510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Modified Coated Vicryl* Rapide (Polyglactin 910) Suture

K033746510(k) SpecialCleared Dec 18, 200317-day reviewClass II21 CFR 878.4493Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

ETHICON, Inc. · GAM Suture, Absorbable, Synthetic, Polyglycolic Acid

Cleared on Dec 18, 2003 after a 17-day review. The FDA summary names its predicate without a number. 6 later devices descend from it. It has been recalled 3 times.

Decided
Dec 18, 2003
Review
17 days
Class
Class II
Product code
GAM
Recalls
3
Predicates
0

6 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2014: 1 descendant cleared2012: 1 descendant cleared2017: 1 descendant cleared2008: 1 descendant cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK033746 Modified Coated Vicryl* Rapide (Polyglactin 910) SutureETHICON, Inc. · 2003
  2. Descendant · cites it directly · 2 of its ownK120556 Vilet Vilet QuickRiverpoint Medical · 2012 · Recalled
  3. Descendant · cites it directly · 1 of its ownK061796 Hepatostat Set, Model 760XVygon Corp. · 2006
  4. Descendant · 2 generations on · 0 of its ownK263073 Fast Absorbing PGLARiverpoint Medical, LLC · 2026
  5. Descendant · cites it directly · 0 of its ownK170166 CARESORB-Polyglactin 910 Surgical Suture, CARESORB-Polyglactin 910…Cpt Sutures Co., Ltd. · 2017
  6. Descendant · cites it directly · 0 of its ownK123940 Demequicktm (Rapid Absorbable) Synthetic Surgical SutureDemetech Corp. · 2014

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2006: 12008: 12012: 12014: 12017: 12026: 1
2006high 1 (2006) · low 0 (2007)2026

Most influential descendants

Showing 5 of 6 · See all
  • K120556Vilet Vilet QuickRiverpoint Medical · 2012 · cites it directly2descendants
  • K061796Hepatostat Set, Model 760XVygon Corp. · 2006 · cites it directly1descendant
  • K263073Fast Absorbing PGLARiverpoint Medical, LLC · 2026 · 2 generations on0descendants
  • K170166CARESORB-Polyglactin 910 Surgical Suture, CARESORB-Polyglactin 910 (fast absorbing)…Cpt Sutures Co., Ltd. · 2017 · cites it directly0descendants
  • K123940Demequicktm (Rapid Absorbable) Synthetic Surgical SutureDemetech Corp. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K033746Modified Coated Vicryl* Rapide (Polyglactin 910) SutureETHICON, Inc.2003—
+1K120556Vilet Vilet QuickRiverpoint Medical201241
+1K061796Hepatostat Set, Model 760XVygon Corp.2006—
+2K263073Fast Absorbing PGLARiverpoint Medical, LLC2026—
+1K170166CARESORB-Polyglactin 910 Surgical Suture, CARESORB-Polyglactin 910 (fast…Cpt Sutures Co., Ltd.2017—
+1K123940Demequicktm (Rapid Absorbable) Synthetic Surgical SutureDemetech Corp.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200810
201210
201410
201710
202610
FIG. 1 — Lineage of K033746, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GAM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GAM device

  1. #1 K252743 STRATAFIX™ Spiral PDS™ Plus Knotless Tissue…score 14.8 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Coated VICRYL* Rapide (Polyglactin 910) Suture
Named without a submission number
—

What descends from it? (6)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120556Vilet Vilet QuickRiverpoint Medical20122
K061796Hepatostat Set, Model 760XVygon Corp.20061
K170166CARESORB-Polyglactin 910 Surgical Suture, CARESORB-Polyglactin 910 (fast absorbing)…Cpt Sutures Co., Ltd.20170
K123940Demequicktm (Rapid Absorbable) Synthetic Surgical SutureDemetech Corp.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 17, 2017Class IIZ-0693-2018Equipment maintenanceOn August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture…
Apr 20, 2010Not classifiedZ-1608-2010OtherChannels in the packaging seal that can lead to suture degradation and loss of packaging integrity. Recall Z-0021-2009 was initiated in 2008. Ethicon had decided to recall all sutures product during the may to June 2007 timeframe.
Aug 11, 2008Not classifiedZ-0021-2009PackagingPackage defect compromised the integrity of the primary seal, which could lead to premature suture degradation and/or impair the sterile barrier of the product.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K120556 (Riverpoint Medical, 2012) has gone furthest, with 2 descendants of its own.

How does its category compare?

All GAM devices

GAM Suture, Absorbable, Synthetic, Polyglycolic Acid: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GAM

3in 2025

2007: 72008: 62009: 32010: 12011: 22012: 52013: 12014: 102015: 52016: 32017: 82018: 52019: 42020: 32021: 22022: 22023: 12024: 22025: 32026 (partial year): 4
2007high 10 (2014) · low 1 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GAM

29in 2025

2007: 50 days median2008: 174 days median2009: 10 days median2012: 31 days median2014: 121 days median2015: 209 days median2016: 169 days median2017: 134 days median2018: 98 days median2019: 90 days median2020: 140 days median2025: 29 days median2026 (partial year): 131 days median
2007high 209 (2015) · low 10 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GAM

1in 2024

2007: 102008: 12009: 62010: 32011: 62016: 12017: 12019: 12020: 442022: 42024: 12026 (partial year): 2
2007high 44 (2020) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GAM

8,970in 2025

2007: 2742008: 1942009: 2872010: 5872011: 6372012: 8682013: 8982014: 1,4102015: 9352016: 1,1122017: 1,4992018: 2,2292019: 3,5642020: 3,4152021: 4,5062022: 4,8142023: 5,0702024: 7,2262025: 8,970
2007high 8,970 (2025) · low 194 (2008)2025

MAUDE reports naming product code GAM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200775010274
200861741194
20093106287
20101—3587
20112—6637
20125310868
20131—0898
20141012101,410
201552090935
2016316911,112
2017813411,499
201859802,229
201949013,564
20203140443,415
20212—04,506
20222—44,814
20231—05,070
20242—17,226
202532908,970
202641312—

What is it indicated for?

Modified Coated VICRYL* Rapide synthetic absorbable suture is indicated only for use in superficial soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required. Coated VICRYL Rapide suture is not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033746: Modified Coated Vicryl* Rapide (Polyglactin 910) Suture.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033746

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026