K030855) EndoViveTMInitial Placement Direct PEJ Kit…
Boston Scientific Corp20032

Also cleared against K030855: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061021Entake Standard and Safety Peg SystemConmedcorp20061
K150679EndoVive Initial Placement Standard PEG Kit with ENFit Connector, EndoVive Initial…Boston Scientific Corp20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 20, 2024Class IIZ-1516-2024Process designInability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events…
Feb 20, 2024Class IIZ-1515-2024Process designInability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events…
Feb 20, 2024Class IIZ-1517-2024Process designInability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events…
May 5, 2009Not classifiedZ-1631-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1625-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1632-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1626-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1630-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1627-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1629-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
May 5, 2009Not classifiedZ-1628-2009Process controlThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
Dec 30, 2008Not classifiedZ-0912-2009OtherThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
Dec 30, 2008Not classifiedZ-0913-2009OtherThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
Dec 30, 2008Not classifiedZ-0914-2009OtherThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
Dec 30, 2008Not classifiedZ-0911-2009OtherThe user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
Jun 2, 2005Not classifiedZ-0963-05OtherTwo blades may be heat staked onto the stiletto instead of the expected one blade.
Jun 2, 2005Not classifiedZ-0966-05OtherTwo blades may be heat staked onto the stiletto instead of the expected one blade.
Jun 2, 2005Not classifiedZ-0968-05OtherTwo blades may be heat staked onto the stiletto instead of the expected one blade.
Jun 2, 2005Not classifiedZ-0965-05OtherTwo blades may be heat staked onto the stiletto instead of the expected one blade.
Jun 2, 2005Not classifiedZ-0964-05OtherTwo blades may be heat staked onto the stiletto instead of the expected one blade.
Jun 2, 2005Not classifiedZ-0967-05OtherTwo blades may be heat staked onto the stiletto instead of the expected one blade.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 1 year and passes through 1 company; a new generation was cleared every 9 months on median.
  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K061021 (Conmedcorp, 2006) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K030855.

How does its category compare?

All KNT devices

KNT Tubes, Gastrointestinal (And Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KNT

4in 2025

2007: 102008: 102009: 112010: 122011: 72012: 122013: 72014: 162015: 102016: 62017: 92018: 72019: 102020: 52021: 62022: 72023: 52024: 92025: 42026 (partial year): 3
2007high 16 (2014) · low 4 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KNT

136in 2025

2007: 118 days median2008: 115 days median2009: 102 days median2010: 104 days median2011: 211 days median2012: 107 days median2013: 128 days median2014: 259 days median2015: 140 days median2016: 99 days median2017: 247 days median2018: 149 days median2019: 126 days median2020: 102 days median2021: 121 days median2022: 110 days median2023: 162 days median2024: 103 days median2025: 136 days median2026 (partial year): 105 days median
2007high 259 (2014) · low 99 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KNT

13in 2025

2007: 12008: 822009: 92010: 32011: 22012: 12015: 12016: 32017: 42018: 22019: 22020: 12021: 22022: 1, 1 Class I2023: 22025: 132026 (partial year): 20
2007high 82 (2008) · low 0 (2013)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · KNT

1,451in 2025

2007: 3002008: 8672009: 8182010: 5172011: 5312012: 5282013: 7472014: 1,0302015: 1,0482016: 1,2512017: 1,6722018: 1,9012019: 1,8382020: 1,6412021: 1,6552022: 1,6502023: 1,8632024: 1,3442025: 1,451
2007high 1,901 (2018) · low 300 (2007)2025

MAUDE reports naming product code KNT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007101181300
20081011582867
2009111029818
2010121043517
201172112531
2012121071528
201371280747
20141625901,030
20151014011,048
201669931,251
2017924741,672
2018714921,901
20191012621,838
2020510211,641
2021612121,655
202271101 (1)1,650
2023516221,863
2024910301,344
20254136131,451
2026310520—

What is it indicated for?

EndoVive Initial Placement PEG Kit is indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. The EndoVive Initial Placement Direct PEJ Kit is indicated for nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated. The EndoVive Initial Placement PEG Safety Kit is indicated for…

Read the full statement (107 words)

EndoVive Initial Placement PEG Kit is indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means.

The EndoVive Initial Placement Direct PEJ Kit is indicated for nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated.

The EndoVive Initial Placement PEG Safety Kit is indicated for providing nutrition directly into the stomach in patients who are unable to consume nutrition by conventional means. The safety components of the kit are designed to reduce the potential for inadvertent sharps injury to medical personnel during and after the procedure.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031538: Endovive Initial Placement Peg Kit, Direct Pej Kit, and Peg Safety Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031538

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.