Philips Viewforum 2003

K032096510(k) TraditionalCleared Jul 18, 200310-day reviewClass II21 CFR 892.2050Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · LLZ System, Image Processing, Radiological

Cleared on Jul 18, 2003 as substantially equivalent to K920950 (Philips Medical Systems, Inc., 1992) and 1 other predicate after a 10-day review. Its family tree reaches back 2 generations to 1992. 6 later devices descend from it. It has been recalled 2 times.

Decided
Jul 18, 2003
Review
10 days
Class
Class II
Product code
LLZ
Recalls
2
Predicates
2

Its longest line of descent runs back 2 generations to K920950 (1992). 6 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2011: 1 descendant cleared2008: 1 descendant cleared2006: 1 descendant cleared2014: 2 descendants cleared2012: 1 descendant clearedK920950 · Philips Easyvision · Philips Medical Systems, Inc. · 1992 · recalledK023137 · Philips Easyvision Workstation Release 6 · Philips Medical Systems North America Co. · 20021990200020102020
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant K023137 · Philips Easyvision Workstation Release 6 · Philips Medical Systems North America Co. · 2002K920950 · Philips Easyvision · Philips Medical Systems, Inc. · 1992 · recalled0
  1. Root · 2 generations backK920950 Philips EasyvisionPhilips Medical Systems, Inc. · 1992 · Recalled
  2. 1 generation backK023137 Philips Easyvision Workstation Release 6Philips Medical Systems North America Co. · 2002
  3. This deviceK032096 Philips Viewforum 2003Philips Medical Systems · 2003
  4. Descendant · cites it directly · 2 of its ownK081262 DC NeuroInvivo Corporation · 2008
  5. Descendant · 2 generations on · 1 of its ownK113701 Visia NeuroMeVis Medical Solutions AG · 2012
  6. Descendant · cites it directly · 1 of its ownK060453 Aze VirtualplaceAze, Ltd. · 2006
  7. Descendant · 3 generations on · 0 of its ownK141514 Synapse 3D Tensor AnalysisFujifilm Meidcal Systems U.S.A.,Inc. · 2014
  8. Descendant · 2 generations on · 0 of its ownK133521 Aze Phoenix SoftwareAze, Ltd. · 2014

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2006: 12008: 12011: 12012: 12014: 2
2006high 2 (2014) · low 0 (2007)2026

Most influential descendants

Showing 5 of 6 · See all
  • K081262DC NeuroInvivo Corporation · 2008 · cites it directly2descendants
  • K113701Visia NeuroMeVis Medical Solutions AG · 2012 · 2 generations on1descendant
  • K060453Aze VirtualplaceAze, Ltd. · 2006 · cites it directly1descendant
  • K141514Synapse 3D Tensor AnalysisFujifilm Meidcal Systems U.S.A.,Inc. · 2014 · 3 generations on0descendants
  • K133521Aze Phoenix SoftwareAze, Ltd. · 2014 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K920950Philips EasyvisionPhilips Medical Systems, Inc.19921
−1K023137Philips Easyvision Workstation Release 6Philips Medical Systems North America Co.2002—
0 (this device)K032096Philips Viewforum 2003Philips Medical Systems2003—
+1K081262DC NeuroInvivo Corporation2008—
+2K113701Visia NeuroMeVis Medical Solutions AG2012—
+1K060453Aze VirtualplaceAze, Ltd.2006—
+3K141514Synapse 3D Tensor AnalysisFujifilm Meidcal Systems U.S.A.,Inc.2014—
+2K133521Aze Phoenix SoftwareAze, Ltd.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200810
201110
201210
201420
FIG. 1 — Lineage of K032096, 1992–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K920950Philips Easyvision
Summary page 2
…are taken. Substantial equivalence: The Philips ViewForum 2003 is substantially equivalent to the Easyvision Workstation systems (K920950) and Easyvision Workstation Release 6 (KO23137.) Philips Medical Systems Re: KO32096 North…
Philips Medical Systems, Inc.19921
*K023137Philips Easyvision Workstation Release 6Philips Medical Systems North America Co.2002—

Also cleared against K920950: 4 other devices →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081262DC NeuroInvivo Corporation20082
K060453Aze VirtualplaceAze, Ltd.20061
K110077NeucoloncareNeusoft Medical Systems Co., Ltd.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 23, 2011Not classifiedZ-1882-2011Software designA software anomaly that occurs when clinicians use Volume Vision to view images in 3D and uses the measurement tool to measure regions of interest. This could lead to an incorrect interpretation of the measurements which could cause an…
Apr 29, 2005Not classifiedZ-0896-05OtherAutomated analysis of vascular obstruction tool in software gives unreliable calibration and analysis results when using unprocessed images acquired on the Philips Allura FD10.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 10.6 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K081262 (Invivo Corporation, 2008) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K920950.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

ViewForum 2003 is an image processing workstation software package designed to run on standard PC hardware. The hardware required is made up of 'off-the-shelf" standard computer components. ViewForum 2003 software receives image data from medical scanning devices such as CT and MRI, or from image archives. It performs viewing, image manipulation, communication, printing and quantification of images.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K032096: Philips Viewforum 2003.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K032096

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026