510(k) Report

510(k) · Cleared

Recalled · 2005

Syndeo Pca Syringe Pump, Model 2L3113

K031566510(k) TraditionalCleared Aug 19, 200392-day reviewClass II21 CFR 880.5725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · MEA Pump, Infusion, Pca

Cleared on Aug 19, 2003 after a 92-day review. The FDA summary names its predicate without a number. It has been recalled once.

Decided
Aug 19, 2003
Review
92 days
Class
Class II
Product code
MEA
Recalls
1
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K031566Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MEA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—PCA II Syringe Infusion Pump
Named without a submission number
—
*—iPump Pain Management System
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 6, 2005Not classifiedZ-1373-05Device DesignDuring an infusion in the 'Basal + PCA' or 'PCA only' modes, the SYNDEO pump may generate false Service Code 919 or Service Code 920, which will cause the pump to stop the infusion. Also, while attempting to turn on the device by pressing…

No ancestor or descendant has a Class I recall.

How does its category compare?

All MEA devices

MEA Pump, Infusion, Pca: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MEA

2in 2017

2008: 12009: 12013: 12016: 12017: 2
2007high 2 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEA

Too few MEA 510(k)s in any year since 2007 to give a median.

Recalls per year · MEA

1in 2024

2008: 12010: 42011: 22014: 12017: 22022: 12024: 1
2007high 4 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MEA

3,278in 2025

2007: 812008: 2372009: 1532010: 2192011: 2452012: 4462013: 4532014: 3822015: 992016: 572017: 442018: 2562019: 1322020: 2,3312021: 3,7762022: 6,4432023: 3,9842024: 4,5592025: 3,278
2007high 6,443 (2022) · low 44 (2017)2025

MAUDE reports naming product code MEA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—081
20081—1237
20091—0153
20100—4219
20110—2245
20120—0446
20131—0453
20140—1382
20150—099
20161—057
20172—244
20180—0256
20190—0132
20200—02,331
20210—03,776
20220—16,443
20230—03,984
20240—14,559
20250—03,278
20260—0—

What is it indicated for?

The Syndeo PCA Syringe Pump is designed for the controlled delivery (continuous, intermittent, and continuous plus intermittent) of analgesic, sedative, and anesthetic solutions through clinically acceptable routes of administration including intravenous, subcutaneous, spinal (or intrathecal), epidural (or subarachnoid), and for regional anesthesia applications.

The Syndeo PCA Syringe Pump is appropriate for use in hospital or alternate medical site environments, but it is not intended for use in homecare settings.

The Syndeo PCA Syringe Pump is not indicated for use with blood or blood products.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031566: Syndeo Pca Syringe Pump, Model 2L3113.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031566

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026