# K032370: Cic Pro Clinical Information Center Central Station

- Pathway: 510(k) (Special)
- Applicant: Ge Medical Systems Information Technologies
- Decided: Aug 13, 2003 after 12 days
- Product code: DSI (Detector And Alarm, Arrhythmia)
- Device class: 2
- Regulation: 21 CFR 870.1025
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K001112 (2000). 1 ancestor in all.
23 later devices descend from it; 1 cites it directly.
0 of 25 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K001112: GE Marquette Clinical Information Center, GE Marquette Cic, General Electric Medical Systems Information Techn (2000)

## Most influential descendants
- K053356: Modification to: Cic Pro Clinical Information Center Central Station (2006), 22 descendants of its own
- K073372: Electrophysiology Systems Integrator (Epsi) (2007), 10 descendants of its own
- K062976: Modification to Cic Pro Clinical Information Center (2007), 10 descendants of its own
- K133882: Carescape Central Station (Formally Known as Cic Pro) (2014), 9 descendants of its own
- K083321: Cardio-View (2008), 9 descendants of its own

## Recalls of this device
- Nov 12, 2019 · Class II · Z-0726-2020: When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) v
- Feb 27, 2009 · Not classified · Z-1232-2009: GE Healthcare is aware of potential network time issues associated with the time change aspects of certain CARESCAPE CIC Pro" software versions. These issues may impact patient safety. No reported pat
- Jan 24, 2007 · Not classified · Z-0280-2008: Two separate issues may occur with operating The CiC Pro v4.1 System with Telemetry. 1) Experience a loss of audible alarms when operating system memory resource reaches approximately 170MG. This will
- Jan 24, 2007 · Not classified · Z-0891-2007: Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of aud
- Oct 11, 2004 · Not classified · Z-0030-05: The following two conditions may occur with the printing of full disclosure strips and reports, printing of the alarm history events stored on a GE Unity Network Patient Data Server (PDS), and printin

## Indications for use
The CIC Pro Clinical Information Center Central Station is intended to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. The CIC Pro Central Station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and / or other non-medical information. Physiological parameters and waveforms from GE Medical Systems Information Technologies monitors and telemetry systems can be displayed and printed from the CIC Pro Central Station. Beat to beat patient information for all parameters and waveforms from the bedside and telemetry systems can be displayed. The CIC Pro Central Station supports the ability to access information from GE Medical Systems Information Technologies’ products in a web browser format. Additionally, CIC Pro Central Station supports the ability to access patient information collected from the Unity network and stored on a network server

Source page: https://510kreport.com/device/K032370