510(k) · Cleared
Recalled · Class II · 2025Kung Shin, Hmef SK203
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Kung Shin Plastics Co., Ltd. · CAH Filter, Bacterial, Breathing-Circuit
Cleared on Sep 20, 2004 after a 371-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 2 times.
- Decided
- Sep 20, 2004
- Review
- 371 days
- Class
- Class II
- Product code
- CAH
- Recalls
- 2
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in CAH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new CAH device
- #1 K251583 Ambu® Virobac II® Exhalation Filterscore 11.6 · 2 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 9, 2025 | Class II | Z-1269-2025 | Process design | The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation.… |
| Jan 9, 2025 | Class II | Z-1268-2025 | Process design | The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation.… |
No ancestor or descendant has a Class I recall.
How does its category compare?
All CAH devicesCAH Filter, Bacterial, Breathing-Circuit: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · CAH
2in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · CAH
237in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · CAH
1in 2025
Class I (7 since 2007) Class II/III not classified
Adverse event reports per year · CAH
77in 2025
MAUDE reports naming product code CAH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 32 |
| 2008 | 1 | — | 0 | 27 |
| 2009 | 4 | 111 | 0 | 18 |
| 2010 | 5 | 216 | 2 | 40 |
| 2011 | 3 | 281 | 0 | 9 |
| 2012 | 3 | 212 | 0 | 16 |
| 2013 | 1 | — | 1 | 47 |
| 2014 | 4 | 250 | 1 | 98 |
| 2015 | 2 | — | 0 | 51 |
| 2016 | 0 | — | 0 | 34 |
| 2017 | 1 | — | 1 | 29 |
| 2018 | 0 | — | 0 | 33 |
| 2019 | 2 | — | 0 | 30 |
| 2020 | 2 | — | 0 | 33 |
| 2021 | 2 | — | 0 | 56 |
| 2022 | 3 | 204 | 6 (6) | 67 |
| 2023 | 3 | 237 | 0 | 40 |
| 2024 | 2 | — | 0 | 51 |
| 2025 | 0 | — | 1 (1) | 77 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
The HMEF SK203 is a disposable single-use device indicated for adult patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross contamination between patient and equipment. The HMEF SK203 is for use in hospital, ICU, anesthesia, respiratory therapy, during transport and with resuscitators. The device can be used on adult patients. The device is indicated for use by qualified medical personnel only.
Indications for use as stated in the FDA 510(k) statement.
https://510kreport.com/device/K032878 · printed Nov 4, 2026
Cite this page
510(k) Report. “K032878: Kung Shin, Hmef SK203.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K032878
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026