510(k) Report

510(k) · Cleared

Cleared

Apex Electric Handpiece System

K031540510(k) TraditionalCleared Jun 30, 200345-day reviewClass I21 CFR 872.4200

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lares Research, Inc. · EKX Handpiece, Direct Drive, Ac-Powered

Cleared on Jun 30, 2003 as substantially equivalent to K984244 (Bien Air USA, Inc., 1999) and 2 other predicates after a 45-day review. Its family tree reaches back 2 generations to 1979.

Decided
Jun 30, 2003
Review
45 days
Class
Class I
Product code
EKX
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K792159 (1979). 5 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2000: 1 other ancestor1997: 1 other ancestor1994: 1 other ancestorK792159 · Fisch Drill System · Jedmed Instrument Co. · 1979 · recalledK984244 · Orl-E-92 Surgical Drill System · Bien Air USA, Inc. · 19991975198019851990199520002005
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor K984244 · Orl-E-92 Surgical Drill System · Bien Air USA, Inc. · 1999K792159 · Fisch Drill System · Jedmed Instrument Co. · 1979 · recalled0
  1. Root · 2 generations backK792159 Fisch Drill SystemJedmed Instrument Co. · 1979 · Recalled
  2. 1 generation backK984244 Orl-E-92 Surgical Drill SystemBien Air USA, Inc. · 1999
  3. This deviceK031540 Apex Electric Handpiece SystemLares Research, Inc. · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K792159Fisch Drill SystemJedmed Instrument Co.19791
−1K984244Orl-E-92 Surgical Drill SystemBien Air USA, Inc.1999—
0 (this device)K031540Apex Electric Handpiece SystemLares Research, Inc.2003—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200010
FIG. 1 — Lineage of K031540, 1979–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in EKX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EKX device

  1. #1 K261987 Reusable SmartPegscore 13.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K984244Orl-E-92 Surgical Drill System
Summary page 1
…ny Device 510(k) No. Date Cleared Bien Air USA, Inc. ORL-E-92 Surgical Drill System K984244 02/23/1999 W&H Synea WA-99LT K993526 01111/2000 NSK…
Bien Air USA, Inc.1999—
*K993526Synea - LS Low-Speed Handpiece
Summary page 1
…n Air USA, Inc. ORL-E-92 Surgical Drill System K984244 02/23/1999 W&H Synea WA-99LT K993526 01111/2000 NSK Speed Increaser Contra Angle K972569 10/08/1997 The ApexTMElectric…
A-Dec, Inc.2000—
*K972569E-Type Speed Increaser Contra Angle, External Irrigation, Internal Irrigation
Summary page 1
…W&H Synea WA-99LT K993526 01111/2000 NSK Speed Increaser Contra Angle K972569 10/08/1997 The ApexTMElectric Handpiece System is a pneumatic/electronic controlled drive for dental drilling applications.…
Nsk Nakanishi, Inc.1997—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 19.2 years on median.
  • 1 of the 5 ancestors has a recall of any class.

How does its category compare?

All EKX devices

EKX Handpiece, Direct Drive, Ac-Powered: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · EKX

5in 2025

2007: 12008: 32009: 42010: 32011: 42012: 22013: 22014: 32015: 32016: 32017: 22018: 32019: 32020: 12021: 52022: 32023: 42024: 42025: 52026 (partial year): 1
2007high 5 (2021) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EKX

244in 2025

2008: 101 days median2009: 186 days median2010: 114 days median2011: 75 days median2014: 251 days median2015: 178 days median2016: 274 days median2018: 293 days median2019: 349 days median2021: 214 days median2022: 359 days median2023: 152 days median2024: 276 days median2025: 244 days median
2007high 359 (2022) · low 75 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EKX

No recall names EKX since 2007.

Adverse event reports per year · EKX

9in 2025

2007: 92008: 102009: 112010: 202011: 62012: 272013: 452014: 1022015: 682016: 662017: 102018: 492019: 562020: 672021: 792022: 372023: 142024: 192025: 9
2007high 102 (2014) · low 6 (2011)2025

MAUDE reports naming product code EKX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—09
20083101010
20094186011
20103114020
201147506
20122—027
20132—045
201432510102
20153178068
20163274066
20172—010
20183293049
20193349056
20201—067
20215214079
20223359037
20234152014
20244276019
2025524409
20261—0—

What is it indicated for?

The Lares Research Apex” Electric Handpiece System (“System” defined as and limited to: A. Contra Angle Dental Instrument B. Electric Motor and C. Control System) device consists of a dental handpiece and accessories as listed in 21 CFR 872.4200, product code EKX that is a Class I, non-exempt hand-held dental device for dental drilling applications. This device is indicated for use where high speeds are required in general dentistry with or without the…

Read the full statement (215 words)

The Lares Research Apex” Electric Handpiece System (“System” defined as and limited to: A. Contra Angle Dental Instrument B. Electric Motor and C. Control System) device consists of a dental handpiece and accessories as listed in 21 CFR 872.4200, product code EKX that is a Class I, non-exempt hand-held dental device for dental drilling applications. This device is indicated for use where high speeds are required in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as cutting a tooth for caries, and/or crown preparation and finishing, cutting and/or finishing for dentures, denture bases, crowns, inlays and metal plates and/or root canal preparation for restoration. The device is also indicated for use in difficult to reach areas in general dentistry with or without the use of coolant where precision control of the cutting bur is desired such as the preparing dental caries for restorations, and for use at low speeds in teeth cleaning applications and dental surface polishing. The Apex” Electric Handpiece System device is driven by an electronically controlled DC electric motor with a pneurnatic/electronicactuated control system that is connected to the Contra Angle dental instrument via an 180-3964 coupling. The Apex” Electric Handpiece System device is indicated for use by dental professionals only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031540: Apex Electric Handpiece System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K031540

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026