510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Vitros 5,1 FS Chemistry System, Vitros Chemistry Products DHDL Reagent Pack, Calibrator Kit 19, FS Calibrator 1

K031924510(k) TraditionalCleared Aug 7, 200345-day reviewClass II21 CFR 862.1150Recalls 29

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics · JIT Calibrator, Secondary

Cleared on Aug 7, 2003 as substantially equivalent to K946090 (Clinical Diagnostic Systems, Inc., 1995) and 2 other predicates after a 45-day review. Its family tree reaches back 1 generation to 1995. It has been recalled 29 times.

Decided
Aug 7, 2003
Review
45 days
Class
Class II
Product code
JIT
Recalls
29
Predicates
3

Its longest line of descent runs back 1 generation to K946090 (1995). 3 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1999: 1 other ancestor1998: 1 other ancestorK946090 · Kodak Ektacchem 950IRC Analyzer · Clinical Diagnostic Systems, Inc. · 1995 · recalled19961998200020022004
Other ancestors, by generation 1 generation back: 2 other ancestors K946090 · Kodak Ektacchem 950IRC Analyzer · Clinical Diagnostic Systems, Inc. · 1995 · recalled0
  1. Root · 1 generation backK946090 Kodak Ektacchem 950IRC AnalyzerClinical Diagnostic Systems, Inc. · 1995 · Recalled
  2. This deviceK031924 Vitros 5,1 FS Chemistry System, Vitros Chemistry Products DHDL…Ortho-Clinical Diagnostics · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K946090Kodak Ektacchem 950IRC AnalyzerClinical Diagnostic Systems, Inc.19951
0 (this device)K031924Vitros 5,1 FS Chemistry System, Vitros Chemistry Products DHDL Reagent Pack…Ortho-Clinical Diagnostics2003—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
199910
FIG. 1 — Lineage of K031924, 1995–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JIT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K946090Kodak Ektacchem 950IRC Analyzer
Summary page 2
…cate devices 1. VITROS Chemistry Products 5,l 1. VITROS 950 Chemistry System FDA FS Chemistry System 5 1O(k) Number K946090 2. Bayer ADVIA 1650 Chemistry System…
Clinical Diagnostic Systems, Inc.19951
*K990346Bayer Advia 1650 Chemistry System
Summary page 2
…946090 2. Bayer ADVIA 1650 Chemistry System FDA 5 1O(k) Number K990346 2. VITROS Chemistry Products 3. Bayer Direct HDL Cholesterol I1 dHDL assay…
Bayer Corp.199914
*K982341Direct HDL-Cholestrol Liquid
Summary page 2
…r K990346 2. VITROS Chemistry Products 3. Bayer Direct HDL Cholesterol I1 dHDL assay FDA Docket Number K982341 1.6 Intended Use Statement(s) The VITROS 5,l FS Chemistry System is a random access, fully automated c…
Randox Laboratories, Ltd.19981

