510(k) Report

510(k) · Cleared

Recalled · 2010

Venturi Phaco Pack

K033653510(k) SpecialCleared Dec 17, 200326-day reviewClass II21 CFR 886.4670Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bausch & Lomb, Inc. · HQC Unit, Phacofragmentation

Cleared on Dec 17, 2003 as substantially equivalent to K955901 (Storz Instrument Co., 1996) after a 26-day review. Its family tree reaches back 2 generations to 1988. It has been recalled 3 times.

Decided
Dec 17, 2003
Review
26 days
Class
Class II
Product code
HQC
Recalls
3
Predicates
2

Its longest line of descent runs back 2 generations to K882240 (1988). No later device cites it as a predicate yet.

  1. Root · 2 generations backK882240 Storz Univ Irrigat/Aspira & Phacoemulsifcation PacStorz Instrument Co. · 1988
  2. 1 generation backK955901 Storz DP4310DELUXE Phaco Pack Without RefluxStorz Instrument Co. · 1996
  3. This deviceK033653 Venturi Phaco PackBausch & Lomb, Inc. · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K882240Storz Univ Irrigat/Aspira & Phacoemulsifcation PacStorz Instrument Co.1988—
−1K955901Storz DP4310DELUXE Phaco Pack Without RefluxStorz Instrument Co.1996—
0 (this device)K033653Venturi Phaco PackBausch & Lomb, Inc.2003—
FIG. 1 — Lineage of K033653, 1988–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HQC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HQC device

  1. #1 K261264 Stellaris Elite™ vision enhancement system…score 13.4 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K955901Storz DP4310DELUXE Phaco Pack Without Reflux
Summary page 1
…ial and irrigant from the eye Substantial Equivalent Basis The modified DP43 10 and new DP4305 Venturi Phaco Packs are substantially equivalent to the existing DP4310 (K955901) and DP4330 and DP4345 (combination K955901 and K952259). See Comparison Matrix.…
Storz Instrument Co.1996—
Kept off the family treeK955901, K952259DP4330
Number could not be read

Also cleared against K955901: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 8, 2010Not classifiedZ-1709-2010Component design/selectionUsers may incorrectly assemble the plastic needle wrench to the phacoemulsification needle resulting in the generation of plastic particulate
Jan 8, 2010Not classifiedZ-1708-2010Component design/selectionUsers may incorrectly assemble the plastic needle wrench to the phacoemulsification needle resulting in the generation of plastic particulate
Jan 8, 2010Not classifiedZ-1711-2010Component design/selectionUsers may incorrectly assemble the plastic needle wrench to the phacoemulsification needle resulting in the generation of plastic particulate

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 2 companies; a new generation was cleared every 7.7 years on median.
  • None of the 2 ancestors has a recall.
  • 3 other devices also cite its predicate K955901.

How does its category compare?

All HQC devices

HQC Unit, Phacofragmentation: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HQC

3in 2025

2008: 42009: 52010: 32011: 52012: 52014: 42015: 32016: 12017: 32019: 42020: 42021: 12022: 42023: 12024: 52025: 32026 (partial year): 3
2007high 5 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HQC

267in 2025

2008: 162 days median2009: 131 days median2010: 142 days median2011: 129 days median2012: 125 days median2014: 135 days median2015: 124 days median2017: 214 days median2019: 126 days median2020: 227 days median2022: 245 days median2024: 197 days median2025: 267 days median2026 (partial year): 115 days median
2007high 267 (2025) · low 124 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HQC

1in 2025

2008: 12009: 12010: 222011: 42012: 12014: 12018: 22022: 52023: 12025: 1
2007high 22 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HQC

1,799in 2025

2007: 3412008: 4442009: 5402010: 2,4312011: 1,8022012: 2,2772013: 2,7722014: 2,7402015: 3,3552016: 3,3742017: 2,7812018: 2,4162019: 2,2872020: 1,2172021: 1,6732022: 1,9072023: 1,7992024: 2,1202025: 1,799
2007high 3,374 (2016) · low 341 (2007)2025

MAUDE reports naming product code HQC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—0341
200841621444
200951311540
20103142222,431
2011512941,802
2012512512,277
20130—02,772
2014413512,740
2015312403,355
20161—03,374
2017321402,781
20180—22,416
2019412602,287
2020422701,217
20211—01,673
2022424551,907
20231—11,799
2024519702,120
2025326711,799
202631150—

What is it indicated for?

The Venturi Phaco Packs are intended for use with a Bausch & Lomb Millennium, Protegd or Premiere microsurgical system for the phacoemulsification of an opacified crystalline lens during anterior segment surgery.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033653: Venturi Phaco Pack.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033653

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026