# K032226: LTV-1000 Ventilator

- Pathway: 510(k) (Traditional)
- Applicant: Pulmonetic Systems, Inc.
- Decided: Oct 16, 2003 after 87 days
- Product code: CBK (Ventilator, Continuous, Facility Use)
- Device class: 2
- Regulation: 21 CFR 868.5895
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
59 later devices descend from it; 4 cite it directly.
19 of 60 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA’s file for it is a 510(k) statement, which names no predicate.

## Most influential descendants
- K051767: LTV 1000 Ventilator (2005), 57 descendants of its own
- K060647: LTV 1200 Ventilator (2006), 29 descendants of its own
- K083526: Trilogy 100 Ventilator (2009), 26 descendants of its own
- K093416: Trilogy 200 Ventilator (2010), 19 descendants of its own
- K070594: Palmtop Ventilator, Models PTV-8, PTV-10 (2007), 12 descendants of its own

## Recalls of this device
- Sep 8, 2004 · Not classified · Z-1485-04: Ventilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation t

## Indications for use
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 Ibs), who require the following types of ventilatory support: Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). Assist/Control, SIMV, or CPAP modes of ventilation. The ventilator is suitable for use in institutional, home, or transport settings.

Source page: https://510kreport.com/device/K032226