Also cleared against K946090: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 9, 2025Class IIZ-2040-2025Software designa software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during…
Jan 26, 2024Class IIZ-1225-2024Software Design ChangeA software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Jan 26, 2024Class IIZ-1230-2024Software Design ChangeA software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not…
Oct 10, 2023Class IIZ-0544-2024Under Investigation by firmDuring planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or…
Aug 9, 2023Class IIZ-0008-2024Under Investigation by firmOrtho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification…
Dec 21, 2015Class IIZ-1075-2016Error in labelingIncreased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition…
Dec 21, 2015Class IIZ-1073-2016Error in labelingIncreased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition…
Nov 3, 2015Class IIZ-0643-2016Software designSoftware anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems…
Nov 3, 2015Class IIZ-0642-2016Software designSoftware anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems…
Sep 10, 2015Class IIZ-0231-2016Software designSoftware Anomaly during ADD Installation on VITROS 5,1 FS Chemistry Systems using Software Versions 2.2.1 through 2.8. The user modified parameters would not be saved as expected, but remained changed to the default values obtained from…
Jun 30, 2015Class IIZ-2544-2015Software designSoftware anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software Versions 2.2.1 through 2.8. If the user is utilizing User- Configured (Modified) parameters, ALL default settings are restored in the VITROS…
Mar 10, 2015Class IIZ-1521-2015Software in the Use EnvironmentSoftware Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series…
Nov 19, 2014Class IIZ-0967-2015Software designSoftware Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Nov 19, 2014Class IIZ-0969-2015Software designSoftware Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Nov 19, 2014Class IIZ-0968-2015Software designSoftware Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Sep 29, 2014Class IIIZ-1004-2015Software Manufacturing/Software DeploymentCalibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
Sep 8, 2014Class IIZ-0077-2015Software designOrtho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied…
Jul 7, 2014Class IIZ-2140-2014Software in the Use EnvironmentOrtho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample…
Apr 10, 2014Class IIZ-1736-2014Software designPotential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with…
Apr 10, 2014Class IIZ-1734-2014Software designPotential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with…
Jun 6, 2013Class IIZ-1948-2013Nonconforming Material/ComponentOrtho Clinical Diagnostics is recalling CM/RT wear pads which are a component in the CM/RT incubator due to becoming worn before their replacement.
May 6, 2013Class IIZ-1647-2013Device DesignThere is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
May 6, 2013Class IIZ-1648-2013Device DesignThere is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Feb 7, 2013Class IIZ-1106-2013Nonconforming Material/ComponentOrtho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal…
Feb 7, 2013Class IIZ-1107-2013Nonconforming Material/ComponentOrtho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal…
May 21, 2012Class IIZ-2067-2012Software designsoftware automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment.
May 21, 2012Class IIZ-2066-2012Software designsoftware automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment.
Mar 15, 2012Not classifiedZ-1469-2012Nonconforming Material/ComponentOrtho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure.
May 17, 2007Not classifiedZ-1167-2007OtherSoftware design. The firm received a customer complaint regarding a qualitative positive THC (Cannabinoids) result obtained from a THC quality control fluid that was known to be negative when using VITROS Chemistry Products THC Reagent. A…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 3 of the 3 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K946090.

How does its category compare?

All JIT devices

JIT Calibrator, Secondary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JIT

7in 2017

2007: 52008: 62009: 22010: 62011: 72012: 22013: 52014: 102015: 42016: 42017: 7
2007high 10 (2014) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JIT

28in 2017

2007: 78 days median2008: 46 days median2010: 65 days median2011: 48 days median2013: 140 days median2014: 33 days median2015: 30 days median2016: 33 days median2017: 28 days median
2007high 140 (2013) · low 28 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JIT

1in 2023

2007: 12009: 22011: 32012: 32014: 22015: 12018: 52020: 22023: 12026 (partial year): 1
2007high 5 (2018) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JIT

7in 2025

2010: 32011: 22012: 102013: 432014: 92015: 62016: 32017: 72018: 82019: 2152020: 122021: 32022: 42023: 62024: 172025: 7
2007high 215 (2019) · low 0 (2007)2025

MAUDE reports naming product code JIT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200757810
200864600
20092—20
201066503
201174832
20122—310
20135140043
2014103329
201543016
201643303
201772807
20180—58
20190—0215
20200—212
20210—03
20220—04
20230—16
20240—017
20250—07
20260—1—

What is it indicated for?

The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.) For in vitro diagnostic use only. The VITROS dHDL Reagent Pack…

Read the full statement (165 words)

The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)

For in vitro diagnostic use only. The VITROS dHDL Reagent Pack quantitatively measures HDL cholesterol (HDLC) concentration in serum and plasma. High Density Lipoprotein (HDL) cholesterol is used to evaluate the risk of developing coronary heart disease (CHD). The risk of CHD increases with lower HDL cholesterol concentrations.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 19 is used on the VITROS 5,1 FS Chemistry System in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate dHDL.

For in vitro diagnostic use only. VITROS Chemistry Products FS Calibrator 1 is a saline solution used on the VITROS 5,1 FS Chemistry System in conjunction with VITROS Calibrator Kit 19 to calibrate dHDL.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K031924: Vitros 5,1 FS Chemistry System, Vitros Chemistry Products DHDL Reagent Pack, Calibrator Kit 19, FS Calibrator 1.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031924

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026